Skip to content

Arthrex® ABS-8981-12SRecalls & reported problems

Arthrex, Inc. · 12MM SINGLE USE OATS DISPOSABLE KIT · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867343894), received between Sep 17, 2024 and Sep 17, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Dull, Blunt2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024Dull, Blunt; Dull, Blunt

    ON 09/03/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN ABS-8981-12S OATS 2.0 SINGLE-USE SET WAS RECEIVED WITH A DULL TIP ON THE HANDLE AND NOT SHARP, AND THE INTERNAL PLUNGE WAS STUCK/TOO TIGHT. THEN, ANOTHER ABS-8981-12S OATS 2.0 SINGLE-USE SET FROM THE SAME LOT WAS OPENED, AND THE SAME ISSUE OCCURRED. THE CASE WAS COMPLETED SUCCESSFULLY USING FORCE AND A BIGGER MALLET TO MALLET THE DISPOSABLE TO DOWNSIZE THE ALLOGRAFT. WHEN FORCE WAS APPLIED THROUGH THE GRAFT, THEY WERE CONCERNED THAT THE GRAFT WOULD BREAK OR THE DEVICE WOULD GO THROUGH. THERE WAS NO IMPACT ON THE PATIENT. THIS WAS DISCOVERED DURING AN OAT PROCEDURE ON THE ELBOW PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238218 ↗
  • MalfunctionSep 17, 2024Dull, Blunt; Dull, Blunt

    ON 09/03/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT (QTY. 2) OF AN ABS-8981-12S OATS 2.0 SINGLE-USE SET WAS RECEIVED WITH A DULL TIP ON THE HANDLE AND NOT SHARP, AND THE INTERNAL PLUNGE WAS STUCK/TOO TIGHT. THEN, ANOTHER ABS-8981-12S OATS 2.0 SINGLE-USE SET FROM THE SAME LOT WAS OPENED, AND THE SAME ISSUE OCCURRED. THE CASE WAS COMPLETED SUCCESSFULLY USING FORCE AND A BIGGER MALLET TO MALLET THE DISPOSABLE TO DOWNSIZE THE ALLOGRAFT. WHEN FORCE WAS APPLIED THROUGH THE GRAFT, THEY WERE CONCERNED THAT THE GRAFT WOULD BREAK OR THE DEVICE WOULD GO THROUGH. THERE WAS NO IMPACT ON THE PATIENT. THIS WAS DISCOVERED DURING AN OAT PROCEDURE ON THE ELBOW PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238003 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page