Skip to content

Arthrex® AR-1941PKRecalls & reported problems

Arthrex, Inc. · PERC kit for 3.9mm Knotless Corkscrew · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867351417), received between Jul 30, 2025 and Feb 24, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1
  • Use of Device Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Material Twisted/Bent

    ON 05-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-1941PK PERC KIT FOR 3.9MM KNOTLESS CORKSCREW HAD A BENT DILATOR OUT OF THE PACKAGE AND WOULD NOT SLIDE WITH THE GUIDE DUE TO THE DIFFICULTY OF SLIDING THE GUIDE OVER THE TOP OF THE DILATOR THE SURGEON NEEDED TO USE A SMALLER SWITCHING STICK TO MOVE FORWARD WITH THE CASE. THIS ISSUE WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428701 ↗
  • MalfunctionJul 30, 2025Use of Device Problem; Use of Device Problem

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1941PK, TRANSTENDON PERCUTANEOUS INSERTION KIT COMPATIBLE WITH 3.9MM KNOTLESS CORKSCREW®/ KNOTLESS FIBERTAK® ANCHORS, THE CANNULATED SWITCHING STICK WAS BLOCKED AND WAS UNABLE TO BE PASSED OVER THE WIRE. THIS WAS DETECTED DURING USE IN A REMPLISSAGE PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS A SECOND KIT WAS OPENED.

    Patient problems: Insufficient Information

    MAUDE report 22663080 ↗