Arthrex® AR-1941PKRecalls & reported problems
Arthrex, Inc. · PERC kit for 3.9mm Knotless Corkscrew · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867351417), received between Jul 30, 2025 and Feb 24, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent1
- Use of Device Problem1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 24, 2026Material Twisted/Bent
ON 05-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-1941PK PERC KIT FOR 3.9MM KNOTLESS CORKSCREW HAD A BENT DILATOR OUT OF THE PACKAGE AND WOULD NOT SLIDE WITH THE GUIDE DUE TO THE DIFFICULTY OF SLIDING THE GUIDE OVER THE TOP OF THE DILATOR THE SURGEON NEEDED TO USE A SMALLER SWITCHING STICK TO MOVE FORWARD WITH THE CASE. THIS ISSUE WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24428701 ↗MalfunctionJul 30, 2025Use of Device Problem; Use of Device Problem
ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1941PK, TRANSTENDON PERCUTANEOUS INSERTION KIT COMPATIBLE WITH 3.9MM KNOTLESS CORKSCREW®/ KNOTLESS FIBERTAK® ANCHORS, THE CANNULATED SWITCHING STICK WAS BLOCKED AND WAS UNABLE TO BE PASSED OVER THE WIRE. THIS WAS DETECTED DURING USE IN A REMPLISSAGE PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS A SECOND KIT WAS OPENED.
Patient problems: Insufficient Information
MAUDE report 22663080 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.