Skip to content

Arthrex® AR-8750-40HRecalls & reported problems

Arthrex, Inc. · COMPR FT SCRW, 5.0 LG, 40MM LGTH · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867262867), received between Mar 9, 2020 and Mar 9, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 9, 2020Break

    IT WAS REPORTED THAT DURING A SUBTALAR FUSION PROCEDURE, THE SURGEON BROKE THE THREADED TIP OF AR-8750KT INSIDE THE PATIENT¿S TALUS. NO ATTEMPT WAS MADE TO RETRIEVE. THE SURGEON WAS ATTEMPTING TO CHANGE THE POSITION OF THE K WIRE AND WHILST DRILLING OUT ON FORWARD THE TIP WAS ADVANCING WHILST THE SHAFT OF THE K WIRE WAS BEING WITHDRAWN FROM THE PATIENT, HENCE A BREAKAGE IS POSSIBLE. IN THE SAME CASE THE SURGEON INSERTED A NEW AR-8750KT SUCCESSFULLY. HOWEVER, BROKE AN AR-8750-40H INSIDE THE PATIENT. THE SURGEON IS VERY FAMILIAR WITH THIS SYSTEM AND ALL ASPECTS OF THE INSERTION APPEARED TO BE COMPLETED CORRECTLY. SURGEON CONTINUED TO INSERT SCREW EVEN AFTER CONSIDERABLE RESISTANCE WAS ENCOUNTERED; HE DIDN¿T STOP TO REMOVE THE SCREW AND RE-DRILL OR ALTER APPROACH. THE SURGEON DRILLED WITH THE CORRECT 1.6 MM GUIDE WIRE (AR-8750KT), MEASURED WITH DEPTH DEVICE (AR-8750-01), OVER DRILLED WITH 3.2 MM DRILL BIT (AR-8750-09), HAND POWERED PROFILE DRILL (AR-8750-05), INSERTED SCREW BY HAND SCREW DRIVER (AR-8750-03) AND THEN FINISHED WITH SOLID DRIVER (AR-8750-06). THE SURGEON DEMANDED THE SOLID DRILL AFTER TWISTING/ WARPING THE END OF THE CANNULATED DRIVER AFTER EXCESSIVE TORSIONAL FORCE DURI

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9805173 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates