Skip to content

Arthrex® AR-7268Recalls & reported problems

Arthrex, Inc. · FiberTape Cerclage w/o Needle · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867369894), received between Jul 6, 2023 and Aug 7, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Split, Cut or Torn1

Patient problems

  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177418 ↗
  • MalfunctionFeb 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE DEVICE TORE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21353905 ↗
  • InjuryFeb 2, 2024Break; Break

    ON 01/23/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A PATIENT UNDERWENT A CLAVICLE FRACTURE PROCEDURE ON (B)(6) 2023 WHERE AN AR-2652CL CLAVICLE FRACTURE PLATE, AN AR-8835-14 LOW PROFILE SCREW, (2) AR-8835-16 LOW PROFILE SCREWS, AN AR-8835CL-12 KREULOCK COMPRESSION SCREW, (2) AR-8835CL-14 KREULOCK COMPRESSION SCREWS, AN AR-8835-10 LOW PROFILE SCREW, AN AR-7268 FIBERTAPE® CERCLAGE WITH TIGERLINK, AND AN AR-7268T TIGERTAPE® CERCLAGE WITH FIBERLINK WERE IMPLANTED. FIVE WEEKS POST-OP, THE PATIENT REPORTED HEARING OR FEELING A POP. AN X-RAY WAS TAKEN THAT CONFIRMED THAT THE AR-2652CL PLATE HAD BROKEN. ON (B)(6) 2024, A REVISION SURGERY WAS PERFORMED WHERE THE IMPLANTED DEVICES WERE REMOVED. THE REVISION WAS COMPLETED USING A DEVICE FROM ANOTHER MANUFACTURER.

    Patient problems: Failure of Implant

    MAUDE report 18634412 ↗
  • InjuryJul 6, 2023Break; Break

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-7268 CERCLAGE BROKE AFTER IT WAS IMPLANTED INTO A PATIENT. INITIAL SURGERY OCCURRED ON (B)(6) 2023, REVISION SURGERY OCCURRED ON (B)(6) 2023. REVISION SURGERY WAS TO REPLACE SHORT STEM REVERSE SHOULDER WITH A LONG STEM SHOULDER PROSTHESIS IN A PATIENT WITH A FRACTURE OF THE PROXIMAL HUMERUS. CERCLAGE WAS USED AS ONE OF THREE ITEMS TO REPAIR. PATIENT SUFFERED ANOTHER ACCIDENT AFTER THE SURGERY PROMPTING THE REVISION TO REPAIR A SECOND TIME. DURING THE SURGERY THE SURGEON NOTICED THE CERCLAGE WAS BROKEN.

    Patient problems: Failure of Implant

    MAUDE report 17267604 ↗

Other WASHER, BOLT NUT devices

All WASHER, BOLT NUT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page