Arthrex, Inc. · DBL LOADED 5.5MM BC SWVLK · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00888867251069), received between Apr 22, 2020 and Oct 27, 2025 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Entrapment of Device1
  • Material Split, Cut or Torn1
  • Physical Resistance/Sticking1

Patient problems

  • Insufficient Information4
  • No Clinical Signs, Symptoms or Conditions4
  • Cellulitis1
  • Device Embedded In Tissue or Plaque1
  • Edema1
  • Erythema1
  • Failure of Implant1
  • Swelling1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2025Break; Break

    ADDITIONAL INFORMATION: AN AR 2323BCT-2 WAS USED TO COMPLETE THE CASE. THE SURGERY EXPERIENCED A MINOR DELAY OF APPROXIMATELY 5 MINUTES. HOWEVER, NO ADDITIONAL ANESTHESIA WAS ADMINISTERED TO THE PATIENT DUE TO THIS DELAY. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED DURING OR FOLLOWING THE PROCEDURE IN RELATION TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23390110 ↗
  • MalfunctionFeb 19, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-2323BCT-2 BIOCOMPOSITE SWIVELOCK C WAS ANCHORED BUT HAD AN ISSUE WITH BEING UNABLE TO REMOVE THE INSERTER AND THE KNOB FROM THE SUTURES AT THE END. THEY WERE ABLE TO CUT THE SUTURES, AND THE REP CHECKED THE INSERTER AFTER THE CASE AND WAS ABLE TO REMOVE THE SUTURES WITH HEMOSTATS. THE REP SAID THE SUTURES WERE INTACT WITHOUT ANY KNOTS. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 02/07/2025: THIS OCCURRED IN A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2025. THERE WAS NO CASE DELAY AND NO ADDED ANESTHESIA ADMINISTERED TO THE PATIENT. NO ADVERSE EFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21412034 ↗
  • MalfunctionMay 13, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR, WHEN THE SURGEON IMPLANTED THE ANCHOR, AND THE REPAIR WAS NEAR COMPLETION, THE SURGEON THEN USED THE EYELET FIBERWIRE FROM THE AR-2323BCT-2, (LOT:11938015), AND AS THE SURGEON PULLED ON THE FIBERWIRE, THE FIBERWIRE SNAPPED IN HALF AND THE ANCHOR PULLED OUT. THIS LED TO A COMPLETE FAILURE OF THE REPAIR AND THE SURGEON HAS TO ESSENTIALLY START OVER.

    Patient problems: Insufficient Information

    MAUDE report 19306095 ↗
  • MalfunctionApr 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 12/22/2021, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT AN AR-2323BCT-2 SWIVELOCK FAILED WHEN SURGEON REMOVED INSERTER HANDLE AND ALL THE SUTURES CAME OUT OF THE PEEK ANCHOR. THIS WAS DISCOVERED DURING A PROCEDURE ON THE (B)(6) 2021. AFTER ADDITIONAL INFORMATION RECEIVED ON 12/22/2021, FACILITY REPRESENTATIVE HAS CONFIRMED THAT THIS OCCURRED DURING A RCR AND ALL THE SUTURES THAT CAME OFF THE ANCHOR AND THE IMPLANTED ANCHOR, WERE REMOVED FROM INSIDE THE PATIENT. SURGEON USED THE SAME BONE SOCKET PREVIOUSLY CREATED TO INSERT A NEW SWIVELOCK ANCHOR FROM THE SAME LOT NUMBER AS THE FAILED ONE. CASE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 19203520 ↗
  • MalfunctionApr 22, 2024Defective Device; Defective Device

    ON 7/11/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2323BCT-2 SWIVELOCK HANDLE WAS SLIPPING INTO THE DEVICE SHAFT AS THE SURGEON WAS ATTEMPTING TO MANUALLY SCREW IN THE ANCHOR. BECAUSE OF THIS, THE ANCHOR COULD NOT BE IMPLANTED. CASE WAS COMPLETED WITH ANOTHER AR-2323BCT-2 SWIVELOCK AND NO FURTHER ISSUES HAS BEEN REPORTED. THIS WAS DISCOVERED DURING AN ACLR PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 19154033 ↗
  • MalfunctionJun 9, 2023Break; Break

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2323BCT-2 BIOCOMPOSITE SWIVELOCK ANCHOR BROKE DURING INSERTION. ALL THE BROKEN FRAGMENTS WERE RETRIEVED, AND THE CASE WAS COMPLETED USING ANOTHER AR-2323BCT-2 BIOCOMPOSITE SWIVELOCK. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 5/17/2023: THIS WAS DISCOVERED DURING AN RCR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17097446 ↗
  • MalfunctionOct 26, 2021Break

    ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT WHILE THE SURGEON WAS TAPPING THE AR-2323BCT-2 ANCHOR DOWN TO BONE THE EYELET SUTURE BROKE OFF THE EYELET AND THE EYELET CAME FREE. THE SURGEON REMOVED THE EYELET WITH A GRASPER. THIS WAS DISCOVERED DURING USE IN A PROCEDURE ON (B)(6) 2021. THE CASE WAS COMPLETED BY USING A AR-2323BCC. NO PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 12696544 ↗
  • MalfunctionJul 6, 2021Break; Entrapment of Device; Detachment of Device or Device Component

    IT WAS REPORTED DURING A ROTATOR CUFF REPAIR PROCEDURE, AFTER INSERTING THE AR-2323BCT-2 SWIVELOCK. ADDITIONAL INFORMATION 6/15/2021. IT WAS REPORTED DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON STARTED TO INSERT THE AR-2323BCT-2 SWIVELOCK MEDIALLY. HALFWAY THROUGH INSERTION, THE SURGEON DECIDED HE NEEDED TO TAKE IT OUT AND RECHECK THE ANGLE. AS HE WAS PULLING THE STILL INTACT ANCHOR OUT, THE EYELET TORE AWAY FROM THE DRIVER AND EYELET SUTURES AND WAS LEFT IN THE BONE. ADDITIONAL INFORMATION 6/17/2021. IT WAS REPORTED DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON STARTED TO INSERT THE AR-2323BCT-2 SWIVELOCK MEDIALLY. HALFWAY THROUGH INSERTION, THE SURGEON DECIDED HE NEEDED TO TAKE IT OUT AND RECHECK THE ANGLE. AS THE SURGEON REMOVED THE DEVICE THE AR0-2323BCT-2 BROKE. THE REMAINING PARTS OF THE ANCHOR AND DRIVER WERE REMOVED WHEN HE PULLED IT OUT AND THE BROKEN EYELET REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; Failure of Implant

    MAUDE report 12124240 ↗
  • MalfunctionJun 14, 2021Material Split, Cut or Torn

    IT WAS REPORTED AFTER PUTTING IN THE AR-2323BCT-2 SWIVELOCK LATERALLY, THE SUTURES TORE WHEN THE SURGEON PULLED THE TAB OFF. THE REP STATED ALL BROKEN SUTURE FRAGMENTS WERE FULLY RETRIEVED, AND THE ANCHOR BODY REMAINED WHOLE AND SECURE WITHIN THE INTENDED IMPLANTATION SITE. THE SUTURES TORE AFTER THE LATERAL ROW ANCHOR WAS IMPLANTED AND DID NOT AFFECT THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11990896 ↗
  • InjuryApr 22, 2020Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SHOULDER REPAIR SURGERY ON (B)(6) 2020 IN WHICH AR-2324BCT-2 & AR-2323BCT-2 WERE IMPLANTED. THE PATIENT HAS BEEN EXPERIENCING POST-OP SYMPTOMS OF REDNESS, SWELLING AND DRAINING. THE PATIENT HAS BEEN REFERRED TO AN IMMUNOLOGIST. THE IMMUNOLOGIST IS REQUESTING SAMPLE PRODUCT OF AR-2324BCT-2 & AR-2323BCT-2. THE IMMUNOLOGY APPOINTMENT IS SCHEDULED FOR (B)(6) 2020. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON 4/01/2020: THE ORIGINAL PROCEDURE PERFORMED WAS A LEFT ROTATOR CUFF REPAIR (TWO ANCHORS SUTURE BRIDGE) SUBACROMIAL DECOMPRESSION AND SUB PECTORALIS BICEPS TENODESIS. THE FACILITY REPORTED AR-2324BCT-2 (LOT: 10431237) & AR-2323BCT-2 (LOT: 10268328) WERE THE ONLY ARTRHEX PRODUCTS IMPLANTED DURING THE PROCEDURE. THE PATIENT'S SYMPTOMS FIRST STARTED ON (B)(6) 2020. POSTOPERATIVELY, THE PATIENT DEVELOPED CELLULITIS OVERLYING THEIR LEFT SHOULDER AND UPPER ARM. THE PATIENT WAS INITIALLY PLACED ON KEFLEX WHICH DID NOT SEEM TO IMPROVE THEIR SYMPTOMS AND WAS THEN TRANSITIONED TO CLINDAMYCIN AGAIN WITH NO SIGNIFICANT CHANGES. THE PATIENT WAS THEN REFERRED TO INFECTIOUS DISEASE WHO STARTED THEM ON IV VANCOMYCIN. AGAIN

    Patient problems: Cellulitis; Edema; Erythema; Swelling

    MAUDE report 9985414 ↗