Skip to content

Arthrex® AR-18716-10Recalls & reported problems

Arthrex, Inc. · Low Profile Screw, 1.6 x 10mm, Cortical · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867313224), received between Dec 27, 2021 and Sep 9, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18716-11 AND A AR-18716-10 LOW PROFILE SCREWS BROKE WHILE BEING IMPLANTED. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004634 ↗
  • MalfunctionJul 14, 2023Break; Break

    ON (B)(6)2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT THE HEAD FROM AN AR-18716-10 LOW PROFILE CORTICAL SCREW BROKE OFF. THE SURGEON REMOVED ALL THE BROKEN FRAGMENTS AND USED A NEW ONE TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING A METACARPAL FRACTURE PROCEDURE ON (B)(6)2023. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 17329648 ↗
  • MalfunctionDec 27, 2021Break

    ON (B)(6) 2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT DURING A FOOT REPAIR TWO AR-18716-10 SCREW HEADS AND ONE AR-18716-15 BROKE OFF DURING INSERTION. ALL FRAGMENTS WERE RETRIEVED BUT THE SCREWS REMAINED IN THE PATIENT HOLDING THE PLATE AMONG THE OTHER SCREWS. THE DOCTOR DETERMINED THE FIXATION WAS SUFFICIENT. THE PATIENT IS FINE TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13087499 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page