Skip to content

Arthrex® AR-6485Recalls & reported problems

Arthrex, Inc. · SYNERGY CW4 ARTHROSCOPY PUMP · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex Synergy CW4 arthroscopy pump is a device that feeds fluid into a joint during arthroscopic surgery. Many of the 113 reports describe the pump not regulating pressure correctly, pressure building up on its own, or the unit losing power or shutting down. Some reports describe fluid leaking into the surrounding tissue, causing swelling. The reports include 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
109
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 109 adverse event reports that list this exact device (DI 00888867251182), received between May 18, 2021 and Mar 13, 2026 — classified as 4 injuries, 105 malfunctions.

The most frequently reported specific device problem is “Pressure Problem” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 21 reportsJan 2024: 10 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 7 reportsMay 2024: 9 reportsJun 2024: 13 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pressure Problem38
  • Increase in Pressure12
  • Defective Device9
  • Intermittent Loss of Power7
  • Excess Flow or Over-Infusion4
  • Key or Button Unresponsive/not Working4
  • Unexpected Shutdown4
  • No Flow3

Patient problems

  • No Clinical Signs, Symptoms or Conditions85
  • Insufficient Information14
  • Extravasation8
  • Swelling/ Edema5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Inflammation1

Recent adverse event reports

The 25 most recent of 109 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Pressure Problem

    IT WAS REPORTED THAT DURING A SURGERY THE PUMP SUDDENLY BUILT-UP PRESSURE CONTINUOUSLY WITHOUT STOPPING. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION WAS RECEIVED. UPDATE (B)(6) IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING AN ARTHROSCOPY SURGERY. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590246 ↗
  • MalfunctionFeb 11, 2026Pressure Problem

    IT WAS REPORTED BY A MEDICAL TECHNICIAN THAT DURING INSPECTION IT WAS FOUND THAT THE PAT REGULATES PRESSURE EVEN WHEN THERE IS RESISTANCE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THERE WAS NO CASE INVOLVEMENT. NO FURTHER INFORMATION RECEIVED. UPDATE ON 4-FEB-2025: FURTHER INFORMATION WAS RECEIVED ON 26-JAN-2026 THAT THE FAILURE OCCURRED DURING AN ARTHROSCOPIC KNEE SURGERY. THE DEVICE DID NOT REGULATE THE PRESSURE CORRECTLY. THE SURGERY ON (B)(6) 2025 WAS COMPLETED SUCCESSFULLY WITH THE DEVICE. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24317110 ↗
  • MalfunctionJan 8, 2026Excess Flow or Over-Infusion

    ON 18-DEC-2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-6485 SYNERGY CW4, ARTHROSCOPY PUMP DELIVERED AN EXCESSIVE AMOUNT OF FLUID INTO THE SHOULDER DURING A, ROTATOR CUFF REPAIR PROCEDURE. THE PUMP WAS SET TO 50, AND AT THE START OF THE CASE, SIGNIFICANT FLUID ACCUMULATION RESULTED IN EXTRAVASATION. THE SURGICAL TEAM REMOVED THE EXTRAVASATION TO THE BEST OF THEIR ABILITY, AND THE PROCEDURE CONTINUED USING A DIFFERENT PUMP. IT IS UNKNOWN WHETHER THE PATIENT WAS KEPT OVERNIGHT.

    Patient problems: Extravasation; Insufficient Information

    MAUDE report 24003435 ↗
  • MalfunctionJun 20, 2025Pressure Problem; Pressure Problem

    ON 06/02/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP WAS PRODUCING A PRESSURE FAULT ERROR CODE. THE ISSUE IS BELIEVED TO BE ISOLATED TO THE PUMP, AS BOTH TUBE SETS ARE WORKING ON DIFFERENT CONSOLES WITH NO ERROR CODE.

