Arthrex® AR-4020C-09Recalls & reported problems
Arthrex, Inc. · Ø9X 20MM BC IF SCRW, VENTED · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex bio-composite interference screw is an implant used to fix soft tissue during surgery. The 26 reports mostly describe the screw breaking during ACL surgery, with broken pieces sometimes retrieved from the patient; another report describes a localized granuloma. There were 2 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00888867276970), received between Mar 29, 2021 and Aug 28, 2026 — classified as 2 injuries, 24 malfunctions.
The most frequently reported specific device problem is “Break” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break21
- Biocompatibility1
- Defective Device1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Installation-Related Problem1
- Material Fragmentation1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- Foreign Body In Patient2
- Insufficient Information2
- Failure of Implant1
- Granuloma1
- Hypersensitivity/Allergic reaction1
- Swelling/ Edema1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2026Patient Device Interaction Problem
On 14-Aug-2026, a sales representative reported via email that a patient who underwent an ACL reconstruction on (b)(6) 2025 developed a caseating granuloma at the tibial screw site. The procedure involved a BTB ACL reconstruction utilizing an AR-4020C-07 FastThread BioComposite Interference Screw (7 x 20 mm) in the femur and an AR-4020C-09 FastThread BioComposite Interference Screw (9 x 20 mm) in the tibia. The reported reaction was localized to the tibial site associated with the AR-4020C-09 screw. No issues were reported with the AR-4020C-07 screw implanted in the femur.Additional information was received on 19-Aug-2026:Follow-up information was received from the surgeon. The surgeon reported that the first signs and symptoms were observed in (b)(6) 2026 and consisted of painful swelling. The surgeon further reported that the diagnosis of a caseating granuloma was based on MRI findings and operative findings. Cyst formation was reportedly identified, and the surgeon suspected the screw contributed to the granuloma because the reaction was localized to the screw location. The surgeon reported that pathology and culture results showed no evidence of infection. Infection was reporte
Patient problems: Granuloma; Swelling/ Edema
MAUDE report 50497111 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT DURING A SURGERY THE SCREW BROKE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232236 ↗MalfunctionNov 5, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SCREW BROKE OFF AT THE TIP. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478900 ↗MalfunctionJun 16, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) RECONSTRUCTION SURGERY WITH SEMITENDINOSUS AND GRACILIS STEMMED IT WAS TRIED TO INSTALL A FEMORAL SCREW AND THE DEVICE BROKE RIGHT AT THE BEGINNING. ALL FRAGMENTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (COMPETITOR¿S DEVICE¿). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22210399 ↗MalfunctionFeb 24, 2025Break; Break
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-4020C-09 FASTTHREAD¿ BIOCOMPOSITE¿ INTERFERENCE SCREW BROKE. THIS OCCURRED DURING AN ACL RECONSTRUCTION ON (B)(6) 2025, WHEN THEY TAPPED TWICE AND IT BROKE. ALL FRAGMENTS WERE ABLE TO BE RETRIEVED FROM THE PATIENT. THE PATIENT HAD DECENT BONE QUALITY, THAT WAS PREPARED USING THE TAP. THE CASE WAS COMPLETED USING ANOTHER AR-4020C-09 FASTTHREAD¿ BIOCOMPOSITE¿ INTERFERENCE SCREW BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21446262 ↗MalfunctionFeb 4, 2025Break; Break
