Arthrex® AR-4020C-06Recalls & reported problems
Arthrex, Inc. · Ø6X 20MM BC IF SCRW, VENTED · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex bioabsorbable interference screw is used to fix soft tissue inside the knee during ligament reconstruction. The 43 reports mostly describe the screw breaking during insertion, and sometimes pieces staying behind in the knee; a few mention the screw sticking, bending, or breaking into fragments. Reports describe 2 injuries and no deaths, and reports naming this screw have risen from 7 in the previous 12 months to 15. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00888867276949), received between Sep 10, 2020 and May 20, 2026 — classified as 2 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Break” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break33
- Material Twisted/Bent3
- Physical Resistance/Sticking3
- Device Dislodged or Dislocated1
- Failure to Advance1
- Fitting Problem1
- Material Fragmentation1
- Material Too Soft/Flexible1
Patient problems
- No Clinical Signs, Symptoms or Conditions35
- Foreign Body In Patient5
- Insufficient Information4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT DURING A MEDIAL PATELLOFEMORAL LIGAMENT RECONSTRUCTION SURGERY THE DEVICE BROKE DURING INSERTION. ALL BROKEN PARTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229474 ↗MalfunctionMay 12, 2026Failure to Advance
ON 24-APR-2026, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-4020C-06 FASTTHREAD BIOCOMPOSITE INTERFERENCE SCREW COULD NOT BE ADVANCED OVER THE NITINOL WIRE, AS RESISTANCE WAS ENCOUNTERED AND THE SCREW APPEARED TO HAVE MATERIAL OBSTRUCTING IT. THIS WAS DISCOVERED DURING AN MPFL PROCEDURE ON (B)(6) 2026.
Patient problems: Insufficient Information
MAUDE report 25150400 ↗MalfunctionApr 8, 2026Break
IT WAS REPORTED THAT DURING THE KNEE SURGERY, THE SCREW BROKE INSIDE THE PATIENT´S KNEE AND A PART OF THE SCREW REMAINED IN THE PATIENT BECAUSE IT COULD NOT BE RETRIEVED. THE OTHER PART OF THE SCREW REMAINED ON THE SCREWDRIVER AND WAS REMOVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 24818207 ↗InjuryMar 18, 2026Break
IT WAS REPORTED THAT DURING THE ANTERIOR CRUCIATE LIGAMENT SURGERY, THE SCREW BROKE INSIDE THE PATIENT. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626231 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED THAT DURING THE ARTHROSCOPY, THE IMPLANT BROKE WHILE BEING INSERTED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 05-MAR-2026 - FURTHER INFORMATION WAS RECEIVED: THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24590997 ↗MalfunctionMar 10, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR SURGERY THE SCREWS AR-4020C-07 (LOT: 15392107) AND AR-4020C-06 (LOT: 15030789) WERE DILATED WITH A TAP BEFORE IMPLANTATION, AS IS CONSISTENT WITH THE CORRECT TECHNIQUE FOR IMPLANTATION INTO HARD BONE, AND YET THE DEVICES BROKE INTO SMALL PIECES WHEN GENTLY SCREWED INTO THE BONE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24563120 ↗MalfunctionFeb 18, 2026Break
IT WAS REPORTED THAT DURING THE KNEE SURGERY, THE SCREW BROKE. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24376410 ↗MalfunctionJan 12, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BECAME STUCK DURING INSERTION FORCING THE SURGEON TO BREAK IT IN ORDER TO REMOVE THE SCREW. ALL FRAGMENTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035482 ↗MalfunctionJan 12, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BECAME STUCK DURING INSERTION FORCING THE SURGEON TO BREAK IT IN ORDER TO REMOVE THE SCREW. ALL FRAGMENTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035100 ↗MalfunctionJan 8, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BECAME STUCK DURING INSERTION FORCING THE SURGEON TO BREAK IT IN ORDER TO REMOVE THE SCREW. ALL FRAGMENTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24012357 ↗MalfunctionJan 8, 2026Break
IT WAS REPORTED THAT DURING A SURGERY THE DEVICE BROKE INTO PIECES WHEN IMPLANTED. NO FURTHER INFORMATION PROVIDED BY THE CUSTOMER.
