Skip to content

Arthrex® AR-2324PSLC-2Recalls & reported problems

Arthrex, Inc. · DBL LOADED 4.75MM PEEK SWVLK · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867251243), received between Sep 19, 2022 and Jun 19, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Split, Cut or Torn2
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON INSERTED AND SEATED THE AR-2324PSLC-2 DOUBLE LOADED PEEK SWIVELOCK, (LOT: 12900335) AS THE SURGEON WAS TIGHTENING DOWN ON THE WHITE SUTURE, HE NOTICE THE SUTURE WAS CUT FROM THE EYELET. THE BROKEN SUTURE WAS REMOVED FROM THE PATIENT. THE BLUE SUTURE WAS USE TO SECURE THE AR-2324PSLC-2. THE SURGEON USED A SECOND AR-2324PSLC-2 (LOT: 13262585) AND THE SAME ISSUE OCCURRED. THE CASE WAS COMPLETED WITHOUT PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19570117 ↗
  • MalfunctionApr 25, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON INSERTED AND SEATED THE AR-2324PSLC-2 DOUBLE LOADED PEEK SWIVELOCK, (LOT: 12900335) AS THE SURGEON WAS TIGHTENING DOWN ON THE WHITE SUTURE, HE NOTICE THE SUTURE WAS CUT FROM THE EYELET. THE BROKEN SUTURE WAS REMOVED FROM THE PATIENT. THE BLUE SUTURE WAS USE TO SECURE THE AR-2324PSLC-2. THE SURGEON USED A SECOND AR-2324PSLC-2 (LOT: 13262585) AND THE SAME ISSUE OCCURRED. THE CASE WAS COMPLETED WITHOUT PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181625 ↗
  • MalfunctionApr 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 09/21/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT QTY. 3 OF AN AR-2324PSLC-2 DOUBLE LOADED SWIVELOCKS BROKE DURING LATERAL ROW IMPLANTATION. FIRST THE EYELET POPPED OFF WHILE THE SURGEON WAS PULLING THE ANCHOR OVER TO THE HOLE. THEN UNWOUND THE EYELET SUTURE AND PULLED TIGHT TO ¿RE-SEAT¿ THE EYELET, CLAMPED THE SUTURE SO THE EYELET WOULDN'T SLIP OFF AGAIN. AS THE SURGEON STARTED TO PULL THE ANCHOR OVER TO THE HOLE A 2ND TIME, THE EYELET SUTURE SEVERED AT THE EYELET/DRIVER ATTACHMENT. THIS OCCURRED 3 TIMES IN A ROW IN THE EXACT SAME SEQUENCE. THE CASE WAS COMPLETED BY SWITCHING TO A STANDARD SWIVELOCK WHICH IMPLANTED PERFECTLY WITH NO ADDITIONAL ISSUE. NO PATIENT HARM ADDITIONAL INFORMATION PROVIDED 09/21/21: THE PROCEDURE BEING PERFORMED WAS A ROTATOR CUFF REPAIR. THE CASE WAS COMPLETED BY USING AN AR-2324PSLC. IMAGES OF THE DEVICE ARE ATTACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19097146 ↗
  • MalfunctionFeb 26, 2024Break; Break

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2324PSLC-2 SWIVELLOCK ANCHOR BROKE UPON INSERTION. THE PATIENT WAS NOT HARMED AND THE PRODUCT WAS DISCARDED. THE SAME ANCHOR WITH A DIFFERENT LOT WAS USED TO COMPLETE THE REPAIR. CASE WAS DELAYED 5 MINUTES BUT NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18779537 ↗
  • MalfunctionSep 19, 2022Break

    IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2324PSLC-2 SWIVELLOCK SUTURES BROKE-OFF AT THE PEEK TIP. THIS WAS DISCOVERED DURING A CASE AND BROKE INSIDE THE PATIENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15445770 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page