    Patient problems: Insufficient Information

    MAUDE report 22298803 ↗
  • MalfunctionApr 21, 2025Pressure Problem; Pressure Problem

    ADDITIONAL INFORMATION WAS RECEIVED ON 04/16/2025: THIS OCCURRED ON (B)(6) 2025, AND THE PROCEDURE NAME IS UNKNOWN. THERE WAS NO DELAY IN THE CASE, NOR WAS ADDITIONAL ANESTHESIA ADMINISTERED. THE ARTHREX TUBING USED WITH THE PUMP IS UNKNOWN. THE PUMP SETTINGS AT THE TIME OF THE EVENT WERE "40." NO CLAMP OR PRESSURE TEST WAS CONDUCTED BEFOREHAND. THERE WAS A BEEPED ERROR MESSAGE, BUT THE DISPLAY DID NOT MENTION ANY ALARM MESSAGE. NO PHOTOS WERE TAKEN, AND NO SALES REP WAS PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876739 ↗
  • MalfunctionApr 18, 2025Pressure Problem; Pressure Problem

    ON 04/02/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(6) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP STOPS RANDOMLY WITH PRESSURE FAULT ERROR. THE PRESSURE FAULT OCCURRED DURING THE CASE. THERE WAS NO CASE INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 21865435 ↗
  • MalfunctionFeb 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ON 01/30/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP WAS EXPERIENCING INACCURATE PRESSURE READINGS.

    Patient problems: Insufficient Information

    MAUDE report 21393789 ↗
  • MalfunctionFeb 11, 2025Pressure Problem; Pressure Problem

    ON 01/21/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMPS PRESSURE WILL JUMP INTERMITTENTLY AND THE HAND CONTROL WILL NOT CONNECT TO THE PUMP. THIS OCCURRED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356853 ↗
  • MalfunctionFeb 11, 2025Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY A "CRITICAL ERROR" HAS OCCURRED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21353939 ↗
  • MalfunctionJan 16, 2025Increase in Pressure; Increase in Pressure

    ON 12/20/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP WAS OVER PRESSURE RISING IN JOINT. NO REPORTED PATIENT HARM. NO FURTHER INFORMATION PROVIDED. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21170235 ↗
  • MalfunctionJan 3, 2025Increase in Pressure; Increase in Pressure

    ON 12/06/2024, A SALES REPRESENTATIVE VIA (B)(4) REPORTED THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP DISPLAYED AN OVERPRESSURE ERROR WHEN IN USE. THE UNIT WAS CONNECTED IN A CASE, AND THE PRESSURE ERROR CODE WOULD NOT LET THE UNIT WORK AS INTENDED. THE UNIT WOULD AUTOMATICALLY STOP WHILE THE PATIENT WAS IN THE ROOM. THE UNIT WAS SWITCHED TO ANOTHER, AND THE CASE WAS COMPLETED SUCCESSFULLY WITHOUT ADVERSE EVENTS OR HARM TO THE PATIENT. THIS WAS DISCOVERED DURING A PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21067183 ↗
  • MalfunctionOct 17, 2024Output Problem; Output Problem

    ON 09/19/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP WAS NOT PUMPING AS IT WAS SET TO PUMP, THEY PUT IN 700 ML FL THROUGH IT, BUT THE MACHINE SAID 1200 ML. THIS OCCURRED DURING THE CASE, AND THE PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20472608 ↗
  • MalfunctionOct 4, 2024Pumping Problem; Pumping Problem

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA *(B)(4) THAT AN AR-6485 PUMP WAS PUMPING 500 ML LESS. NOT AWARE OF A CASE, NO PATIENT WAS AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20368589 ↗
  • InjurySep 23, 2024Pressure Problem; Pressure Problem

