Skip to content

Arthrex® AR-8980BC-1Recalls & reported problems

Arthrex, Inc. · Sutr Anch, DX SwvLk, BC, 1/pack, 4.7 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867319875), received between Sep 15, 2025 and Sep 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Material Deformation1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2025Device Dislodged or Dislocated; Material Deformation; Device Dislodged or Dislocated; Material Deformation

    ON 8/26/2025, AN EMPLOYEE FROM AN ARTHREX SUBSIDIARY REPORTED VIA EMAIL AN ISSUE INVOLVING THE AR-8980BC-1 SWIVELOCK DX DURING A PL TENDON FIXATION PROCEDURE ON (B)(6) 2025. AFTER THE DEVICE WAS INSERTED, IT DISLODGED FROM THE BONE SOCKET WHILE THE EYELET SUTURE WAS BEING REMOVED. THE TIP OF THE SCREW WAS FOUND TO BE DEFORMED AND DAMAGED, RENDERING IT UNUSABLE. DESPITE THE COMPLICATION, THE TAPE AND PL TENDON WERE SUCCESSFULLY SUTURED TO THE JOINT CAPSULE, AND THE OPERATION WAS COMPLETED WITHOUT FURTHER INCIDENT AS THE PATIENT¿S CONDITION STABILIZED. NO SIGNIFICANT SURGICAL DELAY WAS REPORTED, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE BONE SOCKET HAD BEEN DRILLED USING A 2.4 GUIDE PIN AND A 4.5 HOLLOW DRILL, WITHOUT THE USE OF A TAP. BONE QUALITY WAS NOTED TO BE HARD.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23053989 ↗

Other Fastener, fixation, biodegradable, soft tissue devices