Skip to content

Arthrex® AR-1318FT-1Recalls & reported problems

Arthrex, Inc. · SUTURE ANCHOR, MICRO CORKSCREW FT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00888867282322), received between Apr 27, 2021 and Aug 19, 2026 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Material Split, Cut or Torn3
  • Component Misassembled1
  • Device Dislodged or Dislocated1
  • Device Slipped1
  • Loose or Intermittent Connection1
  • Physical Resistance/Sticking1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information3
  • Foreign Body In Patient2

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2026Loose or Intermittent Connection; Component Misassembled

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50325265 ↗
  • MalfunctionJun 2, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A HAND SURGERY THE ANCHOR DID NOT HOLD, TWISTED IN THE BONE AND COULDN'T BE FIXATED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22124120 ↗
  • MalfunctionOct 16, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ULNAR COLATERAL LIGAMENT REPAIR SURGERY THE SUTURE TORE AFTER INSERTING INTO THE PHALANX. THE SURGEON REPORTED THAT THE SUTURE LOADED INTO THE ANCHOR WAS ALREADY WEAKENED BEFORE THE ANCHOR INSERTION. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE PART NUMBER: AR-1317FT-1. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE KPILZ 10-OCT-2024: THE TWO DEVICES AR-1318FT-1 REMAINED INSIDE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20461524 ↗
  • MalfunctionOct 16, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ULNAR COLLATERAL LIGAMENT REPAIR SURGERY THE SUTURE TORE AFTER INSERTING INTO THE PHALANX. THE SURGEON REPORTED THAT THE SUTURE LOADED INTO THE ANCHOR WAS ALREADY WEAKENED BEFORE THE ANCHOR INSERTION. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 10-OCT-2024: THE TWO DEVICES AR-1318FT-1 REMAINED INSIDE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20461499 ↗
  • MalfunctionJun 13, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A TFCC (TRIANGULAR FIBROCARTILAGE COMPLEX) SURGERY AFTER IMPLANTATION OF THE MICRO CORKSCREW IN THE PATIENT'S BONE, ONLY ONE SUTURE WAS RELEASED. THE OTHER SUTURE WAS STUCK IN THE SCREWDRIVER. AN ATTEMPT WAS MADE TO FREE IT BY PIVOTING THE TITANIUM WINGS OF THE SCREWDRIVER, BUT THIS DISRUPTED THE STRUCTURE OF THE SUTURE. IT WAS NECESSARY TO CUT THE THREAD AND EXTRACT THE IMPLANT FROM THE PATIENT'S BONE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19533936 ↗
  • MalfunctionMar 18, 2024Break; Break

    IT WAS REPORTED THAT DURING A SLIL RECONSTRUCTION SURGERY THE IMPLANT BROKE WHILE BEING SCREWED INTO A PRE-DRILLED HOLE, CAUSING THE SUTURE BREAKAGE. THE BROKEN SUTURE WAS REMOVED FROM PATIENT AND THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE SWIT (B)(6) 2024: FIRST THE ANCHOR BROKE AND THIS CAUSED THE SUTURE BREAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18923960 ↗
  • MalfunctionJan 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A WRIST LIGAMENT RECONSTRUCTION THE ANCHOR SLIPPED OUT OF THE BONE AND WAS UNUSABLE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18437109 ↗
  • MalfunctionMar 7, 2022Break

    IT WAS REPORTED THAT DURING A HAND ARTHROSCOPY THE ANCHOR BROKE OFF WHEN IT WAS IMPLANTED. THE BROKEN PART WAS RETRIEVED OUT OF THE PATIENT. ACCORDING TO THE SURGEON THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13684470 ↗
  • MalfunctionJul 27, 2021Break

    IT WAS REPORTED THAT DURING A LATERAL ANKLE LIGAMENT RECONSTRUCTION SURGERY THE TIP OF THE CORKSCREW OF THE DEVICE WAS DAMAGED DURING INSERTION. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12231652 ↗
  • MalfunctionJul 27, 2021Break; Device Slipped

    IT WAS REPORTED THAT DURING A LATERAL ANKLE LIGAMENT RECONSTRUCTION SURGERY THE IMPLANT OF THE DEVICE FELL OFF THE CORKSCREW DURING INSERTION. THE BROKEN OFF PIECE WAS RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12231628 ↗
  • MalfunctionMay 10, 2021Break; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A HAND SURGERY THE ANCHOR CUT THE SUTURE DURING SCREWING IN. THE BROKEN OFF PIECE WAS RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 11798580 ↗
  • MalfunctionApr 27, 2021Break

    IT WAS REPORTED THAT DURING A SURGERY THE SURGEON HAD A PROBLEM WITH THE ITEMS AS TWO ANCHORS BROKE. NO FURTHER INFORMATION WAS RECEIVED.

    Patient problems: Insufficient Information

    MAUDE report 11730831 ↗
  • MalfunctionApr 27, 2021Break

    IT WAS REPORTED THAT DURING A SURGERY THE SURGEON HAD A PROBLEM WITH THE ITEMS AS TWO ANCHORS BROKE. NO FURTHER INFORMATION WAS RECEIVED.

    Patient problems: Insufficient Information

    MAUDE report 11730826 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page