Arthrex® AR-2323KPSLCRecalls & reported problems
Arthrex, Inc. · 5.5MM PEEK KNOTLESS SWIVELOCK · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867351349), received between Apr 25, 2022 and May 2, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Positioning Failure2
- Failure to Advance1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
- Laceration(s)1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2025Positioning Failure; Positioning Failure
ON 04/21/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT (2) AR-2323KPSLC PEEK KNOTLESS SWIVELOCK® SUTURE ANCHORS SUTURES PULLED THROUGH THE TIP OR THE PEEK AS THEY IMPLANTED IT. THIS OCCURRED DURING A CASE. THE FIXATION WAS ABLE TO BE LOCKED IN HOWEVER THE SUTURES WERE UNABLE TO BE UTILIZED DUE TO BEING DEFECTIVE. THERE WAS A LESS THAN FIVE MINUTE DELAY THAT REQUIRED ADDITIONAL ANESTHESIA FOR JUST THOSE FIVE. THERE WAS NO HARM ON THE PATIENT. THIS OCCURRED DURING A CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21950492 ↗MalfunctionMay 2, 2025Positioning Failure; Positioning Failure
ON 04/21/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT (2) AR-2323KPSLC PEEK KNOTLESS SWIVELOCK® SUTURE ANCHORS SUTURES PULLED THROUGH THE TIP OR THE PEEK AS THEY IMPLANTED IT. THIS OCCURRED DURING A CASE. THE FIXATION WAS ABLE TO BE LOCKED IN HOWEVER THE SUTURES WERE UNABLE TO BE UTILIZED DUE TO BEING DEFECTIVE. THERE WAS A LESS THAN FIVE MINUTE DELAY THAT REQUIRED ADDITIONAL ANESTHESIA FOR JUST THOSE FIVE. THERE WAS NO HARM ON THE PATIENT. THIS OCCURRED DURING A CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21950035 ↗MalfunctionApr 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-2323KPSLC PEEK KNOTLESS SWIVELOCK FAILED DURING USE. THE FIRST SWIVELOCK FAILURE OCCURRED WHEN THE LOCKING MECHANISM DID NOT HOLD, COMPROMISING THE FUNCTIONALITY OF THE DEVICE. THE SECOND SWIVELOCK FAILURE INVOLVED AN ISSUE WITH THE SUTURE ON THE LOOP, WHERE IT BUNCHED UP AND WAS UNABLE TO PASS THROUGH THE ANCHOR AS INTENDED. THIS WAS DISCOVERED DURING AN RCR PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION REQUESTED ON 4/19/2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21935097 ↗MalfunctionApr 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-2323KPSLC PEEKS KNOTLESS SWIVELOCK FAILED DURING USE. THE FIRST SWIVELOCK FAILURE OCCURRED WHEN THE LOCKING MECHANISM DID NOT HOLD, COMPROMISING THE FUNCTIONALITY OF THE DEVICE. THE SECOND SWIVELOCK FAILURE INVOLVED AN ISSUE WITH THE SUTURE ON THE LOOP, WHERE IT BUNCHED UP AND WAS UNABLE TO PASS THROUGH THE ANCHOR AS INTENDED. THIS WAS DISCOVERED DURING AN RCR PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION REQUESTED ON 4/19/2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21935088 ↗MalfunctionApr 10, 2024Failure to Advance; Failure to Advance
IT WAS REPORTED THAT THE SUTURE WOULD NOT DEPLOY WITH THE ANCHOR. ANOTHER (B)(6) WAS USED TO COMPLETE THE CASE; BICEPS TENODESIS.
Patient problems: Insufficient Information
MAUDE report 19082002 ↗InjuryApr 25, 2022Structural Problem
SURGEON WAS PERFORMING A DOUBLE ROW ROTATOR CUFF REPAIR USING AR-1927PSF-475 MEDIALLY AND AR2323KPSLC LATERALLY. HE HAS RECENTLY BEEN TRIALING THE AR-2323KPSLC (KNOTLESS SWIVELOCK). AFTER COMPLETING THE MAJORITY OF HIS DOUBLE ROW REPAIR, HE TENSIONED HIS SUTURE TAPES THAT HAD BEEN DUNKED IN THE SWIVELOCKS EYELET INTO THE PUNCHED LATERAL HOLE. TWO OF HIS SUTURE TAPES SNAPPED RIGHT AT THE EYELET (IN THE BOTTOM OF THE PUNCHED SWIVELOCK HOLE) AND CAME LOOSE. THIS COMPLETELY DE THREADED HIS DOUBLE ROW REPAIR WITH THE SURGERY ALMOST COMPLETE. THE ANCHORS REMAINED IN SITU AND HE COMPLETED THE ROTATOR CUFF REPAIR WITH AN OPEN APPROACH. AFTER THE CASE WE INSPECTED THE EYELET OF THE KNOTLESS SWIVELOCK IN A STERILE PACKET AND COMPARED IT THE REGULAR SWIVELOCK EYELET. WE NOTICED THERE IS A BREAK IN THE ROUNDED PLASTIC EYELET TO ALLOW FOR THE KNOTLESS MECHANISM. THIS BREAK ALSO EXPOSES THE METAL INSERTER. WE BELIEVE THAT THIS SMALL GAP MADE THE EYE LIT MUCH SHARPER THAN THE REGULAR VERSION CAUSING THE SUTURE TAPES TO SNAP. IN THE NEXT CASE HE DID HIS REGULAR DOUBLE ROW ROTATOR CUFF REPAIR WITH A REGULAR SWIVELOCK AND THERE WAS NO ISSUE, THUS RULING OUT A TECHNIQUE ISSUE. HE HAS DONE HUNDREDS OF
Patient problems: Laceration(s)
MAUDE report 14192977 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page