Arthrex® AR-3610CTCRecalls & reported problems
Arthrex, Inc. · Curved Spear, FiberTak, Reusable · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867354265), received between Nov 10, 2023 and Sep 23, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Adverse Event Without Identified Device or Use Problem
It was reported that during the shoulder stabilization surgery, the repair suture of two AR-3636 snapped after they had been placed using the placement instrument AR-3610CTC. As a result, the placement instrument AR-3610CTC was replaced, and the remaining new AR-3636 implants were successfully placed. Per complaint reporter there was no harm for patient, operator or third party. The surgery was finished successfully with a new device with the same part number. It was not necessary to switch the surgical technique or do a second surgery.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869628 ↗MalfunctionApr 28, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY SURGERY THE CURVED GUIDE FOR THE FIBERTAK BECAME OBSTRUCTED WITH THE IMPLANT INSIDE, WHICH HAD FRACTURED. DESPITE MULTIPLE ATTEMPTS, INCLUDING HIGH-PRESSURE IRRIGATION, LUBRICANT, AND OTHER AGENTS, IT COULD NOT BE REMOVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 25021481 ↗MalfunctionDec 30, 2024Break; Break
IT WAS REPORTED THAT DURING A LABRUM FIX WITH FIBER TAK 1.8MM SURGERY THE ANCHOR COULD NOT BE INSERTED FROM THE DRILL SLEEVE AND DID BREAK OFF. THE PIECE DID NOT BREAK OFF INSIDE THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE AND PUSHLOCK REFIX SURGICAL TECHNIQUE. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21039678 ↗MalfunctionMar 29, 2024Break; Break
IT WAS REPORTED THAT DURING A SHOULDER ROTATOR CUFF REPAIR SURGERY THE TEETH OF THE SETTING TOOL BENT/BROKE OFF. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19002774 ↗MalfunctionNov 10, 2023Material Twisted/Bent; Material Twisted/Bent
ON 10/17/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3610CTC CURVED FIBERTAKS DRILL GUIDE WAS BENT. THIS OCCURRED DURING A PROCEDURE; THE 1.8 KNOTLESS FIBERTAK WAS INSERTED IN A STANDARD FASHION USING THE REUSABLE CURVED DRILL GUIDE. AFTER THE INSERTION OF THE ANCHOR, THE GUIDE WAS REMOVED, AND IT WAS NOTICED THAT ONE OF THE TIPS ON THE CROWN OF THE GUIDE WAS BENT. THIS WAS NOT NOTICEABLE AT THE BEGINNING OF THE CASE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18113199 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page