Arthrex® AR-1288QAI-90Recalls & reported problems

Arthrex, Inc. · All-Inside QuadLink Kit, 9mm · Class 1 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867473584), received between Mar 20, 2026 and Jul 29, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Break1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2026Dull, Blunt

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50068995 ↗
  • MalfunctionApr 1, 2026Break

    ON 12-MAR-2026, A FACILITY REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-1288QAI-90 FINAL TENSIONING, THE QUAD TIGHT ROPE BROKE DURING FINAL TENSION. GRAFT HAD TO BE REMOVED AND REPLACED, NO PATIENT HARM. APPROXIMATELY 20¿30-MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24755759 ↗
  • MalfunctionMar 25, 2026Material Split, Cut or Torn

    ON 05-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1288QAI-90 ALL-INSIDE QUADLINK KIT EXPERIENCED A FIBERTAG TIGHTROPE FAILURE. THE ISSUE WAS IDENTIFIED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2026. THERE WAS NO REPORTED PATIENT HARM, AND THE SURGERY WAS COMPLETED. ADDITIONAL INFORMATION WAS RECEIVED ON 11-MAR-2026: THE ISSUE OCCURRED INSIDE THE PATIENT DURING GRAFT TENSIONING INTO THE FEMORAL TUNNEL. AS TENSION WAS APPLIED, THE SUTURE FRAYED AND RIPPED OFF THE METAL BUTTON, RESULTING IN FAILURE OF THE DEVICE. ALL DETACHED FRAGMENTS WERE ACCOUNTED FOR AND RETRIEVED FROM THE PATIENT DURING THE PROCEDURE. THE CASE WAS COMPLETED USING AN AR-1588BTB-IB TIGHTROPE II BTB. THE INCIDENT CAUSED A BRIEF DELAY OF APPROXIMATELY 5¿10 MINUTES, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24689429 ↗
  • MalfunctionMar 20, 2026Material Split, Cut or Torn

    ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1288QAI-90 ALL-INSIDE QUADLINK KIT, 9 MM, EXPERIENCED AN ISSUE DURING AN ARTHROSCOPIC-AIDED ACL RECONSTRUCTION PROCEDURE. AT THE FINAL STAGE OF ALL-INSIDE QUAD ACL RECONSTRUCTION, THE SUTURE ON THE FEMORAL TIGHTROPE SNAPPED, RENDERING THE IMPLANT NONFUNCTIONAL. THE GRAFT HAD BEEN SHUTTLED INTO THE JOINT, THE FEMORAL BUTTON WAS FLIPPED ON THE CORTEX, AND THE TIGHTROPE WAS PROVISIONALLY TENSIONED TO DRAW THE GRAFT INTO THE FEMORAL SOCKET. ONCE APPROXIMATELY 10 MM OF GRAFT WAS WITHIN THE FEMORAL SOCKET, THE TIBIAL SIDE WAS PROVISIONALLY TENSIONED UNTIL APPROXIMATELY 15 MM OF GRAFT WAS SEATED WITHIN THE TIBIAL TUNNEL. AT THAT POINT, THE SURGEON REMOVED THE SCOPE AND RETURNED TO THE FEMORAL TIGHTROPE TO COMPLETE FINAL TENSIONING. WHILE PULLING THE WHITE SUTURES INDIVIDUALLY, THE TIGHTROPE SNAPPED. ALL FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED BY REMOVING THE GRAFT FROM THE JOINT AND APPLYING NEW AR-1588RTT2-IB FIBERTAG TIGHTROPE II IMPLANTS TO THE GRAFT. THE GRAFT WAS THEN REINSERTED, AND THE PROCEDURE WAS COMPLETED. A CASE DELAY OF APPROXIMATELY THIRTY MINUTES, WITH ADDITIONAL ANESTHESIA ADMINISTERED, OCCURRED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650853 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →