Arthrex® AR-8835CL-14Recalls & reported problems

Arthrex, Inc. · KreuLock Compression Screw, SS, 3.5x14mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867323162), received between Feb 2, 2024 and May 12, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Use of Device Problem2

Patient problems

  • Failure of Implant4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 12, 2025Use of Device Problem; Use of Device Problem

    ON 3/31/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT A PATIENT'S CLAVICLE FRACTURE PLATE BROKE. NO FURTHER INFORMATION WAS PROVIDED. ON (B)(6) 2025, A PATIENT¿S LEGAL REPRESENTATIVE REPORTED VIA EMAIL THAT A CLAVICLE PLATE FRACTURED. THE PLATE WAS IMPLANTED DURING AN OPEN REDUCTION INTERNAL FIXATION OF THE RIGHT CLAVICLE PERFORMED ON (B)(6) 2024. ON (B)(6) 2024, IMAGING SHOWED A BENT CLAVICLE PLATE. ON (B)(6) 2024, IMAGING SHOWED A FRACTURED CLAVICLE PLATE. ON (B)(6) 2025, THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE HARDWARE. THE RECORDS INDICATED PATIENT NONCOMPLIANCE WITH POSTOPERATIVE PROTOCOL CONTRARY TO THE CLAVICLE PLATE'S DIRECTIONS FOR USE.

    Patient problems: Failure of Implant

    MAUDE report 22002529 ↗
  • InjuryMay 12, 2025Use of Device Problem; Use of Device Problem

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT A PATIENT'S CLAVICLE FRACTURE PLATE BROKE. NO FURTHER INFORMATION WAS PROVIDED. ON (B)(6) 2025, A PATIENT¿S LEGAL REPRESENTATIVE REPORTED VIA EMAIL THAT A CLAVICLE PLATE FRACTURED. THE PLATE WAS IMPLANTED DURING AN OPEN REDUCTION INTERNAL FIXATION OF THE RIGHT CLAVICLE PERFORMED ON (B)(6) 2024. ON (B)(6) 2024, IMAGING SHOWED A BENT CLAVICLE PLATE. ON (B)(6) 2024, IMAGING SHOWED A FRACTURED CLAVICLE PLATE. ON (B)(6) 2025, THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE HARDWARE. THE RECORDS INDICATED PATIENT NONCOMPLIANCE WITH POSTOPERATIVE PROTOCOL CONTRARY TO THE CLAVICLE PLATE'S DIRECTIONS FOR USE.

    Patient problems: Failure of Implant

    MAUDE report 22002433 ↗
  • InjuryFeb 2, 2024Break; Break

    ON 01/23/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A PATIENT UNDERWENT A CLAVICLE FRACTURE PROCEDURE ON (B)(6) 2023 WHERE AN AR-2652CL CLAVICLE FRACTURE PLATE, AN AR-8835-14 LOW PROFILE SCREW, (2) AR-8835-16 LOW PROFILE SCREWS, AN AR-8835CL-12 KREULOCK COMPRESSION SCREW, (2) AR-8835CL-14 KREULOCK COMPRESSION SCREWS, AN AR-8835-10 LOW PROFILE SCREW, AN AR-7268 FIBERTAPE® CERCLAGE WITH TIGERLINK, AND AN AR-7268T TIGERTAPE® CERCLAGE WITH FIBERLINK WERE IMPLANTED. FIVE WEEKS POST-OP, THE PATIENT REPORTED HEARING OR FEELING A POP. AN X-RAY WAS TAKEN THAT CONFIRMED THAT THE AR-2652CL PLATE HAD BROKEN. ON (B)(6) 2024, A REVISION SURGERY WAS PERFORMED WHERE THE IMPLANTED DEVICES WERE REMOVED. THE REVISION WAS COMPLETED USING A DEVICE FROM ANOTHER MANUFACTURER.

    Patient problems: Failure of Implant

    MAUDE report 18636096 ↗
  • InjuryFeb 2, 2024Break; Break

    ON 01/23/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A PATIENT UNDERWENT A CLAVICLE FRACTURE PROCEDURE ON (B)(6) 2023 WHERE AN AR-2652CL CLAVICLE FRACTURE PLATE, AN AR-8835-14 LOW PROFILE SCREW, (2) AR-8835-16 LOW PROFILE SCREWS, AN AR-8835CL-12 KREULOCK COMPRESSION SCREW, (2) AR-8835CL-14 KREULOCK COMPRESSION SCREWS, AN AR-8835-10 LOW PROFILE SCREW, AN AR-7268 FIBERTAPE® CERCLAGE WITH TIGERLINK, AND AN AR-7268T TIGERTAPE® CERCLAGE WITH FIBERLINK WERE IMPLANTED. FIVE WEEKS POST-OP, THE PATIENT REPORTED HEARING OR FEELING A POP. AN X-RAY WAS TAKEN THAT CONFIRMED THAT THE AR-2652CL PLATE HAD BROKEN. ON (B)(6) 2024, A REVISION SURGERY WAS PERFORMED WHERE THE IMPLANTED DEVICES WERE REMOVED. THE REVISION WAS COMPLETED USING A DEVICE FROM ANOTHER MANUFACTURER.

    Patient problems: Failure of Implant

    MAUDE report 18635465 ↗

Other Plate, fixation, bone devices