Arthrex® AR-18800-04Recalls & reported problems

Arthrex, Inc. · Driver Shaft, T6, Non-Retaining, AO · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00888867385573), received between Oct 6, 2023 and Feb 4, 2025 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent8
  • Break6
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information4

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Material Twisted/Bent; Material Twisted/Bent

    ON 01/14/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT AND AN AR-18800-05 SOLID SCREWDRIVER WERE TORQUED. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21306929 ↗
  • MalfunctionDec 6, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 11/08/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-03 DRIVER SHAFT AND AR-18800-04 DRIVER SHAFT ENDS ARE TWISTED. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20863666 ↗
  • MalfunctionNov 15, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 10/21/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT TWISTED DURING THE CASE. THE CASE WAS COMPLETED SUCCESSFULLY WITH BACKUP DRIVERS. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694873 ↗
  • MalfunctionNov 5, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 10/21/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT TWISTED DURING THE CASE. THE CASE WAS COMPLETED SUCCESSFULLY WITH BACKUP DRIVERS. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609792 ↗
  • MalfunctionOct 17, 2024Break; Break

    ON 09/26/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT TIP BROKE OFF INSIDE THE PATIENT INSIDE THE SCREW HEAD, AND AN AR-18800-04 DRIVER SHAFT WAS WARPED. THE CASE WAS NOT AFFECTED. THIS WAS DISCOVERED DURING A PROCEDURE ON 09/19/2024, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20471337 ↗
  • MalfunctionOct 17, 2024Material Twisted/Bent; Material Twisted/Bent

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT TIP BROKE OFF INSIDE THE PATIENT INSIDE THE SCREW HEAD, AND AN AR-18800-04 DRIVER SHAFT WAS WARPED. THE CASE WAS NOT AFFECTED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20471212 ↗
  • MalfunctionMay 17, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 04/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4)THAT DURING A PROCEDURE MULTIPLE DEVICES FAILED. DURING THE PROCEDURE AN AR-18800-04 DRIVER SHAFT BECAME TWISTED, AN AR-18800-04 DRIVER SHAFT SNAPPED OFF, (2) AR-18800-03 DRIVER SHAFTS WERE TWISTED, AN AR-18700-39 TORQUE LIMITING ADAPTER WAS NOT CLICKING WHEN THE TORQUE WAS REACHED, A AR-18800-05 SOLID SCREWDRIVER WAS TWISTED, AND A AR-18700-45 POINTED REDUCTION FORCEPS WOULD NOT CLOSE. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19344086 ↗
  • MalfunctionMay 17, 2024Break; Break

    ON 04/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT DURING A PROCEDURE MULTIPLE DEVICES FAILED. DURING THE PROCEDURE AN AR-18800-04 DRIVER SHAFT BECAME TWISTED, AN AR-18800-04 DRIVER SHAFT SNAPPED OFF, (2) AR-18800-03 DRIVER SHAFTS WERE TWISTED, AN AR-18700-39 TORQUE LIMITING ADAPTER WAS NOT CLICKING WHEN THE TORQUE WAS REACHED, A AR-18800-05 SOLID SCREWDRIVER WAS TWISTED, AND A AR-18700-45 POINTED REDUCTION FORCEPS WOULD NOT CLOSE. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19343918 ↗
  • MalfunctionMay 9, 2024Break; Break

    ON 04/16/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-18800-03 DRIVER SHAFTS AND (2) AR-18800-04 DRIVER SHAFTS WERE WARPED AND BREAKING. THIS OCCURRED DURING AN ORIF ANKLE PROCEDURE ON (B)(6)2024. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19278255 ↗
  • MalfunctionMay 9, 2024Break; Break

    ON 04/16/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-18800-03 DRIVER SHAFTS AND (2) AR-18800-04 DRIVER SHAFTS WERE WARPED AND BREAKING. THIS OCCURRED DURING AN ORIF ANKLE PROCEDURE ON (B)(6) 2024. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19278165 ↗
  • MalfunctionApr 15, 2024Material Twisted/Bent; Material Twisted/Bent

    ADDITIONAL INFORMATION WAS RECEIVED ON 3/25/2024: THIS WAS AFTER A METATARSAL FX PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19111880 ↗
  • MalfunctionMar 4, 2024Break; Break

    ON 02/15/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-03 DRIVER SHAFT AND AR-18800-04 DRIVER SHAFT BROKE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18827564 ↗
  • MalfunctionFeb 23, 2024Material Deformation; Material Deformation

    ON 02/6/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-04 DRIVER SHAFT AND AN AR-18800-03 DRIVER SHAFT WARPED WHEN TIGHTENING THE SCREWS. THIS WAS DISCOVERED DURING A PROCEDURE; THE PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18771075 ↗
  • MalfunctionDec 7, 2023Material Twisted/Bent; Material Twisted/Bent

    ON 11/10/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA A (B)(4) THAT (2) AR-18800-04 DRIVER SHAFTS AND AN AR-18800-03 DRIVER SHAFT BECAME TWISTED. THIS OCCURRED DURING A CASE, WITH NO EFFECT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18286126 ↗
  • MalfunctionOct 6, 2023Break; Break

    ON 9/7/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-18800-03 SELF RETAINING DRIVE SHAFT AND AN AR-18800-04 NON-RETAINING DRIVER SHAFT HAD AN ISSUE. WHILE INSERTING A CORTICAL SCREW, THE AR-18800-03 SELF RETAINING DRIVE SHAFT STARTED TO TORQ OUT AND WASN'T RETAINING SCREWS. THE DEVICE WAS SWITCHED TO AN AR-18800-04 NON-RETAINING DRIVER SHAFT, AND WHILE INSERTING ANOTHER SCREW, THE DRIVER'S HEAD BROKE OFF INSIDE THE SCREW HEAD. THE PIECE WAS RECOVERED, AND THE CASE WAS COMPLETED BY SWITCHING TO A SOLID BLUE HANDLE T6 DRIVER IN THE TRAUMA MINI FRAG TRAY 2.0 SCREW CADDY. THIS WAS DISCOVERED DURING A MET SHAFT FRACTURE PROCEDURE ON (B)(6) 2023, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION RECEIVED ON 9/19/2023: THERE WAS A 1-5 MINUTE DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17890291 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →