Arthrex® AR-18716-13Recalls & reported problems
Arthrex, Inc. · Low Profile Screw, 1.6 x 13mm, Cortical · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867313255), received between Jul 1, 2020 and Aug 8, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Detachment of Device or Device Component1
Patient problems
- Insufficient Information2
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2025Break; Break
ON 07/22/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-18716-13 CORTICAL SCREW BROKE. THIS OCCURRED DURING A CLAVICLE ORIF PROCEDURE ON (B)(6) 2025. THE FRAGMENTS WERE RETRIEVED. THE PATIENT HAD A NORMAL BONE QUALITY. THERE WAS NO DELAY IN THE PROCEDURE, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE CASE WAS COMPLETED USING A NON-ARTHREX DEVICE.
Patient problems: Insufficient Information
MAUDE report 22750240 ↗MalfunctionAug 8, 2025Break; Break
ON 07/22/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-18716-13 CORTICAL SCREW BROKE. THIS OCCURRED DURING A CLAVICLE ORIF PROCEDURE ON (B)(6) 2025. THE FRAGMENTS WERE RETRIEVED. THE PATIENT HAD A NORMAL BONE QUALITY. THERE WAS NO DELAY IN THE PROCEDURE, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE CASE WAS COMPLETED USING A NON-ARTHREX DEVICE.
Patient problems: Insufficient Information
MAUDE report 22750153 ↗MalfunctionAug 31, 2023Break; Break
ON 8/4/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-18716-13, AN AR-18716-12, CORTICAL LOW-PROFILE SCREWS, AND AN AR-18716V-13 VAL SCREW BROKE DURING INSERTION. ALL THREE SCREWS HEAD BROKE OFF AND WERE RETRIEVED; THE BODY OF THE SCREWS REMAIN IMPLANTED IN THE PATIENT. THIS WAS DISCOVERED DURING A FINGER FRACTURE PROCEDURE ON (B)(6) 2023.
Patient problems: Foreign Body In Patient
MAUDE report 17662611 ↗InjuryJul 1, 2020Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A MINI CFS CASE ONE OF THE 1.6MM X 13MM CORTICAL SCREWS, AR-1816-06-13, WAS INSERTED AND WAS A TAD TOO LONG. WHEN THE SURGEON WENT TO REMOVE IT THE HEAD OF THE SCREW BROKE OFF. BROKEN PIECE WILL BE RETURNED FOR EVALUATION. TWO SCREWS WERE USED IN THE CASE, LOT 1065053 AND LOT 2684661. IT IS UNKNOWN WHICH OF THE TWO WAS THE ONE THAT BROKE. THE BODY OF THE BROKEN SCREW, APPROXIMATELY 11-12 MM, WAS LEFT IN THE PATIENT. THE SURGEON HAD DRILLED THE APPROPRIATE SCREW HOLE AS PREP FOR INSERTION. THE APPROPRIATE SELF RETAINING DRIVER WAS USED FOR INSERTION. THE FRACTURE WAS SUCCESSFULLY REDUCED AND COMPLETED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10217878 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page