Arthrex® AR-18716-11Recalls & reported problems

Arthrex, Inc. · Low Profile Screw, 1.6 x 11mm, Cortical · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867313231), received between Mar 12, 2021 and Sep 9, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Defective Device1

Patient problems

  • Foreign Body In Patient4
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-07039383 THAT AN AR-18716-11 AND A AR-18716-10 LOW PROFILE SCREWS BROKE WHILE BEING IMPLANTED. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004612 ↗
  • MalfunctionAug 5, 2025Break; Defective Device; Break; Defective Device

    ON 07/16/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-18716-11 CORTICAL SCREW BROKE DURING REMOVAL. AFTER PLACING ALL SCREWS AND A PLATE, THE CONSTRUCT WAS EVALUATED VIA X-RAY, AND IT WAS DETERMINED THAT THE 11 MM SCREW WAS TOO LONG. THE SURGEON ATTEMPTED TO BACK OUT THE SCREW TO REPLACE IT WITH A SHORTER ONE, BUT THE SCREW FRACTURED DURING REMOVAL. THE BREAK OCCURRED BELOW THE PLATE, MAKING REMOVAL NOT FEASIBLE. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED ON 07/23/2025: THIS OCCURRED DURING A METACARPAL ORIF PROCEDURE. AN AR-18716P-18 T-PLATE (LOT 105642321) WAS SUCCESSFULLY IMPLANTED. NO DAMAGE HAPPENED TO THE PLATE, AND NO SURGICAL INSTRUMENTS BROKE DURING REMOVAL. HOWEVER, ONE AR-18716-11 CORTICAL SCREW FRACTURED AND REMAINED IN THE BONE. THE CORRECT DRIVER WAS USED CONSISTENTLY THROUGHOUT THE PROCEDURE AND FUNCTIONED PROPERLY ON THE REMAINING SIX SCREWS. THE LOT NUMBER FOR THE BROKEN SCREW IS UNKNOWN. THERE WAS NO CHANGE IN SURGICAL TECHNIQUE, AND NO ADDITIONAL PRODUCTS WERE USED TO COMPLETE THE CASE. A DELAY OF APPROXIMATELY FIVE MINUTES OCCURRED, BUT NO EXTRA ANESTHESIA WAS ADMINISTERED. THE AR-18700-12 DRILL BIT WAS USED TO PREP

    Patient problems: Foreign Body In Patient

    MAUDE report 22719822 ↗
  • MalfunctionJun 3, 2025Break; Break

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN (QTY. (B)(4)) OF AN AR-18716-11 LOW PROFILE CORTICAL SCREW AND AN AR-18716V-12 VAL SCREW BROKE OFF DURING IMPLANTATION OF THE SCREW INTO A PLATE. EACH SCREW BROKE AT THE NECK WHILE THEY WERE SCREWING INTO THE PLATE, LEAVING THE SHAFT OF THE SCREW IN THE BONE, AS REMOVAL COULD NOT BE EXTRACTED. THIS WAS DISCOVERED DURING A METACARPAL FRACTURE ORIF PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON (B)(6) 2025.

    Patient problems: Foreign Body In Patient

    MAUDE report 22133652 ↗
  • MalfunctionJun 3, 2025Break; Break

    ON 05/12/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN (QTY. 2) OF AN AR-18716-11 LOW PROFILE CORTICAL SCREW AND AN AR-18716V-12 VAL SCREW BROKE OFF DURING IMPLANTATION OF THE SCREW INTO A PLATE. EACH SCREW BROKE AT THE NECK WHILE THEY WERE SCREWING INTO THE PLATE, LEAVING THE SHAFT OF THE SCREW IN THE BONE, AS REMOVAL COULD NOT BE EXTRACTED. THIS WAS DISCOVERED DURING A METACARPAL FRACTURE ORIF PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 06/02/2025.

    Patient problems: Foreign Body In Patient

    MAUDE report 22133158 ↗
  • MalfunctionNov 30, 2021Break

    ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-18716-08, AN AR-18716-09, AND AN AR-18716-11 CORTICAL SCREWS BROKE ON THE FILE TIGHTENING OF THE SCREW. THIS WAS DISCOVERED DURING A PROCEDURE, THE SCREWS REMAIN INSIDE THE PATIENT. ADDITIONAL INFORMATION REQUESTED. AFTER RECEIVING ADDITIONAL INFORMATION ON (B)(6) 2021, SALES REPRESENTATIVE HAS CONFIRMED THAT DURING THE METACARPAL FRACTURE PROCEDURE, THREE CORTICAL SCREW BROKE AND REMAINED IN PATIENT. SURGEON USED A 1.2MM DRILL BIT TO PREPARE BONE TUNNELS AND PUT IN NEW SCREWS IN OTHER HOLES OF THE PLATE.

    Patient problems: Foreign Body In Patient

    MAUDE report 12897327 ↗
  • MalfunctionMar 12, 2021Break

    IT WAS REPORTED BY THE SALES REP, THAT DURING A METACARPAL FRACTURE, THE AR-18716-11 BROKE UPON INSERTION. ADDITIONAL INFORMATION PROVIDED 2/26/21: ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. THE CORRESPONDING DRILL BIT FOR THE 1.6MM SCREWS WAS USED PRIOR TO INSERTION (1.2MM DRILL BIT). THE PATIENT WAS IN UPPER 20S WITH HEALTHY BONE QUALITY. AFTER REMOVAL OF THE BROKEN SCREW, A NEW SCREW OF SIMILAR SIZE WAS USED TO COMPLETE PLATE FIXATION.

    Patient problems: Failure of Implant

    MAUDE report 11475260 ↗

Other Plate, fixation, bone devices