Skip to content

Arthrex® AR-8963ML-10Recalls & reported problems

Arthrex, Inc. · MEDIAL DISTAL TIBIA PLATE, SS, LEFT, 10H · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867233553), received between Feb 13, 2025 and Feb 13, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025Use of Device Problem; Use of Device Problem

    ON (B)(6) 2025, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT A PLATE AND MULTIPLE SCREWS WERE IMPLANTED IN A PATIENT. TWO ARTHREX SCREWS BROKE DURING IMPLANTATION. ONE SCREW WAS SUCCESSFULLY REMOVED, AND THE OTHER SCREW WAS PARTIALLY REMOVED. THE PARTIAL SCREW WAS LEFT IN PLACE. THE PROCEDURE WAS PERFORMED UNDER FLUOROSCOPY, AND THE PARTIAL SCREW WAS OBSERVED BY THE OPERATING ROOM STAFF AND THE SURGEON. THIS WAS DISCOVERED DURING AN ANKLE FRACTURE PROCEDURE ON (B)(6) 2025. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION WAS PROVIDED ON 2/5/2025: THE FACILITY REPRESENTATIVE PROVIDED PART AR-8835CL-32 KREULOCK COMPRESSION SCREW AS THE SCREW THAT WAS REMOVED ENTIRELY. THE AR-8835CL-36 KREULOCK COMPRESSION SCREW WAS PARTIALLY REMOVED. THE AR-8963ML-10 MEDIAL DISTAL TIBIA PLATE REMAINED IN THE PATIENT AND SUFFERED NO DAMAGE OR MALFUNCTION. THE CASE WAS COMPLETED USING THE SAME PLATE WITH NEW SCREWS. THE CASE WAS NOT DELAYED, AND ADDITIONAL ANESTHESIA WAS NOT NEEDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21375721 ↗