Arthrex® AR-1547BC-CPRecalls & reported problems

Arthrex, Inc. · FDL Implant System, 4.75 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867315150), received between May 12, 2025 and Aug 5, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2025Break; Failure to Advance; Break; Failure to Advance

    ON 07/16/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT DURING AN FHL TRANSFER PROCEDURE PERFORMED ON (B)(6) 2025, AN AR-1547BC-CP FDL IMPLANT SYSTEM EXPERIENCED A MALFUNCTION. SPECIFICALLY, THE 4.75 MM TENODESIS SCREW BROKE IN HALF AT ITS TOP WHILE BEING INSERTED HALFWAY. ALL BROKEN FRAGMENTS WERE RETRIEVED FROM THE PATIENT. TO COMPLETE THE PROCEDURE, THEY UTILIZED THE 5 MM REAMER INCLUDED IN THE KIT TO REAM OVER THE SITE AND SUCCESSFULLY IMPLANTED A 6.25 MM TENODESIS SCREW WITHOUT FURTHER ISSUES. THE CASE WAS DELAYED BY 15 MINUTES, WITH NO ADDITIONAL ANESTHESIA REQUIRED. NO ADVERSE EFFECTS WERE REPORTED DURING OR AFTER THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22720614 ↗
  • MalfunctionMay 12, 2025Break; Break

    A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-1547BC-CP FDL IMPLANT SYSTEM HAD AN ISSUE DURING A FDL TRANSFER PROCEDURE. WHEN THE SURGEON WENT TO PUT IN THE SCREW AND TURNED IT CLOCKWISE, THE SCREW WAS DELAMINATING AND DISINTEGRATING UPON INSERTION, AND BROKE. THERE WAS BIOCOMPOSITE DEBRIS PRODUCED INSIDE THE PATIENT, THE BROKEN SCREW AND ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. ANOTHER AR-1547BC-CP FDL IMPLANT SYSTEM WITH A DIFFERENT LOT NUMBER, 15332204, WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY; THERE WAS NO HARM TO THE PATIENT. THE COMPLAINT DEVICE WAS DISCARDED, BUT A PICTURE WAS TAKEN. THERE WAS A 2-MINUTE CASE DELAY, AND IT COULD NOT BE CONFIRMED IF ADDITIONAL ANESTHESIA WAS ADMINISTERED. THIS OCCURRED DURING USE WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22004324 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates