Arthrex® AR-1288QAI-90Recalls & reported problems
Arthrex, Inc. · All-Inside QuadLink Kit, 9mm · Class 1 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867473584), received between Mar 20, 2026 and Jul 29, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn2
- Break1
- Dull, Blunt1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2026Dull, Blunt
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50068995 ↗MalfunctionApr 1, 2026Break
ON 12-MAR-2026, A FACILITY REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-1288QAI-90 FINAL TENSIONING, THE QUAD TIGHT ROPE BROKE DURING FINAL TENSION. GRAFT HAD TO BE REMOVED AND REPLACED, NO PATIENT HARM. APPROXIMATELY 20¿30-MINUTE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755759 ↗MalfunctionMar 25, 2026Material Split, Cut or Torn
ON 05-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1288QAI-90 ALL-INSIDE QUADLINK KIT EXPERIENCED A FIBERTAG TIGHTROPE FAILURE. THE ISSUE WAS IDENTIFIED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2026. THERE WAS NO REPORTED PATIENT HARM, AND THE SURGERY WAS COMPLETED. ADDITIONAL INFORMATION WAS RECEIVED ON 11-MAR-2026: THE ISSUE OCCURRED INSIDE THE PATIENT DURING GRAFT TENSIONING INTO THE FEMORAL TUNNEL. AS TENSION WAS APPLIED, THE SUTURE FRAYED AND RIPPED OFF THE METAL BUTTON, RESULTING IN FAILURE OF THE DEVICE. ALL DETACHED FRAGMENTS WERE ACCOUNTED FOR AND RETRIEVED FROM THE PATIENT DURING THE PROCEDURE. THE CASE WAS COMPLETED USING AN AR-1588BTB-IB TIGHTROPE II BTB. THE INCIDENT CAUSED A BRIEF DELAY OF APPROXIMATELY 5¿10 MINUTES, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689429 ↗MalfunctionMar 20, 2026Material Split, Cut or Torn
ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1288QAI-90 ALL-INSIDE QUADLINK KIT, 9 MM, EXPERIENCED AN ISSUE DURING AN ARTHROSCOPIC-AIDED ACL RECONSTRUCTION PROCEDURE. AT THE FINAL STAGE OF ALL-INSIDE QUAD ACL RECONSTRUCTION, THE SUTURE ON THE FEMORAL TIGHTROPE SNAPPED, RENDERING THE IMPLANT NONFUNCTIONAL. THE GRAFT HAD BEEN SHUTTLED INTO THE JOINT, THE FEMORAL BUTTON WAS FLIPPED ON THE CORTEX, AND THE TIGHTROPE WAS PROVISIONALLY TENSIONED TO DRAW THE GRAFT INTO THE FEMORAL SOCKET. ONCE APPROXIMATELY 10 MM OF GRAFT WAS WITHIN THE FEMORAL SOCKET, THE TIBIAL SIDE WAS PROVISIONALLY TENSIONED UNTIL APPROXIMATELY 15 MM OF GRAFT WAS SEATED WITHIN THE TIBIAL TUNNEL. AT THAT POINT, THE SURGEON REMOVED THE SCOPE AND RETURNED TO THE FEMORAL TIGHTROPE TO COMPLETE FINAL TENSIONING. WHILE PULLING THE WHITE SUTURES INDIVIDUALLY, THE TIGHTROPE SNAPPED. ALL FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED BY REMOVING THE GRAFT FROM THE JOINT AND APPLYING NEW AR-1588RTT2-IB FIBERTAG TIGHTROPE II IMPLANTS TO THE GRAFT. THE GRAFT WAS THEN REINSERTED, AND THE PROCEDURE WAS COMPLETED. A CASE DELAY OF APPROXIMATELY THIRTY MINUTES, WITH ADDITIONAL ANESTHESIA ADMINISTERED, OCCURRED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24650853 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page