Arthrex® AR-8724V-20SRecalls & reported problems
Arthrex, Inc. · 2.4 X 20MM VAL SCREW, TI, STERILE · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867209800), received between May 15, 2025 and May 15, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Failure to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance1
- Positioning Failure1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 15, 2025Positioning Failure; Failure to Advance; Positioning Failure; Failure to Advance
ON 28TH APR 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT WHEN INSERTING AN AR-8724V-20S (LOW PROFILE VA LOCKING SCREW, 2.4 X 20 MM, STERILE), IT PASSED THROUGH THE PLATE (AR-8916VSL-03S VOLAR DISTAL RADIUS PLATE, TI, STANDARD, LEFT, 3 HOLE) WITHOUT BEING LOCKED AND WAS LEFT INSIDE THE PATIENT'S BONE. THIS OCCURRED DURING A DISTAL RADIUS FRACTURE CASE ON (B)(6) 2025. THE METHOD OF INSERTING THE SCREW WAS FIRSTLY TO AFFIX THE AR-8916-02 (AIMING GUIDE VDR PLATE STANDARD L) TO THE PLATE BEFORE INSERTING THE SCREW. THE SCREW IS NOT SWUNG WITH THE VARIABLE SCREW, AND THE ANGLE IS FIXED. THE DRIVER USED WAS AR-8916TL-011, AND THE SCREW WAS NOT INSERTED WITH EXCESSIVE FORCE. THE OTHER SCREWS WERE LOCKED WITHOUT ANY PROBLEMS. AT THE DOCTOR'S DISCRETION, THE PLATE AND BURIED AR-8724V-20S SCREW WERE LEFT IN THE PATIENT'S BODY AND THE SURGERY WAS COMPLETED. ONLY THE DRIVER AND AIMING GUIDE WILL BE RETURNED TO ARTHREX. THE PROCEDURE WAS NOT CONSIDERED TO BE COMPLETED SUCCESSFULLY AS THE AR-8724V-20S SCREW IN QUESTION WAS IN AN INCOMPLETE FIXED STATE WHILE THERE WERE NO PROBLEMS WITH THE SCREW FIXATION IN OTHER AREAS.
Patient problems: Foreign Body In Patient
MAUDE report 22025298 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page