Arthrex® AR-9508-33Recalls & reported problems

Arthrex, Inc. · Humeral Cup Reamer XS/33 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867406681), received between Jun 5, 2024 and Jun 5, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Dull, Blunt2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2024Dull, Blunt; Dull, Blunt

    ON 05/08/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9508-33 HUMERAL CUP REAMER WAS DULL AND DID NOT REAM AWAY THE BONE SUFFICIENTLY FROM THEIR HUMERAL 2 TRAY. THE CASE WAS COMPLETED USING A BACKUP TRAY WITH NO ISSUES. THIS WAS DISCOVERED DURING A REVERSE TOTAL SHOULDER PROCEDURE, WITH NO PATIENT HARM REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 5/13/2024: THIS OCCURRED ON (B)(6) 2024 DURING REVERSE TOTAL SHOULDER PROCEDURE. THERE WAS NO DELAY IN THE CASE OTHER THAN OPENING ANOTHER TRAY. THE AR-9508-33 HUMERAL CUP REAMER DID NOT BREAK, BUT WAS WORN AND DID NOT WORK AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19473450 ↗
  • MalfunctionJun 5, 2024Dull, Blunt; Dull, Blunt

    ON 05/08/2024 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(3) THAT AN AR-9508-33 HUMERAL CUP REAMER WAS DULL AND DID NOT REAM AWAY THE BONE SUFFICIENTLY FROM THEIR HUMERAL 2 TRAY. THE CASE WAS COMPLETED USING A BACKUP TRAY WITH NO ISSUES. THIS WAS DISCOVERED DURING A REVERSE TOTAL SHOULDER PROCEDURE WITH NO PATIENT HARM REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 5/13/2024: THIS OCCURRED ON (B)(6) 2024 DURING REVERSE TOTAL SHOULDER PROCEDURE. THERE WAS NO DELAY IN THE CASE OTHER THAN OPENING ANOTHER TRAY. THE AR-9508-33 HUMERAL CUP REAMER DID NOT BREAK, BUT WAS WORN AND DID NOT WORK AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19467048 ↗

Other Orthopedic manual surgical instrument devices