Patient problems: Insufficient Information
MAUDE report 24005464 ↗MalfunctionDec 2, 2025Break; Break
IT WAS REPORTED DURING THE ARTHROSCOPY THAT THE IMPLANT HAD BROKEN APART WHEN SCREWED IN. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694202 ↗MalfunctionNov 10, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A KNEE SURGERY THE DEVICE WAS DEFORMED OR DEFORMED IN THE FACE, AS IT WAS TOO FLEXIBLE AND GOT TWISTED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 2025: IT WAS CONFIRMED THAT DURING THE SAME SURGERY IN TOTAL 4 DEVICES FROM THE SAME BATCH FAILED. TWO DEVICES BROKE AND TWO GOT BENT. TWO SCREWS WERE THROWN AWAY AND THE OTHER TWO WERE SUCCESSFULLY INSERTED INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511234 ↗MalfunctionNov 10, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY, THE DEVICE WAS DEFORMED OR DEFORMED IN THE FACE, AS IT WAS TOO FLEXIBLE AND GOT TWISTED. PER COMPLAINT REPORTER, THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE (B)(6) 24-OCT-2025: IT WAS CONFIRMED THAT DURING THE SAME SURGERY IN TOTAL 4 DEVICES FROM THE SAME BATCH FAILED. TWO DEVICES BROKE AND TWO GOT BENT. TWO SCREWS WERE THROWN AWAY AND THE OTHER TWO WERE SUCCESSFULLY INSERTED INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511230 ↗MalfunctionNov 10, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE DEVICE WAS DEFORMED OR DEFORMED IN THE FACE, AS IT WAS TOO FLEXIBLE AND GOT TWISTED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE AVOE 24-OCT-2025 IT WAS CONFIRMED THAT DURING THE SAME SURGERY IN TOTAL 4 DEVICES FROM THE SAME BATCH FAILED. TWO DEVICES BROKE AND TWO GOT BENT. TWO SCREWS WERE THROWN AWAY AND THE OTHER TWO WERE SUCCESSFULLY INSERTED INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511226 ↗MalfunctionOct 27, 2025Material Twisted/Bent; Material Twisted/Bent
UPDATE AVOE 24-OCT-2025. IT WAS CONFIRMED THAT DURING THE SAME SURGERY IN TOTAL 4 DEVICES FROM THE SAME BATCH FAILED. TWO DEVICES BROKE AND TWO GOT BENT. TWO SCREWS WERE THROWN AWAY AND THE OTHER TWO WERE SUCCESSFULLY INSERTED INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23391167 ↗MalfunctionJul 25, 2025Break; Break
IT WAS REPORTED THAT DURING A SURGERY THE DEVICE SHEARED INTO PIECES DURING THE IMPLANTATION. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610976 ↗MalfunctionJul 7, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE DEVICE BROKE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 22414686 ↗MalfunctionMay 5, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE SCREW BROKE WHILE SCREWING. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 05-MAY-2025. ADDITIONAL INFORMATION WAS RECEIVED THAT ALL FRAGMENTS OF THE BROKEN SCREW WERE RETRIEVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21961144 ↗MalfunctionApr 23, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BROKE. NOTHING BROKE OFF INSIDE THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21890684 ↗MalfunctionApr 22, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE IMPLANT BROKE DURING THE INSERTION. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21883317 ↗MalfunctionMar 17, 2025Break; Break
IT WAS REPORTED THAT DURING A LIGAMENTOPLASTY SURGERY THE SCREW BROKE. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21620578 ↗MalfunctionOct 8, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A LIGAMENTOPLASTY SURGERY THE SCREW DISLOCATED AT ITS UPPER PART. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20397949 ↗MalfunctionJul 23, 2024Break; Break
IT WAS REPORTED THAT DURING A KNEE LIGAMENTOPLASTY SURGERY THE DEVICE BROKE INSIDE THE PATIENTS FEMUR. ALL PARTS WERE RETRIEVED FROM THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19809695 ↗MalfunctionJul 1, 2024Break; Break
ON 06/17/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-4020C-06 FASTTHREAD¿ BIOCOMPOSITE¿ INTERFERENCE SCREWS TIP WAS IMMEDIATELY CHIPPED OFF UPON INITIAL INSERTION. THIS OCCURRED DURING A CASE AND DESPITE MULTIPLE ATTEMPTS TO CONTINUE TO INSERT THE SCREW, IT FAILED TO GO IN. THE PATIENT WAS NOT HARMED, AND WE COMPLETED THE CASE SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19644510 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