The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867462038), received between Aug 5, 2024 and Mar 20, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device1
- Material Split, Cut or Torn1
- Material Twisted/Bent1
- Sharp Edges1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Sharp Edges
IT WAS REPORTED THAT DURING THE TRANSFORAMINAL INTERVERTEBRAL DISC SURGERY, IT WAS DISCOVERED THAT THE DILATOR HAD SHARP EDGES. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24650990 ↗MalfunctionDec 2, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ON 10/11/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-S6524-041-220, AR-S6524-051-210, AR-S6524-060-200, AND AR-S6524-071-190 DILATOR IS CHIPPED AND BENT AFTER REPEATED USE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20818877 ↗MalfunctionOct 29, 2024Material Twisted/Bent; Material Twisted/Bent
ON 10/11/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-S6524-041-220, AR-S6524-051-210, AR-S6524-060-200, AND AR-S6524-071-190 DILATOR IS CHIPPED AND BENT AFTER REPEATED USE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20557391 ↗MalfunctionAug 5, 2024Defective Device; Defective Device
ON 07/09/2024, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-S6524-051-210 DILATOR HAD A PIECE OF METAL FROM THE INNER CANNULATION THAT STOPPED IT FROM GOING OVER THE SECOND SEQUENTIAL DILATOR. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT AND THE CASE WAS COMPLETED SUCCESSFULLY. THIS WAS DISCOVERED DURING AN ENDOSCOPIC DISCECTOMY PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19907594 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026