Skip to content

Arthrex® AR-8771-0225SRecalls & reported problems

Arthrex, Inc. · DynaNite® Comp Plate, 2-Hole STR, 25mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867307964), received between Apr 13, 2021 and Apr 13, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2021Break

    IT WAS REPORTED DURING A CC JOINT PROCEDURE THE AR-8771-0225S NITINOL DOUBLE COMPRESSION PLATE WAS ATTACHED TO BONE ON ONE SIDE WITH A BBTAK. THE OTHER SIDE WAS DRILLED WITH A 2.5 DRILL THROUGH VAL GUIDE FOR A 3.0 HYBRID KREULOCK SCREW. THE KREULOCK SCREW DID NOT COMPRESS THE PLATE TO BONE AND DISTRACTED. AS THE SCREW LOCKED INTO THE THREADS OF THE PLATE, THE ROUND HOLE SNAPPED CLEAN OFF THE PLATE, LEAVING THE PLATE UNUSABLE. THERE WAS NO HARM DONE TO THE PATIENT AND THE SCREW WAS EASILY REMOVED.

    Patient problems: Failure of Implant

    MAUDE report 11655733 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates