Arthrex® AR-5207Recalls & reported problems
Arthrex, Inc. · TENSIONLOC COLLAR AND PLUG IMPLANT 7MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888867265813), received between Jun 14, 2021 and May 15, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Difficult to Remove1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Foreign Body In Patient2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2025Break; Break
ON 04/23/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-5207 TENSIONLOC COLLAR AND PLUG IMPLANT HAD A BROKEN COLLAR WHEN INSERTED. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. THE CASE WAS COMPLETED USING ANOTHER AR-5207 TENSIONLOC COLLAR AND PLUG IMPLANT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22030371 ↗MalfunctionMay 1, 2025Break; Break
ON 4/3/2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT THE COLLAR FROM AN AR-5207 TENSIONLOC COLLAR AND PLUG CRACKED WHEN THE PLUG WAS HAMMERED IN. ALL THE BROKEN FRAGMENTS WERE REMOVED. THE CASE WAS COMPLETED USING ANOTHER AR-5207 TENSIONLOC COLLAR AND PLUG. THIS WAS DISCOVERED DURING A PCL RECONSTRUCTION PROCEDURE ON (B)(6) 2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21942112 ↗MalfunctionJan 24, 2024Break; Break
ON 01/05/2024, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5207 TENSION LOCK COLLAR WAS BROKEN WHILE MALLETING. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18573892 ↗MalfunctionJan 24, 2024Break; Break
ON 01/05/2024, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5207 TENSION LOCK COLLAR WAS BROKEN WHILE MALLETING. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18573749 ↗MalfunctionNov 9, 2022Break
ON (B)(6) 2022, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5207 TENSIONLOC COLLAR AND PLUG IMPLANT BROKE DURING INSERTION. ALL BROKEN FRAGMENTS WERE RETRIEVED AND A NEW AR-5207 TENSIONLOC COLLAR AND PLUG IMPLANT WAS USED TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING AN ACLR PROCEDURE ON (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15758790 ↗MalfunctionMay 19, 2022Break
ON (B)(6) 2021 IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5207 TENSION-LOC COLLAR BROKE WHEN SURGEON IMPACTED IMPLANT DURING THE TIBIAL FIXATION. THIS WAS DISCOVERED DURING AN ACL RECONSTRUCTION ON (B)(6) 2021. ALL BROKEN FRAGMENTS WERE RETRIEVE AND PATIENT WAS NOT AFFECTED. A NEW DEVICE WAS USED TO COMPLETE CASE SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14435496 ↗MalfunctionNov 2, 2021Material Separation
ON 10/05/2021 IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5207 TENSIONLOC COLLAR BROKE. THIS OCCURRED DURING AN ACL RECONSTRUCTION. THE CASE WAS COMPLETED USING A NEW PRODUCT. THERE WAS NO PATIENT EFFECT REPORTED. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 10/05/2021 THE DEVICE BROKE IN THE PATIENT, ALL FRAGMENTS WERE REMOVED. THE CASE WAS COMPLETED USING A NEW AR-5207.
Patient problems: Foreign Body In Patient
MAUDE report 12738629 ↗MalfunctionJun 14, 2021Break; Difficult to Remove
IT WAS REPORTED THAT THE AR-5207 TENSIONLOC COLLAR, WAS USED IN ACL RECONSTRUCTION. WHEN THE COLLAR WAS FIXED TO THE BONE HOLE USING A SILICON HAMMER AND THE PLUG OF THE PRODUCT WAS INSERTED INTO THE COLLAR, THE COLLAR OF THE PLUG JOINT BROKE CAUSING THE PRODUCT TO BECOME UNUSABLE. THE SURGEON COULD NOT REMOVE THE PLUG, AND THE PLUG IS STILL LEFT IN THE PATIENT. THE SURGERY WAS COMPLETED USING A PRODUCT FROM ANOTHER MANUFACTURER.
Patient problems: Foreign Body In Patient
MAUDE report 11991234 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page