Skip to content

Arthrex® AR-8980BCRecalls & reported problems

Arthrex, Inc. · DX Knotless SwiveLock, BC, 4.75 mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867319868), received between Apr 10, 2024 and Sep 20, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2

Patient problems

  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 20, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-8980BC DX KNOTLESS SWIVELOCK, AN AR-8913DS MINI TIGHTROPE, AND AN AR-7237-7 FIBERTAPE COULD NOT BE EASILY IMPLANTED. THE PATIENT'S BONE WAS HARD, SO THE SURGEON TAPPED THE IMPLANTS IN WITH A MALLET TO PLACE THEM IN THE TUNNEL. THIS WAS DISCOVERED DURING A LISFRANC PROCEDURE ON (B)(6) 2024. ONE TO TWO WEEKS LATER, ON (B)(6) 2024, THE PATIENT UNDERWENT REVISION SURGERY TO HAVE ALL THE PRODUCTS EXPLANTED. THE AR-8980BC DX KNOTLESS SWIVELOCK ANCHOR HAD PULLED OUT OF THE MEDIAL CUNEIFORM BONE. THE LISFRANC LIGAMENT PLASTY PROCEDURE WAS COMPLETED USING AN AR-8913DS MINI TIGHTROPE FROM LOT NUMBER 15242118, AN AR-7237-7 FIBERTAPE FROM LOT NUMBER 15081694, AND AN AR-8922 SUTURE BUTTON FROM LOT NUMBER 15166049.

    Patient problems: Failure of Implant

    MAUDE report 20268246 ↗
  • MalfunctionApr 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8980BC DX KNOTLESS SWIVELOCK ANCHOR SLID INTO THE SHAFT AND PULLED OUT WHEN SHAFT WAS REMOVED FROM THE PATIENT. SURGEON TAPPED AGAIN AND USED AN AR-2324PSLC PEEK-SWIVELOCK, WHICH THE ANCHOR ALSO SLID INTO THE DRIVER SHAFT AND PULLED OUT WHEN IT WAS REMOVED. FINALLY, AN AR-1540PS PEEK TENODESIS SCREW WAS USED TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING A PATELLA TENDON REPAIR PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 19080693 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page