Arthrex® AR-2324KBCCTRecalls & reported problems
Arthrex, Inc. · 4.75mm BC KNOTLESS SWIVELOCK w/ TAPE · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867351127), received between Sep 17, 2020 and Dec 2, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Device Dislodged or Dislocated3
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information2
- Failure of Implant1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2025Break; Break
ON 11/12/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE EYELET OF AN AR-2324KBCCT KNOTLESS BIOCOMPOSITE SWIVELOCK ANCHOR FRACTURED AND SNAPPED IN HALF WHILE LOADING TWO SUTURETAPE SUTURES. THE ANCHOR WAS NOT OVERLOADED WITH SUTURE AND WAS LOADED CORRECTLY USING THE MEGALOADER. AS A RESULT, A NEW ANCHOR WAS OPENED AND USED. THIS ISSUE WAS IDENTIFIED DURING A ROTATOR CUFF REPAIR PROCEDURE (B)(6) 2025. ADDITIONAL INFORMATION WAS RECEIVED ON 11/17/2025: THIS ISSUE OCCURRED DURING PREPARATION. ALL DETACHED PIECES WERE ACCOUNTED FOR. THE CASE WAS COMPLETED WITH A REPLACEMENT AR-2324KBCCT KNOTLESS BIOCOMPOSITE SWIVELOCK ANCHOR. THERE WAS NO CASE DELAY OR ADDED ANESTHESIA ADMINISTERED TO THE PATIENT. NO ADVERSE EFFECTS WERE REPORTED FOR THIS EVENT. NO PHOTOS WERE TAKEN OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694979 ↗MalfunctionMay 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION WAS RECEIVED ON 05/02/2025: THIS OCCURRED DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED USING ONE 2.6 RC FIBERTAK FOR THE MEDIAL ROW AND AR-2324KBCCT FOR THE LATERAL ROW, AND BEFORE INSERTION, THE TAPE WAS REMOVED FROM THE EYELET TO ALLOW THE FIBERTAK SUTURE TO PASS THROUGH INSTEAD. THERE WAS A CASE DELAY DUE TO THE NEED TO RETRIEVE THE EYELET AND ANCHOR ARTHROSCOPICALLY AND PREPARE FOR A NEW ANCHOR. BONE QUALITY WAS DECENT, AND THE FIBERTAK INSERTION PROCEEDED WITHOUT ISSUES. THE SILVER PUNCH, AR-1927PB, WAS USED TO CREATE THE PILOT HOLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21953060 ↗MalfunctionFeb 22, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 2/2/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (2) AR-2324KBCCT SWIVELOCK ANCHORS WERE PULLED OUT OF THE BONE WHEN ATTEMPTING TO TENSION THE KNOTLESS MECHANISM IN A MEDIAL ROW FOR A DOUBLE PULLEY RIPSTOP WITH ROTATOR CUFF REPAIR. THE ANCHORS WERE REMOVED AND NO PIECES BROKE INSIDE THE PATIENT. THE CASE WAS COMPLETED WITH NO FURTHER INFORMATION PROVIDED. THIS WAS DISCOVERED DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18758684 ↗MalfunctionFeb 22, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 2/2/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (2) AR-2324KBCCT SWIVELOCK ANCHORS WERE PULLED OUT OF THE BONE WHEN ATTEMPTING TO TENSION THE KNOTLESS MECHANISM IN A MEDIAL ROW FOR A DOUBLE PULLEY RIPSTOP WITH ROTATOR CUFF REPAIR. THE ANCHORS WERE REMOVED AND NO PIECES BROKE INSIDE THE PATIENT. THE CASE WAS COMPLETED WITH NO FURTHER INFORMATION PROVIDED. THIS WAS DISCOVERED DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18758561 ↗MalfunctionOct 6, 2023Break; Break
ON 9/7/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2324KBCCT KNOTLESS BIOCOMPOSITE SWIVELOCK EYELET BROKE COMPLETELY DURING INSERTION, RELEASING THE FIBERTAPE LOOP. WHEN THE SURGEON OPENED THE PACKAGE, IT WAS NOTICED THAT THE EYELET OF THE SWIVELOCK LOOKED DIFFERENT; IT WAS SOMEHOW DEFECTIVE AND APPEARED TO HAVE A CRACK WITH THE FIBERTAPE LOOP PARTIALLY LODGED IN IT. THE SURGEON ATTEMPTED TO USE THE SWIVELOCK ANYWAY, IN WHICH THE EYELET BROKE. EVERYTHING WAS REMOVED, AND THE CASE WAS COMPLETED USING ANOTHER AR-2324KBCCT KNOTLESS BIOCOMPOSITE SWIVELOCK FROM THE SAME LOT NUMBER. THIS WAS DISCOVERED DURING AN RCR PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 9/7/2023: THE SURGEON REMOVED ALL THE BROKEN FRAGMENTS. THE CASE WAS ONLY DELAYED TWENTY TO THIRTY SECONDS AND ADDITIONAL ANESTHESIA WAS NOT NECESSARY.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 17890289 ↗MalfunctionSep 21, 2021Break
ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT WHEN THE SURGEON INSERTED THE AR-2324KBCCT SWIVELOCK ANCHOR INTO THE PATIENT, THE SCREW BROKE MID SHAFT. THE SALES REP EXPLAINED THE PATIENT HAD SOFT BONE. THE FRAGMENT WAS REMOVED AND THE CASE WAS COMPLETED BY USING A NEW AR-2324KBCCT.
Patient problems: Insufficient Information
MAUDE report 12507938 ↗InjurySep 17, 2020Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE THE SURGEON EXPERIENCED THE FOLLOWING ISSUES WITH ARTHREX DEVICES, WHEN INSERTING AN AR-2324KBCCT (LOT 10679428) THE SWIVELOCK WAS INSERTED AND UPON REMOVAL OF THE DRIVER, THE PORTION OF THE EYELET THAT CONTAINS THE KNOTLESS MECHANISM REMAINED IN THE CANNULATION OF THE DRIVER. WHEN INSERTING AN AR-2324KBCCTT (LOT 10941879) UPON FINAL TENSIONING OF THE KNOTLESS MECHANISM THE PORTION OF THE EYELET BROKE COMING UP THROUGH THE SCREW, RESULTING IN CUTTING OF THE SUTURE AND REMOVING THE PEEK PORTION OF THE EYELET. THE DISTAL END OF THE EYELET THAT HAS THE FIBERTAPES THROUGH THEM AND THE SCREWS THEMSELVES STAYED IN THE PATIENT FROM EACH ANCHOR. AN AR-1922P GREEN PUNCH HAD BEEN USED TO PREP THE BONE FOR INSERTION. BONE QUALITY OF THE PATIENT WAS REPORTED TO BE GOOD. THE REPORTER STATES THE PATIENT WAS NOT HARMED IN ANY WAY AND THE SPEEDBRIDGE WAS COMPLETED PER TECHNIQUE GUIDE WITH TWO 5.5MM SWIVELOCKS LATERALLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10546269 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page