Skip to content

Arthrex® AR-8737-61Recalls & reported problems

Arthrex, Inc. · Easy-out, Micro Compression FT · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867263802), received between Jan 20, 2020 and Jan 6, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1
  • Entrapment of Device1
  • Patient Device Interaction Problem1

Patient problems

  • Device Embedded In Tissue or Plaque2
  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2025Break; Break

    ON 12/10/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8737-61 SCREW REMOVER, EASY-OUT, 2.5 MICRO COMPRESSION FT HAD THE TIP BREAK (FROM THEIR HEADLESS COMPRESSION SCREW SET #3). THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21084157 ↗
  • InjuryDec 18, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ON NOVEMBER 3, 2023, IT WAS REPORTED BY THE SURGEON THAT HIS PATIENT MAY BE HAVING AN ALLERGIC REACTION TO THE DEVICES AND/OR INSTRUMENTS USED IN A PRIOR SURGERY. NO FURTHER INFORMATION WAS PROVIDED. ON (B)(6) 2023, MEDICAL RECORDS WERE PROVIDED CONFIRMING THAT THE SURGEON PERFORMED A BUNIONECTOMY, AND 1ST METATARSAL OSTEOTOMY OF THE LEFT FOOT WAS PERFORMED ON (B)(6) 2020. ARTHREX¿S SALES RECORDS FOR THE PATIENT¿S (B)(6) 2020 PROCEDURE INDICATE THE SCREWS IMPLANTED DURING THE SURGERY WERE ARTHREX¿S AR-8725-24H COMPR FT SCREW,2.5 MICRO,24MM LGTH (LOT 10546665), AR-8730-28H COMPR FT SCREW,3.5 MINI,28MM LGTH (LOT 10279485) AND AR-8730-36H COMPR FT SCRW,3.5 MINI, 36MM (LOT 10443285). FOLLOWING THE INITIAL SURGERY, THE RECORDS INDICATE THAT ON (B)(6) 2023, A SECOND SURGERY WAS PERFORMED BY THE SAME SURGEON TO REMOVE A PROMINENT SCREW IN THE 1ST LEFT METATARSAL. DURING THIS SURGERY, MULTIPLE ARTHREX INSTRUMENTS WERE USED TO ATTEMPT TO REMOVE THE SCREW HOWEVER THE PROCEDURE WAS ABANDONED AS THE SCREW WAS UNABLE TO BE REMOVED FROM THE BONE AFTER MULTIPLE ATTEMPTS WITH DIFFERENT REMOVAL INSTRUMENTS. ALL SCREWS REMAIN IMPLANTED. THOSE INSTRUMENTS HAVE BEEN IDENTIFIED AS ARTHREX¿S AR-8737-35

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 18352186 ↗
  • MalfunctionDec 13, 2021Break; Entrapment of Device

    ON 11/24/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT A AR-8737-37 SHAFT TIP BROKE OFF, AND A AR-8737-61 SCREW REMOVER TIP BROKE. THIS OCCURRED ON (B)(6) 2021 DURING A METACARPAL FRACTURE WHILE INSERTING THE 2.5MMX50MM INTO THE 4TH METACARPAL THE CANAL OF THE METACARPAL IT WAS VERY TIGHT. THE METACARPAL WAS PREPARED WITH THE CORRECT CORRESPONDING DRILL BIT FOR THE 2.5MM HEADLESS COMPRESSION FT SCREW. WHILE INSERTING THE SCREW THE CANNULATED SCREW DRIVER TIP (AR-8737-37) BROKE OFF AND WAS THEN REMOVED. THE SURGEON THEN PROCEEDED TO USE THE SOLID SCREW DRIVER IN THE SET TO CONTINUE INSERTING THE SCREW. THE SCREW BECAME TIGHT IN THE SHAFT OF THE 4TH METACARPAL AND OF THE APPROXIMATELY 4MM OF THE SCREW BROKE. THE HEADLESS, BROKEN PORTION OF THE SCREW WAS REMOVED. THE SURGEON CONTINUE USING THE CORRESPONDING EASY OUT (AR-8737-61) TO TRY AND REMOVE THE PORTION OF THE SCREW THAT WAS STILL IN THE 4TH METACARPAL, AND THE EASY OUT TIP BROKE UPON TRYING TO REMOVE THE SCREW AND THE SMALL PORTION OF THE TIP WAS LEFT INSIDE OF THE CANNULATED PORTION OF THE SCREW. THE SCREW WAS LEFT IN THE 4TH METACARPAL AS COMPRESSION ACROSS THE FRACTURE SITE WAS ACHIEVED. THE SCREW REMAINS IN

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 12984855 ↗
  • MalfunctionMay 18, 2021Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INSERTER TIP BROKE OFF DURING THE ANCHOR IMPLANT. THE SITE HAD BEEN PRE-DRILLED. THE TIP WAS RETRIEVED. THEY TRIED TO USE THE AR-8737-61 TO REMOVE THE ANCHOR, BUT THE METAL IN THE DEVICE STARTED TO SPLINTER. THE ANCHOR WAS REMOVED WITH PLIERS. AND THE ANCHOR WAS REMOVED WITH PLIERS. ANOTHER ANCHOR WAS USED TO COMPLETE THE BUNION CASE. PATIENT IS FEMALE, (B)(6) YEARS WITH HARD BONE.

    Patient problems: Insufficient Information

    MAUDE report 11841670 ↗
  • InjuryJan 20, 2020Break

    IT WAS REPORTED BY THE SALES REP, THAT DURING A METATARSAL FRACTURE REPAIR PROCEDURE, THE SURGEON PREPARED THE INSERTION SITE USING A 2.2 DRILL, THE AR-8725-50H SCREW WAS INSERTED, AS THE SURGEON MADE THE FINAL TURNS, TO COUNTERSINK THE SCREW IT BROKE APPROXIMATELY 19-20 MM FROM THE DRIVE INSERTION POINT. THE SURGEON USED AN AR-8737-61, EASY OUT TOOL, ON POWER TO TRY AND RELEASE THE SCREW AND THE TIP OF THE DRIVER BROKE. THE PORTION OF THE SCREW WITH THE DRIVER INSERTION WAS REMOVED. THE REMAINDER OF THE SCREW WAS LEFT IN WITH THE TIP OF THE EASY OUT TOOL STILL IN THE CANNULATION OF THE SCREW. THE CASE WAS COMPLETED WITH THE REMAINDER OF THE SCREW STILL ACROSS THE FRACTURE SITE WITH THE FRACTURE REDUCED. ADDITIONAL INFORMATION OBTAINED 01/06/2020: THE SCREW BROKE BUT THE DRIVER FOR THE SCREW DID NOT BREAK. THE EXTRACTION TOOL BROKE LATER IN THE PROCEDURE. THE REMAINING 30MM OF THE SCREW WAS LEFT IN THE PATIENT, STILL ACROSS THE FRACTURE SITE, WITH THE TIP OF THE EASY OUT TOOL IN THE SCREW'S CANNULATION. THE SURGEON DID NOT NEED TO USE ANY ADDITIONAL SCREWS OR ADDITIONAL INCISIONS. IT WAS A METACARPAL FRACTURE AND THERE WERE LIMITED ACCESS POINTS. THE SURGEON DID NOT NEED TO DO ANYT

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9606551 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page