Skip to content

Arthrex® AR-18714P-07Recalls & reported problems

Arthrex, Inc. · Straight Plate, 1.4mm, 10 Hole · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867315556), received between Feb 16, 2026 and Feb 16, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2026Device Damaged by Another Device

    ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-18714-12 CORTICAL SCREW (1.4 MM × 12 MM) EXPERIENCED A HEAD FRACTURE DURING INSERTION. THE SCREW WAS BEING PLACED INTO A HOLE OF A 10-HOLE 1.4 MM PLATE AND HAD BEEN ADVANCED NEARLY FLUSH TO THE PLATE WHEN THE HEAD SEPARATED FROM THE SHAFT. THE SURGEON REMOVED ONLY THE SCREW HEAD, LEAVING THE SCREW SHAFT IN SITU BECAUSE IT CROSSED THE FRACTURE SITE. THE EVENT OCCURRED DURING A METACARPAL FRACTURE INTERNAL FIXATION PROCEDURE ON (B)(6) 2026. THE ISSUE DID NOT CAUSE PROCEDURAL DELAY, AND NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 03-FEB-2026: THERE WAS NO REPORTED DAMAGE TO THE AR-18714P-07 STRAIGHT PLATE, 1.4 MM, 10-HOLE ASSOCIATED WITH THIS EVENT. THE ONLY ISSUE IDENTIFIED INVOLVED A SINGLE AR-18714-12 CORTICAL SCREW (1.4 MM × 12 MM), AS SHOWN IN THE ATTACHED PHOTO. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THE AFFECTED SCREW REMAINED IMPLANTED. ADDITIONAL SCREWS MEASURING 10 MM, 11 MM, 12 MM, AND 15 MM IN LENGTH WERE ALSO UTILIZED AS PART OF THE CONSTRUCT. NO ADVERSE PATIENT EFFECTS WERE REPORTED IN RELATION TO THIS ISSUE.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24344277 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page