Skip to content

Arthrex® AR-8959TDSRecalls & reported problems

Arthrex, Inc. · TightRope XP Buttress Plate Implant Sys · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867352278), received between Apr 27, 2021 and Apr 30, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1
  • Material Split, Cut or Torn1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2026Positioning Failure

    IT WAS REPORTED THAT DURING A SURGERY TIGHTROPE HASN'T FLIPPED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25044267 ↗
  • MalfunctionFeb 19, 2026Break

    ON 02-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT DURING A SYNDESMOSIS REPAIR PROCEDURE ON (B)(6) 2026, BOTH AN AR-8959TDS SYNDESMOSIS TIGHTROPE XP BUTTRESS PLATE IMPLANT SYSTEM AND AN AR-8925SS SYNDESMOSIS TIGHTROPE XP WERE INSERTED THROUGH THE PLATE WITH THE BUTTONS FLIPPED AS INTENDED. WHEN THE INCLUDED TENSIONING HANDLES WERE USED, THE SUTURES BROKE BEFORE ADEQUATE REDUCTION COULD BE ACHIEVED. THE SAME ISSUE OCCURRED WITH BOTH INITIAL IMPLANTS. THE SURGEON THEN OPENED TWO ADDITIONAL SINGLE-IMPLANT KITS, AND THE SUTURE BREAKAGE RECURRED WITH ONE OF THOSE IMPLANTS AS WELL. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 05-FEB-2026: ON OPENING THE AR-8959TDS SYNDESMOSIS TIGHTROPE XP BUTTRESS PLATE IMPLANT SYSTEM (LOT 15119012), THE TWO INCLUDED TIGHTROPE IMPLANTS EXPERIENCED SUTURE BREAKAGE DURING TENSIONING. THE FACILITY THEN OPENED TWO AR-8925SS SYNDESMOSIS TIGHTROPE XP IMPLANTS, AND ONE OF THESE (LOT 15292557) EXHIBITED THE SAME SUTURE-FRACTURE ISSUE. THE SURGEON REPORTED THAT THE LATERAL ROUND BUTTON WAS GUIDED TO THE PLATE, AND APPROXIMATELY TWO TO THREE TENSIONING CYCLES WERE PERFORMED BEFORE THE SUTURES FRACTURED. TENSIONING WAS APPLIED AL

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24387234 ↗
  • MalfunctionMay 13, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ON 6/10/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT THE TIGHTROPES FROM AN AR-8959TDS TIGHTROPE XP BUTTRESS PLATE IMPLANT SYSTEM RIPPED AND THE IMPLANTS FAILED. SURGEON WAS ABLE TO REMOVED ALL FRAGMENTS AND OPENED INDIVIDUAL TIGHTROPES TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING AN ANKLE FRACTURE PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 19297797 ↗
  • MalfunctionMar 7, 2022Break

    ON 2/14/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A DRILL BIT FROM ANAR-8929TDS SYNDESMOSIS TIGHTROPE KIT BROKE OFF AND REMAINS IN PATIENT. THIS WAS DISCOVERED DURING A SYNDESMOSIS TIGHTROPE PROCEDURE ON (B)(6) 2022. SURGEON DID NOT WANT TO REMOVE THE BROKEN PIECE BECAUSE AFTER SURGERY, PATIENT WILL NOT BE ABLE TO WALK FOR 8 WEEKS AND WILL REASSESS IF NEED BE AFTER PATIENT STARTS WALKING AGAIN.

    Patient problems: Foreign Body In Patient

    MAUDE report 13684340 ↗
  • MalfunctionNov 30, 2021Break

    IT WAS REPORTED THAT DURING A SYNDESMOSIS REPAIR SURGERY WITH INTERNAL BRACE THE DEVICE DID NOT WORK PROPERLY. AFTER TIGHTENING THE SUTURES THE OTHER SUTURE WAS CUT AND THE SURGEON COULD NOT TIGHTEN IT. THE DEVICE BROKE INSIDE THE PATIENT AND ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. HE NEEDED TO CHANGE THE DEVICE TO A SINGLE PACKED TIGHTROPE IMPLANT TO CONTINUE WITH THE PROCEDURE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-8926T). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12896568 ↗
  • MalfunctionApr 27, 2021Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SURGEON USED A AR-8959TDS BB TAK AND PLACED IN THE CENTRAL HOLE OF THE BUTTRESS PLATE TO KEEP THE PLATE STEADY. AFTER USING FLUOROSCOPY, THE SURGEON WANTED TO RELOCATE THE POSITION OF THE PLATE AND WENT TO REMOVE THE BB TAK. AS THE BB TAK WAS BEING REMOVED, THE DISTAL PORTION BROKE OFF AND BECAME EMBEDDED IN THE PATIENT. THE SURGEON WAS ABLE TO USE THE CANNULATED DRILL BIT FROM THE SET TO POP THE CORTEX AND REMOVE A SMALL FRAGMENT OF THE BB TAK; HOWEVER, A SMALL PIECE REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 11730902 ↗

Other WASHER, BOLT NUT devices

All WASHER, BOLT NUT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page