Skip to content

Arthrex® AR-8750-46HRecalls & reported problems

Arthrex, Inc. · COMPR FT SCRW, 5.0 LG, 46MM LGTH · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867262942), received between Sep 26, 2022 and Nov 5, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Fitting Problem1
  • Material Deformation1
  • Material Twisted/Bent1
  • Physical Resistance/Sticking1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 10/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8750-46H 5.0 LARGE COMPRESSION FT SCREW WELDED TO THE K-WIRE AND WAS REMOVED. THE SCREW AND K-WIRE COULD NOT BE TAKEN APART. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20611044 ↗
  • MalfunctionMay 23, 2024Fitting Problem; Fitting Problem

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8750-46H LARGE COMPRESSION FT SCREW DID NOT FIT INTO ANY T15 HEX DRIVERS. A NEW AR-8750-46H LARGE COMPRESSION FT SCREW WAS OPENED TO COMPLETE THE CASE SUCCESSFULLY WITHOUT FURTHER ISSUES. THIS WAS DISCOVERED DURING A CALCANEUS OSTEOTOMY PROCEDURE ON (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19382063 ↗
  • MalfunctionSep 26, 2022Break; Material Twisted/Bent

    ON 9/6/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-8750-46H, AN AR-8750-42H, AND AN AR-8750-38H HAD ISSUES. DURING A FULKERSON OSTEOTOMY PROCEDURE ON (B)(6)2022, ALL THREE COMPLAINT DEVICES WERE IMPLANTED, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH NO IMPACT TO THE PATIENT. ON (B)(6) 2022, THE X-RAY WAS SHOWING THAT 2 OF 3 SCREWS WERE BENDING. ON (B)(6) 2022, ON THE NEXT X-RAY, 2 OF THE 3 SCREWS WERE BROKEN INSIDE THE PATIENT POST-OP. ACCORDING TO THE X-RAY AND THE SALES REPRESENTATIVE, IT WAS ASSUMED THAT THE AR-8750-46H AND AR-8750-42H WERE BROKEN. THE BROKEN SCREWS WERE SECURELY AFFIXED INSIDE THE PATIENT, AND THE SURGEON WAS NOT PLANNING A REVISION SURGERY AND DEEMED IT SAFE TO LEAVE THE BROKEN SCREWS INSIDE THE PATIENT, BUT JUST WANTED TO HAVE THE COMPLAINT REPORTED. THE PATIENT IS FINE TO DATE. ON 09/14/2022, ANOTHER SALES REPRESENTATIVE CALLED AND PROVIDED VIA PHONE THE FOLLOWING INFORMATION: ON (B)(6) 2022, A REVISION SURGERY WAS PERFORMED BY THE SAME SURGEON TO REMOVE THE BROKEN AR-8750-46H AND AR-8750-42H FROM THE PATIENT SUCCESSFULLY. BOTH COMPLAINT DEVICES WERE AVAILABLE FOR RETURN.

    Patient problems: Foreign Body In Patient

    MAUDE report 15484138 ↗
  • MalfunctionSep 26, 2022Break; Material Deformation

    ON 9/6/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-8750-42H, AN AR-8750-46H, AND AN AR-8750-50H HAD ISSUES. DURING A FULKERSON OSTEOTOMY PROCEDURE ON (B)(6) 2022, ALL THREE COMPLAINT DEVICES WERE IMPLANTED, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH NO IMPACT TO THE PATIENT. IT WAS REPORTED THAT THE PATIENT HAD A POST-OP INFECTION THREE WEEKS AFTER SURGERY. ON (B)(6) 2022, THE X-RAY WAS SHOWING THAT 2 OF 3 SCREWS WERE BENDING. ON (B)(6) 2022, ON THE NEXT X-RAY, ONE OF THE SCREWS HAD BROKEN POST-OP INSIDE THE PATIENT. ON (B)(6) 2022, TWO OF THE SCREWS WERE BROKEN INSIDE THE PATIENT POST-OP. ACCORDING TO THE X-RAY AND THE SALES REPRESENTATIVE, IT WAS ASSUMED THAT THE AR-8750-50H AND AR-8750-46H WERE BROKEN. THE BROKEN SCREWS WERE SECURELY AFFIXED INSIDE THE PATIENT, AND THE SURGEON WAS NOT PLANNING A REVISION SURGERY AND DEEMED IT SAFE TO LEAVE THE BROKEN SCREWS INSIDE THE PATIENT, BUT JUST WANTED TO HAVE THE COMPLAINT REPORTED. THE PATIENT IS FINE TO DATE.

    Patient problems: Unspecified Infection

    MAUDE report 15484127 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates