Arthrex® AR-1941PSVRecalls & reported problems

Arthrex, Inc. · PEEK, 3.9mm Knotless Corkscrew w/ #5 MTS · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867344921), received between Jan 19, 2021 and Apr 26, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Crack1
  • Device Dislodged or Dislocated1

Patient problems

  • Insufficient Information2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 09/09/2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1941PSV 3.9 MM KNOTLESS CORKSCREW®FAILED. THE TEAM USED A REUSABLE GUIDE AND PUNCH AND FOLLOWED PROPER TECHNIQUE FOR INSERTION. AFTER THE SURGEON PASSED THE REPAIR STITCH FROM HER FIRST ANCHOR, THEY HEARD A SOUND WHILE SHE WAS TENSIONING AND REALIZED THE ANCHOR PULLED OUT. SHE THEN REMOVED THE PROTRUDING ANCHOR, CUT OUT THE SUTURE, AND LEFT THE SECOND ONE SINCE THE CONSTRUCT COULD NO LONGER BE COMPLETED. THE PATIENT WAS 28 YEARS OLD AND APPEARED TO HAVE GOOD BONE. THE DOCTOR DECIDED TO NOT MOVE FORWARD WITH THE REMPLISSAGE AFTER THE ANCHOR FAILED. ADDITIONAL INFORMATION PROVIDED 09/10/2021 THE DEVICE WILL NOT BE RETURNING. THE INSTRUMENTS USED TO PREPARE THE BONE WERE AR-1941P (PUNCH), AND AR-1941CT (GUIDE)

    Patient problems: Insufficient Information

    MAUDE report 19191539 ↗
  • MalfunctionFeb 22, 2024Break; Break

    ON 2/5/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1941PSV PEEK KNOTLESS CORKSCREW ANCHOR BROKE WHEN SHUTTLING THE REPAIR STITCH THROUGH THE ANCHOR BODY. THE SURGEON WAS UNABLE TO TENSION DOWN FULLY. THE ANCHOR WAS REMOVED, AND THE CASE WAS COMPLETED USING ANOTHER AR-1941PSV PEEK KNOTLESS CORKSCREW ANCHOR. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 18761655 ↗
  • InjuryJan 19, 2021Break; Crack

    IT WAS REPORTED THAT A SURGEON WAS DOING A REMPLISSAGE USING 3.9 KNOTLESS CORKSCREWS (AR-1941PSV, LOT 10747047). SURGEON USED A DISPOSABLE KIT (AR-1941DS, LOT 11238107), WITH GUIDE AND PUNCH AND FOLLOWED PROPER TECHNIQUE FOR INSERTION. AS SURGEON BEGAN TWISTING THE ANCHOR IN, SHE HEARD A SOUND AND THEN SAW THE ANCHOR WAS CRACKED. THE ANCHOR WAS ALMOST FULLY INSERTED BUT SITTING A BIT PROUD. SURGEON TRIED TO REMOVE THE CRACKED ANCHOR BODY BUT HAD A HARD TIME SO SHE ENDED UP REMOVING AND CUTTING OUT THE SUTURES AND THE ANCHOR BODY REMAINED IN THE PATIENT BUT DID NOT CAUSE HARM. THE CASE WAS COMPLETED BY USING 2 NEW ANCHORS WITH THE SAME LOT NUMBER.

    Patient problems: Foreign Body In Patient

    MAUDE report 11192523 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →

Sources and updates