    ON 11TH SEPTEMBER 2024, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2024 AT AROUND 3.30PM, SWELLING IN THE CHEST CAVITY (UP TILL THE NECK) WAS DETECTED BY THE ANESTHETIST SHORTLY AFTER THE INSERTION OF THE FIRST 2 ANCHORS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE ARTHREX EMPLOYEE HAD CHECKED THAT ALL THE DEVICES (I.E., SCOPE, PUMP, SHAVER AND RF) WERE CONNECTED PROPERLY AND NOTICED THAT THE PUMP WAS ALREADY AT 65MMHG BUT WAS UNSURE WHO HAD SET THE PRESSURE TO 65MMHG. SHE OBSERVED THROUGH THE SCOPE AND FOUND OUT THAT THE VISIBILITY WAS POOR AND THERE WERE MAJOR IMPEDIMENTS TO LOCATE THE CUFF TEAR. AS A RESULT, THE SURGEON THEN INSTRUCTED TO ACTIVATE LAVAGE ON MULTIPLE OCCASIONS AND WAS ABLE TO LOCATE THE CUFF TEAR. HE THEN PREPARED THE FOOTPRINT AND INSERTED THE FIRST TWO ANCHORS ON THE MEDIAL ROW. THIS WAS WHEN THE SWELLING WAS DETECTED BY THE ANESTHETIST WHO ALERTED THE SURGEON AND QUESTIONED HIM HOW MUCH LONGER HE REQUIRED TO COMPLETE THE SURGERY. THE SURGEON RESPONDED THAT HE NEEDED ANOTHER 30 MINUTES TO FINISH UP THE SURGERY AND DECIDED TO PRESS ON. THE ARTHREX EMPLOYEE THEN INFORMED THE SURGEON THAT SHE WOULD DECREA

    Patient problems: Extravasation

    MAUDE report 20283741 ↗
  • MalfunctionSep 9, 2024Pressure Problem; Pressure Problem

    IT WAD REPORTED THAT DURING A ROTATOR CUFF SURGERY THE TUBING WORKED GREAT BEGINNING OF THE SURGERY, BUT AFTER A WHILE THE PUMP IN COMBINATION WITH THE TUBING STARTED TO PUMP 1000ML/PER MINUTE. DURING SURGERY THE TUBING WAS CHANGED AND IT WORKED GREAT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20175071 ↗
  • MalfunctionSep 5, 2024Increase in Pressure; Increase in Pressure

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THE PUMP OR SUCTION REGULATOR IS NOT WORKING. THE PUMP IS SUCKING AT A TOO HIGH A SUCTION PRESSURE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-6475). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20145929 ↗
  • MalfunctionJul 5, 2024Intermittent Loss of Power; Intermittent Loss of Power

    ON 06/10/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06590763 THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP WOULD STOP AND FLOW IN MED CASE. THIS OCCURRED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19680354 ↗
  • MalfunctionJun 21, 2024Defective Device; Infusion or Flow Problem; Defective Device; Infusion or Flow Problem

    ON 05/28/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMPS INFLOW SIDE WAS NOT WORKING. THIS OCCURRED DURING A CASE WHERE THEY SWITCHED TO GRAVITY TUBING.

    Patient problems: Insufficient Information

    MAUDE report 19589926 ↗
  • MalfunctionJun 21, 2024Device Alarm System; Device Alarm System

    ON 05/27/2024 IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT AN AR-6485 SYNERGY CW4 ARTHROSCOPY PUMP INDICATES THAT THE DOOR WAS OPEN. THIS OCCURRED DURING AN ANKLE INSTABILITY WHERE ANOTHER CONSOLE WAS USED TO PROCEED.

    Patient problems: Insufficient Information

    MAUDE report 19587554 ↗
  • MalfunctionJun 20, 2024Device Sensing Problem; Device Sensing Problem

    ON 04/10/2023 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6485 ARTHROSCOPY PUMPS DOOR IS CLOSED, HOWEVER, IS STILL READING AS OPENED AND NOT ALLOWING THE PUMP TO RUN. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19573850 ↗
  • MalfunctionJun 19, 2024Intermittent Loss of Power; Intermittent Loss of Power

