Arthrex® AR-3210-0043Recalls & reported problems
Arthrex, Inc. · NANONEEDLE SCOPE, 125MM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888867375574), received between Aug 18, 2023 and Feb 11, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Display or Visual Feedback Problem3
- Defective Device1
- Erratic or Intermittent Display1
- Failure to Calibrate1
- Material Deformation1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, WHEN THE MONOPOLAREN ABLATOR WAS USED, THE BUTTONS OF THE NANO NEEDLE WERE TRIGGERED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE DW 29-JAN-2025: FURTHER INFORMATION WAS PROVIDED THAT EVERY TIME AN ABLATOR FROM ANOTHER MANUFACTURER WAS USED, THE FUNCTIONS ASSOCIATED WITH PRESSING A BUTTON WERE TRIGGERED UNCONTROLLABLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21354587 ↗MalfunctionSep 3, 2024Poor Quality Image; Poor Quality Image
ON 08/08/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA SEMS(B)(4) THAT AN AR-3210-0043 NANOSCOPE NEEDLE BECAME BLURRED AFTER REPEATED INSERTION AND REMOVAL. THIS WAS DISCOVERED DURING A CASE. THIS CASE WAS COMPLETED BY NEW PRODUCT OF SAME PRODUCT NUMBER. NO PATIENT HARMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20127187 ↗MalfunctionJun 15, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON 05/01/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-3210-0043 NANONEEDLE FAILED WHEN WENT TO CALIBRATE CAMERA, IT WOULD NOT WHITE BALANCE. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19544084 ↗MalfunctionJun 14, 2024Failure to Calibrate; Failure to Calibrate
ON 2/6/2023 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3210-0043 NANONEEDLE WAS DISPLAYING AN ERROR MESSAGE. THE ERROR MESSAGE HAD A RED X ON THE CONSOLE AND SAID "UNABLE TO CALIBRATE. TRY AGAIN." THIS WAS DISCOVERED DURING A PROCEDURE AND WAS COMPLETED WITH ANOTHER NANONEEDLE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19540594 ↗MalfunctionJun 4, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A WRIST ARTHROSCOPY THERE WERE GROOVES AT THE LOWER END OF THE DEVICE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19463978 ↗MalfunctionMay 21, 2024Defective Device; Defective Device
ON 04/24/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3210-0043 NANONEEDLE SCOPES LIGHT SOURCE DID NOT WORK. THIS OCCURRED DURING A CASE ON 04/17/2024.
Patient problems: Insufficient Information
MAUDE report 19367132 ↗MalfunctionOct 12, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING A DEMO PERONEUS RECON SURGERY OF THE NANONEEDLE AN IMAGE SCATTERING IN THE FORM OF HORIZONTAL LINES ON THE NANONEEDLE MONITOR WAS NOTICED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED WITH THE SAME DEVICE ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17919508 ↗MalfunctionAug 18, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON 7/20/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3210-0043 NANONEEDLE SCOPE AND AN AR-3210-0041 NANOSCOPE OPERATIVE HIGH-FLOW SHEATH KIT HAD AN ISSUE. THE VISUAL QUALITY AND LIGHT SOURCE OUTPUT WAS NOT SUFFICIENT TO ADEQUATELY SEE. TO TROUBLESHOOT, THE NANO WAS RECALIBRATED TWICE AND PLUGGED THE CONSOLE INTO DIFFERENT POWER SOURCES AND THE ISSUE WAS NOT RESOLVED. THE CASE WAS COMPLETED BY OPENING A TRADITIONAL 2.7MM SCOPE. THIS WAS DISCOVERED DURING AN SUBTALAR ARTHROSCOPY PROCEDURE ON (B)(6) 2023, WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17572567 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page