Skip to content

Arthrex® AR-3350-4031Recalls & reported problems

Arthrex, Inc. · 4K Arthroscope, 30°, 4 mm x 152.5 mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867291690), received between Jul 6, 2023 and Feb 18, 2025 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component2
  • Crack1
  • Device Dislodged or Dislocated1
  • Display or Visual Feedback Problem1
  • Leak/Splash1
  • No Display/Image1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Burn(s)1
  • Insufficient Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2025No Display/Image; Display or Visual Feedback Problem; No Display/Image; Display or Visual Feedback Problem

    NO COMPLAINT ALLEGATION WAS PROVIDED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION WAS RECEIVED. UPDATE DW 12-FEB-2025: FURTHER INFORMATION WAS PROVIDED THAT DURING A SURGERY IT WAS FOUND THAT NOTHING COULD BE SEEN THROUGH THE USED OPTIC. AS NO REPLACEMENT DEVICE WAS AVAILABLE THE CUSTOMER HAD TO ABORT THE SURGERY AND THE PATIENT HAD TO BE AWAKENED. UPDATE DW 13-FEB-2025: IT WAS FURTHER REPORTED THAT THE REPORTED DEVICE STOPPED WORKING DURING A KNEE SURGERY ON THE MENISCUS. IT IS NOT KNOWN IF THE PATIENT RECEIVED AN ADDITIONAL TREATMENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21396586 ↗
  • MalfunctionFeb 18, 2025Sharp Edges; Sharp Edges

    ON 01/31/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3350-4031 SCOPE HAS SHARP METAL PIECE ON THE DISTAL TIP. DISCOVERED DURING A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21393378 ↗
  • MalfunctionSep 6, 2024Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC KNEE SURGERY A RUBBER STRIP HAS COME OFF THE OCULAR AND ALWAYS PRODUCED A BLACK STRIPE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155904 ↗
  • InjuryJul 17, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SURGERY THE SCOPE, LIGHT CABLE OR BOTH OVER HEATED AND BURNED THE PATIENTS LEG. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Burn(s)

    MAUDE report 19760717 ↗
  • MalfunctionJun 7, 2024Break; Break

    IT WAS REPORTED ON 05/04/2022 BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-3350-4031 SCOPES BLACK EYEPIECE FELL OFF. THIS WAS DISCOVERED DURING AN UNSPECIFIED TIME WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19488955 ↗
  • MalfunctionJun 7, 2024Break; Break

    ON 01/16/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3350-4031 SCOPE AND AN AR-3350-4071 SCOPE BOTH HAVE A DETACHED BLACK EYE PIECE. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19486910 ↗
  • MalfunctionJun 6, 2024Break; Break

    IT WAS REPORTED ON 1/11/2022 BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-3350-4031 SCOPE IS FALLING APART "GLUE IS COMING LOOSE". THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19475385 ↗
  • MalfunctionJan 26, 2024Crack; Crack

    ON 01/17/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-3350-4031 SCOPE BASE IS CRACKED AND DIFFICULT TO STAY ATTACHED TO CAMERA. THIS OCCURRED DURING AN ARTHROSCOPIC PROCEDURE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18587430 ↗
  • MalfunctionNov 27, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 11/6/2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS-06392638 THAT AN AR-3350-4031 ARTHROSCOPE CAME APART DURING SURGERY. THE EYEPIECE DETACHED ITSELF FROM THE SCOPE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18209574 ↗
  • MalfunctionJul 13, 2023Leak/Splash; Leak/Splash

    ON (B)(6) 2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-3350-4031 ARTHROSCOPE IS EXPERIENCING WATER INVASION, IT IS FILLING UP THE BOTTOM PORTION OF THE LENS. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17314952 ↗
  • MalfunctionJul 6, 2023Break; Break

    ON 6/15/2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-3350-4031 4K ARTHROSCOPE LENS WAS CRACKED AND NOT GIVING A CLEAR IMAGE. THE CASE WAS COMPLETED BY SWITCHING OUT THE SCOPE FOR ANOTHER. THIS WAS DISCOVERED DURING AN ARTHROSCOPY PROCEDURE, WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17272927 ↗

Other Arthroscope devices

All Arthroscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page