Arthrex® AR-1941PSRecalls & reported problems

Arthrex, Inc. · 3.9MM KNOTLESS CORKSCREW, PEEK · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00888867270794), received between Jun 27, 2019 and May 24, 2024 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Patient Device Interaction Problem2
  • Patient-Device Incompatibility2
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Material Deformation1
  • Material Split, Cut or Torn1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available2
  • Device Embedded In Tissue or Plaque1
  • Fluid Discharge1
  • Tissue Damage1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT THE ANCHOR FROM THE AR-3638DS AND THE AR-3638 WOULD NOT SEAT. THEY ATTEMPTED WITH AN AR-1941PS WHICH WOULD NOT SEAT. THEY COMPLETED THE CASE (LATERAL EPICONDYLE) WAS ANOTHER AR-1941PS.

    Patient problems: Insufficient Information

    MAUDE report 19392270 ↗
  • MalfunctionMay 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 11/23/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (2) AR-1938PSS PEEK KNOTLESS SUTURETAK ANCHOR AND AN AR-1941PS PEEK KNOTLESS CORKSCREW ANCHOR, PULLED OUT FROM IMPLANTATION SITE. THIS WAS DISCOVERED DURING AN MPFL RECONSTRUCTION PROCEDURE ON (B)(6) 2021. WHEN SURGEON IMPLANTED BOTH SUTURETAK, THE ANCHORS PULLED OUT WITHOUT ANY RESISTANCE. THEN, SURGEON REMOVED THE SUTURETAK ANCHORS AND SWAPPED IT FOR THE CORKSCREW ANCHOR BUT IT ALSO PULLED OUT WITHOUT RESISTANCE. FINALLY, THE SURGEON HAD TO DRILL A BONE TUNNEL TO ACHIEVE ADEQUATE FIXATION AND COMPLETE CASE. AFTER RECEIVING ADDITIONAL INFORMATION ON 11/30/2021, SALES REPRESENTATIVE HAS CONFIRMED THAT ALL THREE FAILED ANCHORS WERE REMOVED FROM THE PATIENT AND THERE WAS NOT BROKEN PIECES LEFT TO RETRIEVED. SURGEON COMPLETED PROCEDURE SUCCESSFULLY AND PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19286067 ↗
  • MalfunctionMar 7, 2024Material Deformation; Material Deformation

    ON 02/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1941PS KNOTLESS CORKSCREW INSERTER WAS STRIPPING THE INSIDE OF THE ANCHOR AND THE ANCHOR WOULD NOT BITE TO BE INSERTED. CASE COMPLETED USING ANOTHER CORKSCREW. 3-MINUTE DELAY. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18859034 ↗
  • MalfunctionJul 26, 2023Break; Break

    ON (B)(6)2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1941PS PEEK KNOTLESS CORKSCREW ANCHOR PULLED OUT. THE PEEK KNOTLESS CORKSCREW WAS INSERTED INTO THE LATERAL EPICONDYLE AND THE BONE WAS PREPPED USING AN AR-1941DS INSTRUMENT KIT IN STANDARD FASHION. THE ANCHOR WAS INSERTED WITH NO PROBLEM, BUT IT PULLED OUT WHEN SHUTTLING THE REPAIR STITCH THROUGH THE KNOTLESS MECHANISM. THIS WAS DISCOVERED DURING AN ECRB DEBRIDEMENT AND REPAIR PROCEDURE ON(B)(6)2023. ADDITIONAL INFORMATION RECEIVED ON (B)(6)2023: THE CASE WAS COMPLETED USING AN AR-3636 KNOTLESS FIBERTAK.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17395395 ↗
  • MalfunctionApr 27, 2021Break

    IT WAS REPORTED AFTER THE KNOTLESS SUTURETAK MALFUNCTIONED, THE SURGEON DECIDED TO UPSIZE TO THE AR-1941PS 3.9 KNOTLESS CORKSCREW TO FINISH THAT PORTION OF THE REPAIR. THE SURGEON USED THE AR-1941D TO DILATE THE SOCKET, BUT THE ANCHOR BROKE IMMEDIATELY UPON INSERTION. THE SURGEON WAS ABLE TO PICK OUT THE LOOSE PIECES USING A GRASPER AND A SHAVER. NO PIECE FROM THE UNSUCCESSFUL ANCHOR WAS LEFT IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 11730921 ↗
  • InjuryJun 16, 2020Break; Failure to Advance; Positioning Problem

    IT WAS REPORTED THAT THE AR-1941PS, 3.9 CORKSCREW ANCHOR (S704108), WAS UNABLE TO BE FULLY IMPLANTED. THE SCREW DRIVER STRIPPED THE ANCHOR CAUSING IT TO NOT BE ABLE TO BE ADVANCED TO A SEATED POSITION AND WAS NOT ABLE TO BE REMOVED. THE ANCHOR NEEDED TO BE BURRED DOWN AND ANOTHER ANCHOR WAS PUT IN IT'S PLACE. INITIAL REPORT ALSO STATED THAT NO PART OF THE IMPLANT WAS LEFT IN THE PATIENT. SURGEON HAS EXPERIENCE USING THE DEVICE. THERE IS NO DEVICE TO RETURN. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION PROVIDED 6/1/2020: THE PROCEDURE WAS A RWMPLISSAGE. A 3.9 PUNCH WAS USED TO PREPARE THE BONE. THE BONE QUALITY WAS AVERAGE AND ACCEPTABLE FOR USE OF A PUNCH VS. A DRILL. THE ANCHOR WAS PARTIALLY REMOVED. THE BOTTOM HALF WAS LEFT IN THE PATIENT AND NOT ABLE TO BE REMOVED. THE TOP HALF OF THE ANCHOR WAS BURRED OFF. THE SALES REP HAS CONFIRMED THAT THE NEW ANCHOR WAS A 3.9 SUTURETAK WHICH WAS PLACED NEXT TO THE ORIGINAL 3.9 CORKSCREW ANCHOR.