    ON (B)(6) 2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6485 PUMP, WAS WORKING INTERMITTENTLY. THIS WAS DISCOVERED DURING USE IN A PROCEDURE. NO PATIENT AFFECT REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19568165 ↗
  • MalfunctionJun 14, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT AFTER THE INITIAL EVENT ( REPORTED IN (B)(4)) THE CUSTOMER CONTINUED TO USE THE DEFECTIVE PUMP. THE INITIAL REPORTED EVENT REOCCURRED AND DURING AN ARTHROSCOPIC SURGERY THE TUBING LATCH OPENED AND THE ERROR MESSAGE "DOOR OPEN" APPEARED AND THE FLOW STOPPED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE CUSTOMER CLOSED THE DOOR AND RESTARTED THE PUMP TO FINISHED THE SURGERY SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19537497 ↗
  • MalfunctionJun 13, 2024Unexpected Shutdown; Unexpected Shutdown

    ON 13-FEB-2023, IT WAS REPORTED BY A SALES REP VIA EMAIL THAT AN AR-6485, SYNERGY CW4 ARTHROSCOPY PUMPSYNERGY CW4 ARTHROSCOPY PUMP CHANGED FROM RUNNING TO STOPPED AND NEEDED TO BE RE-STARTED. THIS WAS DISCOVERED DURING USE AND A CASE WAS INVOLVED. TO COMPLETE THE PROCEDURE, AN EXISTING PUMP WAS USED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19527202 ↗
  • MalfunctionJun 12, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT AFTER THE INITIAL EVENT ( REPORTED IN (B)(4)) THE CUSTOMER CONTINUED TO USE THE DEFECTIVE PUMP. THE INITIAL REPORTED EVENT REOCCURRED AND DURING AN ARTHROSCOPIC SURGERY THE TUBING LATCH OPENED AND THE ERROR MESSAGE "DOOR OPEN" APPEARED AND THE FLOW STOPPED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE CUSTOMER CLOSED THE DOOR AND RESTARTED THE PUMP TO FINISHED THE SURGERY SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19524413 ↗
  • MalfunctionJun 12, 2024Restricted Flow rate; Restricted Flow rate

    ON 22-NOV-2022, IT WAS REPORTED BY A SALES REP VIA EMAIL THAT WHILST USING AN AR-6485, SYNERGY CW4 ARTHROSCOPY PUMP IT WAS NOTED THAT THE FLOW RATE WAS READING INCORRECTLY. THE SHEATH AND SCOPE WERE TAKEN OUT AND SET TO 80MMHG WHICH SHOULD HAVE HAD A HIGH FLOW BUT INSTEAD IT WAS JUST A SLOW POUR OF FLUID. ALL FITTINGS ETC WERE CHECKED AND NO LEAKS WERE OBSERVED. POST PROCEDURE THE SALES REP TOOK THE PUMP BACK TO WA OFFICE AND RAN THE SELF-CHECK TEST AND PUMP SHOWED FLUID FAILURE AFTER RUNNING FOR AROUND 2 MINS. THIS WAS DISCOVERED DURING USE AND A CASE WAS INVOLVED. TO COMPLETE THE PROCEDURE, THE SAME PUMP WAS USED, AND IT WAS NOTED AFTER THE PROCEDURE THAT THE FLUID FLOW WAS A LOT SLOWER THAN SHOWING ON THE SCREEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522065 ↗

FDA clearance and classification

The FDA regulates product code HRX (arthroscope) under 21 CFR 888.1100 as a Class II device. 21 CFR 888.1100 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Arthrex® AR-6485

Has the Arthrex® AR-6485 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867251182), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-6485?

The FDA's MAUDE database lists 109 adverse event reports for this device, received between May 18, 2021 and Mar 13, 2026. In those reports, the most common specific device problems are “Pressure Problem” (38), “Increase in Pressure” (12) and “Defective Device” (9); the most common patient problems are “Extravasation” (8), “Swelling/ Edema” (5) and “Inflammation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-6485?

None of the reports are classified as deaths, 4 as injuries and 105 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Arthrex® AR-6485, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-6485). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Arthroscope devices

All Arthroscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page