ON 01/13/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-4020C-09 FASTTHREAD BIOCOMPOSITE INTERFERENCE SCREW BROKE INTO THREE PIECES DURING USE. THE CASE WAS COMPLETED USING ANOTHER ONE WITH THE SAME LOT NUMBER WITHOUT ISSUES. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 01/20/2025: AR-4020C-09 FASTTHREAD BIOCOMPOSITE INTERFERENCE SCREW BROKE INTO THREE PIECES WHILE INSERTING THE SCREW. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THEY USED ANOTHER 9X20 INTERFERENCE SCREW TO COMPLETE THE CASE. THERE WAS NO CASE DELAY OR ADDED ANESTHESIA ADMINISTERED TO THE PATIENT. THE GRAFT USED IN THE PROCEDURE WAS A QUAD TENDON. THIS WAS DISCOVERED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6)2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21304205 ↗MalfunctionJul 1, 2024Break; Break
IT WAS REPORTED THAT DURING AN ACL SURGERY THE DEVICE BROKE WHILE SCREWING IN. ALL PARTS WERE RETRIEVED FROM THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH ANOTHER DEVICE. A FAST THREAD SCREW WITH 8X23MM WAS USED. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19647014 ↗MalfunctionJun 27, 2024Break; Break
IT WAS REPORTED THAT THE SCREW BROKE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE DW 17-JUN-2024: FURTHER INFORMATION WAS PROVIDED THAT THE REPORTED EVENT OCCURRED DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WHILE THE SCREW WAS BEING INSERTED INTO THE DRILL CHANNEL. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PARTS AND FINISHED THE SURGERY SUCCESSFULLY WITH A DIFFERENT SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624444 ↗MalfunctionJun 10, 2024Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THERE WAS A BREAKAGE OF THE FAST THREAD INTERFERENCE SCREW DURING INSERTION OF THE DEVICE INTO THE FEMORAL BONE TUNNEL. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. ACCORDING TO THE SURGEON, NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19497352 ↗MalfunctionMay 7, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE IMPLANT SPLIT AT THE END. NOTHING BROKE OFF INSIDE THE PATIENT. THE SURGEON REPORTED THAT THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19257691 ↗MalfunctionMay 1, 2024Defective Device; Defective Device
ON 12/27/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-4020C-09 FASTTHREAD SCREW HAD STRUCTURAL DAMAGE AND LOST SHAPE AFTER IT WAS INSERTER. THIS WAS DISCOVERED DURING AN ACL PROCEDURE ON (B)(6) 2021. SURGEON USED THE PROPER TAP PRIOR OF SCREW INSERTION AND THEN WHEN SCREW WAS PUTTING IT DID NOT WORK AND SURGEON REMOVED IT. IT WAS NOTICED THAT THE LOWER THREADS HAD LOST ITS SHAPE AND SURGEON WAS UNABLE TO USE IT AGAIN. ADDITIONAL INFORMATION RECEIVED ON 12/27/2021: NOTHING BROKE INSIDE THE PATIENT AND CASE WAS COMPLETED USING AN AR-1390E FROM LOT 13435877 WITHOUT FURTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220681 ↗MalfunctionJan 29, 2024Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BROKE IN THE FEMUR BONE DURING INSTALLATION. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18597607 ↗InjuryNov 21, 2023Biocompatibility; Biocompatibility
ON (B)(6) 2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT A RIGHT KNEE ARTHROSCOPY AND ACLR WITH ALLOGRAFT PROCEDURE ON 12/16/2022. ON (B)(6) 2022, THE PATIENT STARTED TO EXPERIENCE ITCHINESS AND SMALL HIVES FORMING ON THE RIGHT LEG NEAR THE SCAR. BY (B)(6) 2022, THE PATIENT NOTICED REDNESS AROUND THE INCISION SITE. THE PATIENT INFORMED THE SURGEON AND WAS GIVEN ANTIBIOTICS AND BENADRYL. BY (B)(6) 2023, THE ISSUE WAS RESOLVED AND THE SKIN LOOKED NORMAL. ON (B)(6) 2023, THE PATIENT NOTICED SWELLING AND REDNESS AT THE INCISION SITE. THE SURGEON PRESCRIBED ANTIBIOTICS BUT, EVENTUALLY, AN ABSCESS-LIKE BUMP FORMED IN THE LEG, WHICH WAS DRAINED IN AN IN-OFFICE PROCEDURE. LATER, THE ABSCESS RE-FORMED AND BURST ON ITS OWN, AND ANOTHER ROUND OF ANTIBIOTICS WAS PRESCRIBED. THE PATIENT UNDERWENT I &D. AS OF (B)(6) 2023, THE PATIENT IS STILL EXPERIENCING REDNESS, SWELLING, PITTING EDEMA IN THE RIGHT LEG, ABSCESS, DRAINAGE, AND WOUND DEHISCENCE. THE PATIENT HAS HAD BLOODWORK AND CULTURES TO FIND THE REACTION CAUSE.