Skip to content

Arthrex® AR-18716-13Recalls & reported problems

Arthrex, Inc. · Low Profile Screw, 1.6 x 13mm, Cortical · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867313255), received between Jul 1, 2020 and Aug 8, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component1

Patient problems

  • Insufficient Information2
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2025Break; Break

    ON 07/22/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-18716-13 CORTICAL SCREW BROKE. THIS OCCURRED DURING A CLAVICLE ORIF PROCEDURE ON (B)(6) 2025. THE FRAGMENTS WERE RETRIEVED. THE PATIENT HAD A NORMAL BONE QUALITY. THERE WAS NO DELAY IN THE PROCEDURE, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE CASE WAS COMPLETED USING A NON-ARTHREX DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 22750240 ↗
  • MalfunctionAug 8, 2025Break; Break

    ON 07/22/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-18716-13 CORTICAL SCREW BROKE. THIS OCCURRED DURING A CLAVICLE ORIF PROCEDURE ON (B)(6) 2025. THE FRAGMENTS WERE RETRIEVED. THE PATIENT HAD A NORMAL BONE QUALITY. THERE WAS NO DELAY IN THE PROCEDURE, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE CASE WAS COMPLETED USING A NON-ARTHREX DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 22750153 ↗
  • MalfunctionAug 31, 2023Break; Break

    ON 8/4/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-18716-13, AN AR-18716-12, CORTICAL LOW-PROFILE SCREWS, AND AN AR-18716V-13 VAL SCREW BROKE DURING INSERTION. ALL THREE SCREWS HEAD BROKE OFF AND WERE RETRIEVED; THE BODY OF THE SCREWS REMAIN IMPLANTED IN THE PATIENT. THIS WAS DISCOVERED DURING A FINGER FRACTURE PROCEDURE ON (B)(6) 2023.

    Patient problems: Foreign Body In Patient

    MAUDE report 17662611 ↗
  • InjuryJul 1, 2020Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A MINI CFS CASE ONE OF THE 1.6MM X 13MM CORTICAL SCREWS, AR-1816-06-13, WAS INSERTED AND WAS A TAD TOO LONG. WHEN THE SURGEON WENT TO REMOVE IT THE HEAD OF THE SCREW BROKE OFF. BROKEN PIECE WILL BE RETURNED FOR EVALUATION. TWO SCREWS WERE USED IN THE CASE, LOT 1065053 AND LOT 2684661. IT IS UNKNOWN WHICH OF THE TWO WAS THE ONE THAT BROKE. THE BODY OF THE BROKEN SCREW, APPROXIMATELY 11-12 MM, WAS LEFT IN THE PATIENT. THE SURGEON HAD DRILLED THE APPROPRIATE SCREW HOLE AS PREP FOR INSERTION. THE APPROPRIATE SELF RETAINING DRIVER WAS USED FOR INSERTION. THE FRACTURE WAS SUCCESSFULLY REDUCED AND COMPLETED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10217878 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →