Arthrex® AR-1938PS-1Recalls & reported problems

Arthrex, Inc. · SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867282421), received between Oct 5, 2021 and Mar 30, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2
  • Device Damaged by Another Device2
  • Failure to Advance2
  • Positioning Failure2
  • Use of Device Problem2
  • Defective Device1
  • Fitting Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Insufficient Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Positioning Failure

    IT WAS REPORTED THAT DURING A SURGERY THE ANCHOR DID NOT HOLD AND CAME LOOSE DURING TIGHTENING. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24713226 ↗
  • MalfunctionApr 21, 2025Positioning Failure; Fitting Problem; Positioning Failure; Fitting Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY THE DEVICE COULD NOT BE ANCHORED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21873442 ↗
  • MalfunctionApr 18, 2025Break; Device Damaged by Another Device; Break; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE SIGHT GOT DISINTEGRATED AND THE ANCHOR BECAME DAMAGED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21865795 ↗
  • MalfunctionApr 18, 2025Break; Device Damaged by Another Device; Break; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE SIGHT GOT DISINTEGRATED AND THE ANCHOR BECAME DAMAGED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21865577 ↗
  • MalfunctionFeb 20, 2025Use of Device Problem; Failure to Advance; Use of Device Problem; Failure to Advance

    IT WAS REPORTED THAT DURING A KNEE CAPSULE REPAIR SURGERY THE SUTURE DID NOT TIGHTEN. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21425184 ↗
  • MalfunctionFeb 20, 2025Use of Device Problem; Failure to Advance; Use of Device Problem; Failure to Advance

    IT WAS REPORTED THAT DURING A KNEE CAPSULE REPAIR SURGERY THE SUTURE DID NOT TIGHTEN. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21424979 ↗
  • MalfunctionMay 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A LABRUM REFIXATION SURGERY WITH KNOTLESS SUTURETAK WHEN INSERTING THE ANCHOR, THE PEEKHART BODY HELD UNDER TENSION, BUT WHEN SHUTTLING AND INSERTING THE TENSION AND MAIN SUTURE, THE ANCHORS CAME LOOSE AND WERE DISCARDED AND REPLACED BY 2 NEW ANCHORS. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19329625 ↗
  • MalfunctionMay 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A LABRUM REFIXATION SURGERY WITH KNOTLESS SUTURETAK WHEN INSERTING THE ANCHOR, THE PEEKHART BODY HELD UNDER TENSION, BUT WHEN SHUTTLING AND INSERTING THE TENSION AND MAIN SUTURE, THE ANCHORS CAME LOOSE AND WERE DISCARDED AND REPLACED BY 2 NEW ANCHORS. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19329518 ↗
  • MalfunctionApr 25, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT DURING A BANKART AND SLAP REPAIR SURGERY WHEN DRIVING THE ANCHOR IN, IT WAS DAMAGED AND NO STABILITY WAS OBTAINED. THE BONE WAS PROPERLY PREPARED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE, AR-2325BSS-1. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19180389 ↗
  • MalfunctionApr 10, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY SURGERY THERE WAS A RUPTURE OF THE BINDING SUTURE OF THE DEVICE DURING COLLARBONE STABILISATION. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19080855 ↗
  • MalfunctionOct 5, 2021Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A SURGERY IT WAS NOT POSSIBLE TO IMPLANT THE DEVICES COMPLETELY. THE DEVICES GOT STUCK AND NO FURTHER TIGHTENING OF THE SUTURES WAS POSSIBLE. NO PART OF THE DEVICES BROKE OFF. THE TWO DEVICES WITH THE SAME ERROR REMAINED INSIDE THE PATIENT AND THE SURGEON SWITCHED THE PROCEDURE FROM A KNOTLESS TO A KNOTTED PROCEDURE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (1934BC). IT WAS NOT NECESSARY TO DO A SECOND SURGERY. UPDATE 22-SEP-2021: IT WAS CONFIRMED THAT IT WAS NOT POSSIBLE TO PERFORM THE FINAL TIGHTENING OF THE AFFECTED DEVICES AND THESE WERE NOT REMOVED BECAUSE DRILLING OVER THEM WOULD ONLY HAVE CAUSED MORE DAMAGE TO THE PATIENT. NEW TUNNELS HAD TO BE DRILLED FOR THE KNOTTED ANCHORS. UPDATE 27-SEP-2021: IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING A SHOULDER (LABRUM) SURGERY. AFTER THE SUTURES GOT STUCK, THEY WERE CUT FROM THE ANCHORS AND REMOVED. THE ANCHOR BODIES REMAINED FULLY IMPLANTED INSIDE THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 12576867 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →