Arthrex® AR-3650DSRecalls & reported problems

Arthrex, Inc. · FiberTak RC Disposables Kit · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867291799), received between Sep 10, 2024 and Jul 27, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Contamination” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination1
  • Defective Device1
  • Device Dislodged or Dislocated1
  • Material Fragmentation1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2026Positioning Failure; Defective Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50006838 ↗
  • MalfunctionMar 24, 2026Contamination

    ON 04-MAR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3650DS DISPOSABLES KIT FOR 2.6 FIBERTAK HAD A PIECE OF FOAM IN THE 2.6 FIBERTAK RC GUIDE. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM. THE FOAM PIECE WAS REMOVED THE PATIENT. ADDITIONAL INFORMATION PROVIDED 13-MAR-2026: IT WAS FURTHER REPORTED THAT THE FOAM PIECE CAME OUT WHEN OBTURATOR WAS PLACED INTO THE GUIDE. THIS OCCURRED DURING A ROTATOR CUFF REPAIR PROCEDURE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674078 ↗
  • MalfunctionFeb 26, 2025Material Fragmentation; Material Fragmentation

    ON 02/12/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3650DS DISPOSABLES KIT FOR 2.6 FIBERTAK ANCHORS AND (QTY. 3) OF AN AR-3632SP 2.6 FIBERTAK SP SOFT ANCHOR HAD AN ISSUE WHEN REPAIRING A SUPRASPINATUS TEAR. THE FIRST AR-3632SP 2.6 FIBERTAK SP SOFT ANCHOR WAS DEPLOYED SUCCESSFULLY, BUT THE DRILLING WAS PROBLEMATIC FOR THE FIRST ANCHOR. THE 2.6 DRILL WAS CONFIRMED TO BE APPROPRIATELY CHUCKED. THE SECOND SOCKET WAS MORE DIFFICULT, AND SOME METAL SHAVINGS WERE NOTICED AROUND THE DRILL SITE. THE SURGEON HAD TO REVERSE THE DRILL TO REMOVE IT FROM THE DRILL GUIDE, WHICH CAUGHT THE SUTURES FROM THE FIRST ANCHOR AND PULLED IT FROM THE SOCKET. THE SURGEON ASKED FOR TWO MORE AR-3632SP 2.6 FIBERTAK SP SOFT ANCHORS AND ATTEMPTED TO DEPLOY THEM IN EACH SOCKET, BUT BOTH PULLED OUT. THE SURGEON ABANDONED THE FIBERTAKS AND MOVED ON TO 5.5MM CORKSCREWS, WHICH WERE SUCCESSFULLY DEPLOYED IN THE PREVIOUSLY DRILLED SOCKETS. ALL METAL SHAVINGS WERE REMOVED FROM THE SITE, AND THE MEDIAL ROW SUTURES WERE PASSED, TIED, AND BRIDGED TO TWO 5.5MM SWIVELOCKS AS PLANNED. THIS WAS DISCOVERED DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21467544 ↗
  • MalfunctionSep 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2024, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-2600FSB-1 FIBERTAK SPEED BRIDGE IMPLANT SYSTEM, WHILE USING THE AR-3650DS DISPOSABLES KIT FOR 2.6 FIBERTAK, HAD AN ISSUE WITH TWO OUT OF THE THREE 2.6 FIBERTAK RC SELF-PUNCHING SOFT ANCHORS RIPPING AND PULLING OUT AND NEVER FULLY SETTING UPON IMPLANTING. THE CASE WAS COMPLETED USING A 4.75 SWIVELOCK. THERE WAS UNDER 15 MINUTES OF DELAY IN REMOVING THE TWO SUTURE ANCHORS THAT HAD RIPPED. THIS WAS DISCOVERED DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20181196 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices