Skip to content

Arthrex® AR-18714V-13Recalls & reported problems

Arthrex, Inc. · VA Locking Screw, 1.4x13mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867303652), received between Aug 18, 2023 and May 13, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Installation-Related Problem1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2024Installation-Related Problem; Installation-Related Problem

    ON 3/6/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (3) 1.4MM LOCKING SCREWS WOULD NOT LOCK INTO THE PLATE AND WOULD SPIN. THE SURGEON WAS USING AN EIGHT-HOLE TRIANGLE PLATE IMPLANTED IN THE PATIENT. THE SCREW WOULD NOT LOCK IN THE SHAFT AND ON THE TRIANGLE PORTION OF THE PLATE. WHEN THE SURGEON ATTEMPTED TO LOCK THE SCREW IN THE LOCKING HOLE AND THE SCREW WOULD SPIN; HOWEVER, THE SURGEON DRILLED A HOLE RIGHT NEXT TO THE PREVIOUS ONE, AND THE SCREW WOULD LOCK WITHOUT ANY ISSUES. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION RECEIVED ON 3/13/2023: THE THREE FAULTY SCREWS WERE AN AR-18714V-13 VA LOCKING SCREW, 1.4X13MM, AN AR-18714V-08 VA LOCKING SCREW, 1.4X8MM, AND AN AR-18714V-12 VA LOCKING SCREW, 1.4X12MM. ALL THREE WERE ATTEMPTED TO BE INSERTED AND WERE NOT IMPLANTED DUE TO SPINNING AND NOT LOCKING INTO AN AR-18714P-18 TRIANGLE PLATE, 1.4MM. THE CASE WAS COMPLETED BY USING TWO AR-18714V-12 VA LOCKING SCREW, 1.4X12MM, AN AR-18714V-06 VA LOCKING SCREW, 1.4X6MM, THREE AR-18714V-10 VA LOCKING SCREW, 1.4X10MM AND TWO AR-18714V-09 VA LOCKING SCREW, 1.4X9MM. THIS WAS DISCOVERED DURING A 4TH AND 5TH PROXIMAL PHALANX FRACTURE PROCEDURE ON (B)(6) 2023.

    Patient problems: Insufficient Information

    MAUDE report 19296633 ↗
  • MalfunctionAug 18, 2023Break; Break

    ON 07/21/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-18714-12 SCREW AND AN AR-18714V-13 SCREW BROKE ABOUT HALFWAY DOWN THE THREADS. THIS WAS DISCOVERED DURING A CASE DEVIATION LONGER THAN 15 MINUTES, SCREWS WERE UNABLE TO BE REMOVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17573787 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page