Skip to content

Arthrex® AR-8980DSRecalls & reported problems

Arthrex, Inc. · Disposable Kit for DX Swivelock, 4.75 mm · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867319851), received between Nov 22, 2023 and Nov 22, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Component Misassembled” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Misassembled1
  • Patient Device Interaction Problem1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 22, 2023Patient Device Interaction Problem; Component Misassembled; Patient Device Interaction Problem; Component Misassembled

    ON 31ST OCTOBER 2023, IT WAS REPORTED BY A SALES REP. VIA EMAIL THAT AN AR-8980DS, DISPOSABLE KIT FOR DX SWIVELOCK®, 4.75MM, WAS FAULTY AS IT HAD A COMPLETELY SQUARED TIP INSTEAD OF A POINTY TIP, WAS NOT DRILLING THROUGH BONE AND WAS CAUSING HEAT BUILDUP AND SOME BURNING OF BONE EDGES THROUGH FRICTION. THIS WAS DISCOVERED DURING USE IN A ACHILLES RECONSTRUCTION ON 27TH OCTOBER 2023. THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DRILL KIT WITH THE SAME LOT NUMBER.

    Patient problems: Burn(s)

    MAUDE report 18192005 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page