Skip to content

Arthrex® AR-9943BL-04Recalls & reported problems

Arthrex, Inc. · Locking Distal Fibula Plate, Ti ,Left,4H · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867255609), received between Aug 17, 2023 and Aug 13, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1
  • Positioning Failure1

Patient problems

  • Hypersensitivity/Allergic reaction1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025Positioning Failure; Positioning Failure

    ON 07/28/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-8935CL-14 KREULOCK COMPRESSION SCREW (3.5 MM X 14 MM) FAILED TO ENGAGE WITH THE LOCKING MECHANISM OF AN AR-9943BL-04 LOCKING DISTAL FIBULA PLATE DURING AN ANKLE FRACTURE PROCEDURE ON (B)(6) 2025. DESPITE PROPER DRILLING USING THE DESIGNATED DRILL TOWER, BOTH SCREWS SPUN FREELY WITHIN THE HOLE AND DID NOT LOCK INTO THE PLATE. A 4.0 MM CORTICAL SCREW WAS ULTIMATELY USED TO FILL THE HOLE; HOWEVER, LOCKING WAS NOT ACHIEVED, AND THEY SUSPECT THAT THE ISSUE IS WITH THE LOCKING THREADS ON THE PLATE. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 08/01/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22793773 ↗
  • InjuryAug 17, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGERY ON (B)(6) 2018 FOR AN OPEN REDUCTION INTERNAL FIXATION OF A LEFT TRIMALLEOLAR ANKLE FRACTURE WITH FIXATION OF THE DISTAL FIBULA FRACTURE, POSTERIOR MALLEOLUS FRACTURE, AND SYNDESMOTIC FIXATION. DURING THE (B)(6) 2018 SURGERY, THE FOLLOWING PRODUCTS WERE USED AND/OR IMPLANTED: AR-8740-52PTS QCKFIX SCRW,TI,CANN ST,CANC.,4.0X 52MM, LOT/SERIAL (B)(6). AR-8925T K-LESS T-ROPE W/DRV, SYN REPR, TI, LOT/SERIAL (B)(6). AR-8933-20 LO-PRO SCRW,TI,3.0MMX 20MMCORT, LOT/SERIAL (B)(6). AR-8933V-12 3.0 X 12MM VAL SCREW, TI, LOT/SERIAL (B)(6). AR-8933V-14 3.0 X 14MM VAL SCREW, TI, LOT/SERIAL (B)(6). AR-8933V-16 3.0 X 16MM VAL SCREW, TI, LOT/SERIAL (B)(6). AR-8933V-18 3.0 X 18MM VAL SCREW, TI, LOT/SERIAL (B)(6). AR-8933V-20 3.0 X 20MM VAL SCREW, TI, LOT/SERIAL (B)(6). AR-8935-12 LO-PRO SCRW,TI,3.5MMX 12MM, LOT/SERIAL (B)(6). AR-8935-14 LO-PRO SCRW,TI,3.5MMX 14MM, LOT/SERIAL (B)(6). AR-8935-14 LO-PRO SCRW,TI,3.5MMX 14MM, LOT/SERIAL (B)(6). AR-8935-26 LO-PRO SCRW,TI,3.5MMX 26MM, LOT/SERIAL (B)(6). AR-8943-01 GUIDE WIRE, TROCAR TIP, 1.35 MM, LOT/SERIAL (B)(6). AR-8943-02 DRILL BIT, CANNULATED, 2.6MM, LOT/SERIAL (B)(6). AR-8943-16 2.0MM DRILL BIT, CALIBR

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 17563229 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page