Arthrex® AR-2323BCT-2-1Recalls & reported problems
Arthrex, Inc. · DBL LOADED 5.5MM BC SWVLK · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00888867264557), received between Nov 22, 2021 and Oct 28, 2024 — classified as 11 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Material Frayed3
- Appropriate Device Problem Term/Code Not Available1
- Difficult to Remove1
- Fitting Problem1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Foreign Body In Patient6
- Device Embedded In Tissue or Plaque2
- Insufficient Information1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2024Break; Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY THE SURGEON WAS PERFORMING THE SUTUREBRIDGE TECHNIQUE USING SWIVELOCK 5,5MM ANCHORS FOR THE LATERAL RAW. AFTER IMPACTING THE ANCHOR AND RETRIEVING THE BLUE GUIDE THE FIBERWIRE SUTURES CAME OUT AS WELL. THE FIBERWIRE SUTURES WERE NOT HOLD IN THE EYELET. IT HAPPENED WITH TWO ANCHORS WITH THE SAME LOT NUMBER. THE SCREW OF SWIVELOCK 5,5 WAS WELL PLACED AND BLOCKED THE SUTURES OF MEDIAL, SO THE ANCHORS STAYED IN THE PATIENT. THERE WAS NO NEED TO REPAIR THE EARS OF THE ROTATOR CUFF THIS TIME, SO THERE WAS NO NEED TO USE THE FIBER WIRES OF THE SWIVELOCK ANCHORS. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICES. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20549378 ↗MalfunctionOct 24, 2024Positioning Failure; Fitting Problem; Positioning Failure; Fitting Problem
IT WAS REPORTED THAT DURING A SURGERY THE SURGEON WAS USING A SWIVELOCK ANCHOR AND THE FIBERWIRE SUTURES WERE NOT PRELOADED. AFTER IMPACTING THE ANCHOR AND RETRIEVING THE BLUE GUIDE THE FIBERWIRE SUTURES CAME OUT AS WELL. THE FIBERWIRE SUTURES WERE NOT HOLD IN THE EYELET. IT HAPPENED WITH TWO ANCHORS WITH THE SAME LOT NUMBER. THE SCREW OF SWIVELOCK 5,5 WAS WELL PLACED AND BLOCKED THE SUTURES OF MEDIAL, SO THE ANCHORS STAYED IN THE PATIENT. THE USED ANCHOR WAS IMPLANTED WITH NO DAMAGE TO THE PATIENT. THE SURGEON DID NOT NEED TO USE THE FIBERWIRE (WHICH WERE NOT PRELOADED) FOR THESE TWO CASES. NO FURTHER INFORMATION RECEIVED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 20524435 ↗InjuryMar 1, 2023Break
IT WAS REPORTED THAT DURING A SHOULDER THE SUTURE THREADS CONTAINED IN THE ANCHOR BROKE CLEANLY AFTER SCREWING AND REMOVING THE BLUE HANDLE OF THE SWIVELOCK. THE ANCHOR REMAINED FIXATED IN THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM ANOTHER MANUFACTURER. DUE TO THE CHANGE OF THE DEVICE A DIFFERENT TECHNIQUE WAS NECESSARY TO FINISH THE SURGERY SUCCESSFULLY. NO FURTHER INFORMATION WERE PROVIDED BY THE CUSTOMER WHAT KIND OF TECHNIQUE CHANGE HAS BEEN PERFORMED.
Patient problems: Foreign Body In Patient
MAUDE report 16459033 ↗MalfunctionMar 14, 2022Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR THE INTERNAL WIRES BROKE WHILE THE SURGEON WAS TIGHTENING THE KNOTS. THE BROKEN WIRE COULD NOT BE RETRIEVED AND REMAINED INSIDE THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 13758091 ↗InjuryFeb 7, 2022Break; Material Frayed; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A ROTATOR CUFF SHOULDER ARTHROSCOPY THE SUTURE THREADS CONTAINED IN THE ANCHOR BROKE AFTER SCREWING AND REMOVING THE GREEN HANDLE FROM THE SWIVELOCK. THE THREADS FRAYED AT THE JUNCTION BETWEEN THE EYELET AND THE METAL GUIDE DURING IMPACTION AND SCREWING. DEVICE REMAINED IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM ANOTHER MANUFACTURER. A CHANGE OF THE SURGICAL TECHNIQUE WAS NECESSARY BECAUSE THE ANCHOR COULD NOT BE USED TO REPAIR THE CAP DUE TO THE ABSENCE OF SUTURE. IT WAS NOT NECESSARY DO A SECOND SURGERY. UPDATE 25-JAN-2022: THE ANCHOR STAYS IN PATIENT WITHOUT THE BROKEN SUTURES, THEY WERE REMOVED FROM PATIENT. THE SURGEON HAD TO USE ANOTHER ANCHOR FROM ANOTHER MANUFACTURER, JUST NEXT TO THE FIRST ONE TO FINISH THE PROCEDURE. SURGEON HAD TO CHANGE THE SURGICAL TECHNIQUE A LITTLE BIT BECAUSE THE ARTHREX ANCHOR TOOK ONE PLACE ON SHOULDER.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13471305 ↗InjuryFeb 7, 2022Break; Material Frayed
IT WAS REPORTED THAT DURING A ROTATOR CUFF SHOULDER SURGERY THE SUTURE THREADS CONTAINED IN THE ANCHOR BROKE AFTER SCREWING AND REMOVING THE GREEN HANDLE FROM THE SWIVELOCK. THE THREADS FRAYED AT THE JUNCTION BETWEEN THE EYELET AND THE METAL GUIDE DURING IMPACTION AND SCREWING. THE DEVICE REMAINS IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM ANOTHER MANUFACTURER. ACCORDING TO THE CUSTOMER IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE, BECAUSE THE ANCHOR COULD NOT BE USED TO REPAIR THE CAP DUE TO ABSENCE OF SUTURE. A SECOND SURGERY WAS NOT NECESSARY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13471068 ↗InjuryFeb 7, 2022Material Frayed
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURE THREADS CONTAINED IN THE ANCHOR BROKE AFTER SCREWING AND REMOVING THE GREEN HANDLE FROM THE SWIVELOCK. THE THREADS FRAYED AT THE JUNCTION BETWEEN THE EYELET AND THE METAL GUIDE DURING IMPACTION AND SCREWING. THE DEVICE REMAINS IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM ANOTHER MANUFACTURER. ACCORDING TO CUSTOMER IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE, BECAUSE THE ANCHOR COULD NOT BE USED TO REPAIR THE CAP DUE TO THE ABSENCE OF SUTURE. IT WAS NOT NECESSARY TO DO A SECOND SURGERY. UPDATE 25-JAN-2022: THE ANCHOR STAYS IN PATIENT WITHOUT THE BROKEN SUTURES, THEY WERE REMOVED FROM PATIENT. THE SURGEON HAD TO USE ANOTHER ANCHOR FROM ANOTHER MANUFACTURER, JUST NEXT TO THE FIRST ONE TO FINISH THE PROCEDURE. SURGEON HAD TO CHANGE THE SURGICAL TECHNIQUE A LITTLE BIT BECAUSE THE ARTHREX ANCHOR TOOK ONE PLACE ON SHOULDER.
Patient problems: Foreign Body In Patient
MAUDE report 13471015 ↗MalfunctionFeb 7, 2022Break; Difficult to Remove
IT WAS REPORTED THAT DURING A ROTATOR CUFF SHOULDER ARTHROSCOPY TWO ANCHORS AR-2323BCT-2-1 WERE USED. THE SUTURES WERE DAMAGED AFTER THE ANCHOR WAS PLACED AND THREE SUTURES BROKE. SURGEON HAD DIFFICULTIES TO REPAIR THE PATIENT'S ROTATOR CUFF. HE COULDN'T SQUEEZE AS HARD AS HE WANTED TO. THE ANCHOR WITH THE BROKEN SUTURES WERE NOT RETRIEVED AND REMAINS IN PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 19-JAN-2022: THE BROKEN SUTURES AND THE ANCHOR, WHERE THE SUTURES BROKE, REMAINS IN PATIENT. IT WAS NOT RETRIEVED. AND WITH THIS ANCHOR THE SURGERY WAS FINISHED SUCCESSFULLY.
Patient problems: Foreign Body In Patient
MAUDE report 13470959 ↗MalfunctionJan 24, 2022Break
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY SURGERY THE DEVICE BROKE WHEN IT HIT THE BONE. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 13333463 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984300 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984246 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984227 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984225 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984124 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. BOTH ANCHORS REMAINED INSIDE THE PATIENT AND WERE NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984103 ↗InjuryDec 13, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE SUTURES CONTAINED IN THE ANCHOR BROKE AFTER THE SWIVELOCK BODY WAS REMOVED. THE REPORTED ANCHOR REMAINED INSIDE THE PATIENT AND WAS NOT RETRIEVED. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM A DIFFERENT MANUFACTURER. IT WAS NECESSARY TO SWITCH THE SURGICAL TECHNIQUE TO USE THE ANCHOR FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984057 ↗MalfunctionNov 22, 2021Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY MULTIPLE WIRES BROKE AS SOON AS THE SURGEON REMOVED THE IMPLANT GUIDE FROM THE SWIVELOCK. THE WIRES BROKE AT THE JUNCTION OF THE WIRE COLOR CHANGE. THE ANCHORS REMAINED INSIDE THE PATIENT WITHOUT BEING USED FOR REPAIR. OTHER TUNNELS HAD TO BE DRILLED AND THE SURGERY WAS FINISHED WITH A DIFFERENT APPROACH AND A DIFFERENT DEVICE FROM ANOTHER MANUFACTURER.
Patient problems: Insufficient Information
MAUDE report 12853844 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page