Arthrex® AR-4068HLRecalls & reported problems
Arthrex, Inc. · SwiftStitch Hip Suture Passer · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00888867300217), received between Dec 15, 2023 and Mar 30, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Material Twisted/Bent3
- Device Damaged by Another Device1
- Device Dislodged or Dislocated1
- Material Split, Cut or Torn1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information3
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Break
IT WAS REPORTED THAT DURING A HIP LABRUM SURGERY THE TIP OF THE DEVICE BROKE. NO PART OF THE DEVICE REMAINED INSIDE THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24716838 ↗MalfunctionFeb 23, 2026Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC HIP SURGERY THE FRONT PART OF THE DEVICE GOT BENT AND COULD NO LONGER BE USED WITHOUT BREAKING OFF. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413635 ↗MalfunctionFeb 4, 2026Break
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY ONE DEVICE BROKE AND A SECOND DEVICE BENT AT THE TIP. THE BROKEN PART WAS REMOVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24258530 ↗MalfunctionFeb 4, 2026Material Twisted/Bent
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY ONE DEVICE BROKE AND A SECOND DEVICE BENT AT THE TIP. THE BROKEN PART WAS REMOVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24258485 ↗MalfunctionJan 23, 2026Break
IT WAS REPORTED THAT DURING THE HIP ARTHROSCOPY, THE TIP OF THE DEVICE BROKE WHEN IT WAS ATTACHED TO THE HIP BEFORE PRESSURE WAS APPLIED. NO PARTS REMAINED IN THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24160185 ↗MalfunctionJan 5, 2026Break
ON 15-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT THE AR-4068HL SWIFTSTITCH HIP SUTURE PASSER TIP OF THE SWIFTSTITCH BROKE WHILE TRYING TO PASS THE REPAIR SUTURE THROUGH THE LABRUM FOR AN ARTHROSCOPIC HIP LABRAL REPAIR. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23957151 ↗MalfunctionMay 5, 2025Material Twisted/Bent; Material Twisted/Bent
ON 04/22/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT (2) AR-3636H SELF BUNCHING 1.8 KNOTLESS HIP FIBERTAK®SOFT ANCHORS BACKED OUT, AN AR-3638DHS-2 DISPOSABLES KITS HANDLE BENT, AND AN AR-4068HL SWIFTSTITCH¿ SUTURE PASSER BENT. THIS OCCURRED DURING A HIP ARTHROSCOPY ON (B)(6) 2025 WHERE ADDITIONAL AR-3636H WAS USED, AND A BIRDBEAK WAS USED TO COMPLETE THE CASE.
Patient problems: Insufficient Information
MAUDE report 21959301 ↗MalfunctionMar 20, 2025Break; Break
ON (B)(6)2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-4068HL SWIFTSTITCH SUTURE PASSER WAS INTRODUCED TO THE HIP JOINT FOR AN ACETABULAR LABRAL REPAIR PROCEDURE USING 1.8 KNOTLESS FIBERTAK ANCHOR ON (B)(6)2025. THE REPAIR SUTURE WAS GRASPED BY THE SWIFTSTITCH, AND UPON PASSING THROUGH THE LABRUM, THE BASE OF THE SWIFTSTITCH NEEDLE BROKE. THE SWIFTSTITCH DID NOT PENETRATE THE ACETABULUM TO CAUSE THIS, AND THE SURGEON BELIEVED THIS WAS DUE TO A FAULTY DEVICE. THIS OCCURRED DURING USE, WITH NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 03/11/2025: ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THERE WAS NO CASE DELAY OR KNOWN ANESTHESIA ADMINISTERED TO THE PATIENT. A NEW AR-4068HL WAS OPENED TO COMPLETE THE CASE. NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21653892 ↗MalfunctionOct 9, 2024Use of Device Problem; Device Damaged by Another Device; Use of Device Problem; Device Damaged by Another Device
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-4068HL SWIFTSTITCH SUTURE PASSER FRAYED THE SUTURE. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION RECEIVED ON 9/19/2024: THIS WAS DISCOVERED DURING A HIP ARTHROSCOPY LABRAL REPAIR PROCEDURE ON 9/12/2024. THE SWIFTSTITCH SUTURE PASSER FRAYED AN AR-7216 FIBERWIRE, WHICH BROKE. EVERYTHING WAS REMOVED FROM INSIDE THE PATIENT. THE CASE WAS COMPLETED USING PRODUCTS FROM ANOTHER MANUFACTURER. IT WAS DELAYED ABOUT TWO MINUTES WITHOUT ADDITIONAL ANESTHESIA.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20408358 ↗MalfunctionSep 10, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2024, A SALES REPRESENTATIVE VIA SEMS-06642019 REPORTED THAT AN AR-4068HL SWIFTSTITCH HIP SUTURE PASSER HAD CUT THROUGH THE SUTURE NUMEROUS TIMES. NO ADVERSE EFFECTS ON THE PATIENT WERE REPORTED. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20185178 ↗MalfunctionJun 24, 2024Break; Break
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY THE TIP BROKE OFF DURING THE FIRST STITCH. THE BROKEN TIP WAS RETRIEVED FROM THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19595455 ↗MalfunctionDec 15, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 11/17/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-3636H SELF BUNCHING 1.8 KNOTLESS HIP FIBERTAK PULLED OUT OF THE BONE. THE KNOTLESS HIP FIBERTAK WAS INITIALLY INSERTED USING THE 20-DEGREE CURVED DRILL GUIDE AND THE NEW 1.8 FLEXIBLE FLUTED DRILL AR-3600ND-2HF. THE ANCHOR WAS INSERTED, AND THE REPAIR SUTURE WAS PASSED THROUGH THE LABRUM. THE REPAIR STITCH WAS CONVERTED INTO THE ANCHOR AND TENSIONED DOWN TO PULL THE LABRUM UP. WHEN TENSIONING THE REPAIR SUTURE, THE ANCHOR PULLED OUT OF THE BONE. ADDITIONALLY, WHEN USING AN AR-4068HL SWIFTSTITCH SUTURE PASSER SHREDDED THE REPAIR SUTURES UPON PASSING THROUGH. THE SWIFTSTITCH SUTURE PASSER WAS SWITCHED FOR ANOTHER ONE, AND THE PROCEDURE WAS CONTINUED WITHOUT FURTHER ISSUES. THIS WAS DISCOVERED DURING A HIP LABRAL REPAIR PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 11/22/2023: THE CASE WAS COMPLETED USING ANOTHER AR-3636H SELF BUNCHING 1.8 KNOTLESS HIP FIBERTAK. THE PATIENT HAD NORMAL BONE QUALITY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18340577 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page