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 18181851 ↗MalfunctionJun 7, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY DURING THE INSTALLATION OF THE SCREW, IT BROKE IN TWO PIECES. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 17081179 ↗MalfunctionMay 2, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREWS BROKE INSIDE THE PATIENT WHEN SCREWING IN WITH THE SCREWDRIVER. THE BROKEN PIECE WAS RETRIEVED FROM PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16857696 ↗MalfunctionJan 30, 2023Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY DURING INSERTING THE SCREW THE DISTAL TIP BROKE OFF AND PREVENTED THE IMPLANTATION. THIS INCIDENT OCCURRED WITH A 9X20MM SCREW AND THEN WITH A 9X30MM SCREW. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16263368 ↗MalfunctionNov 7, 2022Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE SCREW BROKE AT THE BEGINNING OF THE INSERTION. THE BROKEN OFF PART WAS RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON, NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15743096 ↗MalfunctionOct 31, 2022Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY DURING TIBIAL FIXATION THE DEVICE BROKE OFF. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15698339 ↗MalfunctionSep 26, 2022Break; Material Fragmentation
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT PATELLAR TENDON SURGERY THE INTERFERENCE SCREW BROKE DURING INSTALLATION. THE SCREW WAS PARTIALLY REMOVED AND FRAGMENTS REMAINED IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-4020C-08). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 15484126 ↗MalfunctionJul 25, 2022Break; Installation-Related Problem
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THERE WAS A FAILURE OF INSTALLATION AT HPJM. THE SCREW BROKE OFF INSIDE THE PATIENT IN THE BONE TUNNEL AT THE BEGINNING OF SCREWING. THE BROKEN PIECE WAS REMOVED FROM PATIENT AND A NEW ONE HAD TO BE USED TO FINISH THE SURGERY SUCCESSFULLY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15093640 ↗MalfunctionJul 5, 2022Break
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY (ANTERIOR CRUCIATE LIGAMENT) THE SCREW BROKE WHEN IT WAS INSERTED IN THE FEMUR. IT WAS REMOVED FROM PATIENT AND A NEW SCREW HAD TO BE USED TO FINISH THE SURGERY SUCCESSFULLY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14914611 ↗MalfunctionApr 4, 2022Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT LIGAMENTOPLASTY SURGERY THE HEAD OF THE DEVICE BECAME DETACHED FROM THE BODY AT THE TIME OF SCREWING. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-4020C-06). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13995364 ↗MalfunctionOct 12, 2021Break
ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-4020C-09 BIO COMPOSITE 9 X 20 INTERFERENCE SCREW BROKE DURING INSERTION. ALL BROKEN FRAGMENTS WERE REMOVED FROM THE PATIENT. THIS WAS DISCOVERED DURING USE IN A ACL REPAIR ON (B)(6) 2021. THE CASE WAS COMPLETED BY A 9 X 23 INTERFERENCE SCREW.
Patient problems: Insufficient Information
MAUDE report 12615560 ↗MalfunctionSep 28, 2021Break; Difficult to Remove
IT WAS REPORTED THAT DURING KNEE ARTHROSCOPY SURGERY THE SCREW BROKE OFF WHILE INSERTING. THE BROKEN OFF PIECE HAS NOT BEEN RETRIEVED FROM THE PATIENT. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. NO FURTHER INFORMATION RECEIVED. UPDATE, (B)(6) 2021 RECEIVED FURTHER INFORMATION THAT THE SCREW IS SECURE INSIDE THE FEMUR BONE AND CANNOT BE MOVED/ IS NOT LOOSENING.
Patient problems: Foreign Body In Patient
MAUDE report 12540056 ↗MalfunctionMar 31, 2021Break
IT WAS REPORTED THAT DURING A PATELLAR TENDON ALLOGRAFT ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE DEVICE (AR-4020C-09, LOT 11764659) BROKE INSIDE THE PATIENT IN THE MIDDLE OF IMPLANTING IN A 9 MM DIAMETER TIBIAL TUNNEL. THE SURGEON USED A TAP BEFORE INSERTING THE DEVICE AS A BONE TENDON BONE GRAFT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11599597 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page