Arthrex® AR-8730-60HRecalls & reported problems
Arthrex, Inc. · COMPR FT SCRW,3.5 MINI, 60mm · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867303508), received between Jul 27, 2021 and Jul 27, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2021Break
IT WAS REPORTED THAT DURING A METACARPAL FRACTURE, BOTH 3.5 AND 2.5 HEADLESS SCREWS BROKE WHILE BEING IMPLANTED. WHEN SCREW WAS IMPLANTED AND GOING PAST THE ISTHMUS OF THE METACARPAL THE HEADLESS SCREW BROKE. THE SURGEON ATTEMPTED TO REMOVE THE SCREW USING THE DRIVER AND BROKE THE 1.5 HEX DRIVER. THEN HAD TO USE THE SYNTHES SCREW REMOVAL WHICH DIDN¿T WORK AS WELL AND RESORTED TO HAVING TO BURR DOWN THE SCREW FLUSH TO THE METACARPAL. ADDITIONAL INFORMATION PROVIDED 7/7/21: THE LOT NUMBER OF THE SCREW'S WERE NOT LOCATED AS THE DEVICES WERE DISCARDED DURING THE CASE
Patient problems: Insufficient Information
MAUDE report 12231478 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.