Arthrex® AR-5210Recalls & reported problems
Arthrex, Inc. · TENSIONLOC COLLAR AND PLUG IMPLANT 10MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867265820), received between May 31, 2023 and Mar 2, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Dislodged or Dislocated1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Failure of Implant1
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2026Break
ON 10-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-5210 TENSIONLOC COLLAR AND PLUG IMPLANT (10 MM) EXPERIENCED AN ISSUE DURING A PROCEDURE PERFORMED ON (B)(6) 2026. A PORTION OF THE COLLAR BROKE WHILE IT WAS BEING HAMMERED IN. THE BROKEN PIECE WAS REMOVED USING AN ELEVATOR, AND A REPLACEMENT AR-5210 IMPLANT WAS SUCCESSFULLY INSERTED. THE PROCEDURE WAS COMPLETED WITHOUT SIGNIFICANT DELAY. NO ADDITIONAL ANESTHESIA WAS REQUIRED, AND NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24482219 ↗MalfunctionAug 25, 2025Break; Break
ON 08/05/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-5210 TENSIONLOC COLLAR AND PLUG IMPLANT HAD AN ISSUE WHEN THE COLLAR WAS INSERTED; IT SANK INTO THE BONE, CAUSING DAMAGE AROUND THE COLLAR. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED USING A REPLACEMENT AR-5210 TENSIONLOC COLLAR AND PLUG IMPLANT. THIS ISSUE WAS DISCOVERED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 08/13/2025: THERE WAS NO SIGNIFICANT CASE DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 08/21/2025: IT WAS FURTHER CLARIFIED THAT THE DAMAGE WAS LIMITED TO THE DEVICE AND THAT THE PATIENT'S BONE TISSUE REMAINED UNAFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22888663 ↗MalfunctionMay 12, 2025Positioning Failure; Positioning Failure
ON 04/14/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-5210 TENSIONLOC¿ COLLAR AND PLUG IMPLANT HAD A FIXATION FORCE THAT WAS WEAK AND AS A RESULT COULD NOT BE USED. THE CASE WAS COMPLETED BY USING ANOTHER DEVICE FROM A DIFFERENT COMPANY. THIS WAS DISCOVERED DURING A CASE (B)(6) 2025, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22001834 ↗MalfunctionSep 8, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 08/14/2023, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5210 TENSIONLOC COLLAR AND PLUG COULD NOT BE SECURED AFTER IMPLANTATION. THIS WAS DISCOVERED DURING A TIBAL FIXATOR FOR ACLR PROCEDURE ON (B)(6) 2023. A NEW PRODUCT WAS USED TO COMPLETE THE CASE. ADDITIONAL INFORMATION RECEIVED ON 8/23/2023: AFTER THE GRAFT WAS IMPLANTED AND PASSED THROUGH THE TENSIONLOC COLLAR, THERE WAS A LOSS IN FIXATION, AND THE PLUG COULD NOT BE SECURED.
Patient problems: Failure of Implant
MAUDE report 17710215 ↗MalfunctionMay 31, 2023Break; Break
ON (B)(4) 2023, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-5210 COLLAR AND PLUG IMPLANT BROKE. THIS OCCURRED DURING A CASE ON (B)(6) 2023, DURING INSERTION A PORTION OF THE COLLAR WAS BROKEN. ALL FRAGMENTS OF THE BROKEN DEVICE WERE RETRIEVED. THE CASE WAS COMPLETED USING ANOTHER AR-5210.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17030504 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page