    Patient problems: No Code Available

    MAUDE report 10159959 ↗
  • InjuryOct 11, 2019Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A ROTATOR CUFF REPAIR PROCEDURE (DATE OF SURGERY CURRENTLY UNKNOWN). THE PATIENT RE-TORE THEIR ROTATOR CUFF AND HAD A REVISION RCR ON (B)(6) 2019. DURING THE REVISION IT WAS DISCOVERED THAT THE PATIENT'S CUFF WAS IRREPARABLE AND THE SURGEON SWITCHED TO A SUPERIOR CAPSULE RECONSTRUCTION. DUE TO THE SOFT BONE QUALITY THE FOLLOWING ARTHREX PARTS BEING USED WOULD NOT STAY SEATED IN THE BONE: AR-1938BC // LOT: 10265961 // QTY.: 1, AR-1938BC // LOT: F189852 // QTY.: 1. THE REP STATED THE PROCEDURE WAS ARTHROSCOPIC. IT IS UNKNOWN IF ARTHREX PARTS WERE USED DURING THE ORIGINAL RCR; HOWEVER, NO ARTHREX PRODUCTS WERE REMOVED DURING THE REVISION ON (B)(6) 2019. THE REP STATED THERE WERE NO FURTHER ISSUES WITH ANY OTHER ARTHREX PRODUCTS DURING THE CASE. THE HUSBAND OF THE PATIENT CALLED THE SURGEON ON (B)(6) 2019, STATING THE LATERAL PORTAL WOUND WAS DRAINING FLUID AND THEY WERE WORRIED ABOUT INFECTION. THE PATIENT CAME IN FOR A FOLLOW-UP VISIT WITH THE SURGEON ON (B)(6) 2019 TO HAVE THE WOUND LOOKED AT. THE REP STATED THEY CANNOT CONFIRM IF CULTURES WERE TAKEN. AT THE MOMENT NO REVISION PROCEDURE IS SCHEDULED, BUT THE PATIENT HAS BEEN SCHEDULED TO CO

    Patient problems: No Code Available; Tissue Damage

    MAUDE report 9179265 ↗
  • InjuryOct 11, 2019Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A ROTATOR CUFF REPAIR PROCEDURE (DATE OF SURGERY CURRENTLY UNKNOWN). THE PATIENT RE-TORE THEIR ROTATOR CUFF AND HAD A REVISION RCR ON (B)(6) 2019. DURING THE REVISION IT WAS DISCOVERED THAT THE PATIENT'S CUFF WAS IRREPARABLE AND THE SURGEON SWITCHED TO A SUPERIOR CAPSULE RECONSTRUCTION. DUE TO THE SOFT BONE QUALITY THE FOLLOWING ARTHREX PARTS BEING USED WOULD NOT STAY SEATED IN THE BONE: AR-1938BC // LOT: 10265961 // QTY.: 1, AR-1938BC // LOT: F189852 // QTY.: 1. THE REP STATED THE PROCEDURE WAS ARTHROSCOPIC. IT IS UNKNOWN IF ARTHREX PARTS WERE USED DURING THE ORIGINAL RCR; HOWEVER, NO ARTHREX PRODUCTS WERE REMOVED DURING THE REVISION ON (B)(6) 2019. THE REP STATED THERE WERE NO FURTHER ISSUES WITH ANY OTHER ARTHREX PRODUCTS DURING THE CASE. THE HUSBAND OF THE PATIENT CALLED THE SURGEON ON (B)(6) 2019, STATING THE LATERAL PORTAL WOUND WAS DRAINING FLUID AND THEY WERE WORRIED ABOUT INFECTION. THE PATIENT CAME IN FOR A FOLLOW-UP VISIT WITH THE SURGEON ON (B)(6) 2019 TO HAVE THE WOUND LOOKED AT. THE REP STATED THEY CANNOT CONFIRM IF CULTURES WERE TAKEN. AT THE MOMENT NO REVISION PROCEDURE IS SCHEDULED, BUT THE PATIENT HAS BEEN SCHEDULED TO CO

    Patient problems: Fluid Discharge

    MAUDE report 9179207 ↗
  • InjuryJun 27, 2019Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A SUPERIOR CAPSULE REPAIR (SCR) WITH FOUR ANCHORS THE THREADS TORE OFF. TWO ANCHORS WERE LEFT IN THE PATIENT AND TWO WERE REMOVED FROM THE PATIENT. IN ORDER TO REMOVE THE ANCHORS, THE SURGEON NEEDED TO CHANGE THE SURGICAL TECHNIQUE TO AN OPEN PROCEDURE. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO DO A SECOND SURGERY. UPDATE 21-JUN-2019: THE ANCHORS LEFT IN THE PATIENT ARE REPORTED TO BE FIXED AND CANNOT MOVE. NO FURTHER ANCHORS WERE PUT ON IT. IT WAS ALSO REPORTED THAT NO CONSEQUENCES ARE EXPECTED FOR THE PATIENT. AN ADDITIONAL CUT WAS MADE TO BETTER ACCESS THE SURGICAL AREA.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8739514 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →