Arthrex® AR-2386-09Recalls & reported problems
Arthrex, Inc. · QuadPro™ Harvester, 9 MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00888867344662), received between Feb 6, 2023 and May 20, 2025 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Defective Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device3
- Dull, Blunt2
- Failure to Cut2
- Physical Resistance/Sticking2
- Use of Device Problem2
- Break1
- Improper or Incorrect Procedure or Method1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information2
- Unspecified Tissue Injury1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2025Failure to Cut; Failure to Cut
ADDITIONAL INFORMATION HAS BEEN RECEIVED ON 5/20/2025: AN AUTOGRAFT QUAD TENDON ACL RECONSTRUCTION PROCEDURE WAS PERFORMED. DURING THE PROCEDURE, A DELAY OF APPROXIMATELY 10 MINUTES WAS REPORTED DUE TO IDENTIFIED ISSUES. HOWEVER, NO ADDITIONAL ANESTHESIA WAS ADMINISTERED BEYOND THE EXPECTED DURATION, WITH THE PATIENT REMAINING UNDER FOR AN ADDITIONAL 10 MINUTES. NO SURGICAL INSTRUMENTS OR IMPLANTS BROKE, AND NO FRAGMENTS WERE ENCOUNTERED DURING THE PROCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22052517 ↗MalfunctionMay 19, 2025Failure to Cut; Physical Resistance/Sticking; Failure to Cut; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE SURGEON WAS NOT ABLE TO AMPUTATE THE GRAFT USING THE QUADPRO DEVICE AR-2386-09 (LOT: 23F22). THE TIP OF THE STRIPPER WAS BLOCKING IN THE TRANSPARENT TUBE AND WAS NOT AMPUTATING THE QUAD TENDON. THE SURGEON HAD TO FORCE THE DEVICE AND CUT THE SMALLER LENGTH THEN NEEDED. HE HAD TO CHANGE THE SURGICAL TECHNIQUE AND HARVESTED THE HAMSTRING TENDONS INSTEAD. TWO TENDONS WAS HARVESTED ON THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-10300F). IT WAS NOT NECESSARY TO DO A SECOND SURGERY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22043894 ↗MalfunctionApr 22, 2025Use of Device Problem; Use of Device Problem
ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-2386-09 QUADPRO¿ HARVESTER MALFUNCTIONED DURING A CASE ON (B)(6) 2025. THERE WAS NO IMPACT ON THE PATIENT OR CASE AS THEY HAD TWO HARVESTERS PRESENT FOR THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21883994 ↗MalfunctionFeb 18, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 01/30/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-2386-09 QUADPRO¿ HARVESTER DID NOT FULLY AMPUTATE THE GRAFT. THIS OCCURRED DURING AN ACL RECONSTRUCTION ON (B)(6) 2025 WHEN THE 9MM QUADPRO HARVESTER DID NOT FULLY AMPUTATE THE GRAFT. THE PUSH ROD BECAME STUCK INSIDE THE HARVESTER AND HAD TO BE FORCED APART USING A MALLET AND OSTEOTOME. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21392181 ↗MalfunctionApr 25, 2024Defective Device; Defective Device
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT (2) AR-2386-09 QUADPRO¿ HARVESTERS WERE NOT WORKING. THIS OCCURRED DURING A CASE WITH NO EFFECT ON THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19188317 ↗MalfunctionApr 25, 2024Defective Device; Defective Device
ON 04/02/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-2386-09 QUADPRO¿ HARVESTERS WERE NOT WORKING. THIS OCCURRED DURING A CASE WITH NO EFFECT ON THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19185263 ↗MalfunctionApr 19, 2024Defective Device; Defective Device
ON (B)6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-2386-09 QUADPRO¿ HARVESTER WOULD NOT HARVEST THE TENDON. THE DEVICE WAS BRAND NEW AND OUT OF THE BOX. THIS OCCURRED DURING A CASE WHERE ANOTHER DEVICE WAS OPENED, AND THE SAME OUTCOME. THE PATIENT WAS IN THE ROOM, BUT NO ADVERSE EVENT. NOTHING BROKE, AND THE PATIENT IS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19144853 ↗MalfunctionApr 18, 2024Dull, Blunt; Dull, Blunt
ON 03/21/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT QTY. 2 OF AN AR-2386-09 QUADPRO HARVESTER, 9 MM WOULD NOT HARVEST THE TENDON, BRAND NEW OUT OF BOX. THE SALES REPRESENTATIVE STATED THAT THEY TRIED THE SAME PART AND LOT NUMBER WITH THE SAME OUTCOME. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19139371 ↗MalfunctionApr 17, 2024Dull, Blunt; Dull, Blunt
ON 03/21/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT QTY. 2 OF AN AR-2386-09 QUADPRO HARVESTER, 9 MM WOULD NOT HARVEST THE TENDON, BRAND NEW OUT OF BOX. THE SALES REPRESENTATIVE STATED THAT THEY TRIED THE SAME PART AND LOT NUMBER WITH THE SAME OUTCOME. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19126705 ↗MalfunctionOct 24, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY, THE TIP, STUB AND HANDLE OF THE DEVICE BROKE OFF WHILE SCREWING IN. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. NO FURTHER INFORMATION WAS PROVIDED. (B)(6) 2023. IT WAS REPORTED THAT DURING A KNEE SURGERY, THE HANDLE OF THE DEVICE BROKE OFF WHILE SCREWING IN. THE SURGEON STATED THAT THIS MIGHT HAVE BEEN RELATED TO THE BLUNT TIP. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Insufficient Information
MAUDE report 17995550 ↗MalfunctionJul 24, 2023Loose or Intermittent Connection; Material Separation; Loose or Intermittent Connection; Material Separation
ON 6/27/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-2386-09 QUAD PRO HARVESTER HAD AN ISSUE DURING AN ACL PROCEDURE ON (B)(6) 2023. THE QUAD HARVESTER TIP SEEMED LOOSE ON THE HARVESTING DEVICE. A SECOND DEVICE WAS OPENED TO COMPLETE THE CASE. THE DEVICE WAS DISCARDED BY THE FACILITY. THIS OCCURRED DURING USE WITH NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED. ON 7/6/2023, THE SALES REPRESENTATIVE STATED THE FOLLOWING INFORMATION OVER THE PHONE: THIS OCCURRED DURING THE USE OF THE DEVICE. WHEN THE SURGEON INSERTED THE QUAD PRO HARVESTER, THE METAL CUTTING BLADE WAS ROCKING/WIGGLING WITHIN THE PLASTIC CANNULA. THE DEVICE WAS REMOVED IN ONE PIECE FROM THE PATIENT, NOTHING BROKE OFF. A SECOND DEVICE WITH AN UNKNOWN LOT NUMBER WAS USED TO COMPLETE THE CASE SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17380086 ↗InjuryFeb 6, 2023Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem
IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-2386-09 QUAD PRO HARVESTER WAS MISUSED BY THE SURGEON. SURGEON PREMATURELY SEVERED TENDON. WAS DOING BONE BLOCK TECHNIQUE. LIKELY POINTING TO ANTERIOR. PATIENT BONE BLOCK AND TRUNCATED TENDON WAS STITCHED BACK INTO PATIENT KNEE AND SURGEON CONVERTED TO HAMSTRING TENDON HARVEST. CASE INVOLVEMENT, NO PATIENT EFFECT. ADDITIONAL INFORMATION REQUESTED ON (B)(6) 2022, RESPONSE (B)(6) 2022. PATIENT WAS NOT KEPT UNDER ADDITIONAL ANESTHESIA TO COMPLETE THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury
MAUDE report 16303264 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026