Arthrex® AR-18714P-21Recalls & reported problems

Arthrex, Inc. · Hook Plate, 1.4mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867303751), received between May 28, 2025 and May 28, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device1
  • Use of Device Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2025Use of Device Problem; Defective Device; Use of Device Problem; Defective Device

    ON 05/06/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE TIP OF THE AR-18700-29 DRIVER SHAFT BROKE OFF IN THE HEAD OF THE AR-18714V-06 VAL SCREW AS THE SURGEON WAS ADVANCING IT INTO THE AR-18714P-21 HOOK PLATE. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. THE AR-18700-29 DRIVER SHAFT WAS BEING USED WITH THE JEWELER'S KNOB HANDLE AR-18700H. THE SURGEON WANTED TO REPLACE THE SCREW, BUT THE ONLY WAY TO REMOVE IT WAS TO CUT THE TINES OFF THE AR-18714P-21 HOOK PLATE, REMOVE EACH TINE FROM THE PATIENT, AND ROTATE THE PLATE COUNTERCLOCKWISE TO BACK OUT THE LOCKING SCREW. THE CASE WAS COMPLETED, AND NO FURTHER DETAILS WERE PROVIDED. THIS WAS DISCOVERED DURING AN ORIF MIDDLE PHALANX FRACTURE PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 05/27/2025.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095933 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →