DJO SURGICAL 010-55-025Recalls & reported problems
Encore Medical, L.P. · SCREW, CANCELLOUS, 25MM · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888912069977), received between Jul 27, 2016 and Nov 14, 2017 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K931665); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Detachment Of Device Component2
- Device Slipped1
- Inadequacy of Device Shape and/or Size1
Patient problems
- No Information2
- Device Embedded In Tissue or Plaque1
- Discomfort1
- No Code Available1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 14, 2017Inadequacy of Device Shape and/or Size
REVISION SURGERY - DUE TO THE PATIENT COMPLAINING OF DISCOMFORT. THE SCREW WAS A LITTLE TOO LONG.
Patient problems: Discomfort; No Code Available
MAUDE report 7029653 ↗InjurySep 27, 2017Break; Detachment Of Device Component
COMPLAINT/PRIMARY SURGERY - THE SCREW HEAD BROKE OFF THE SHAFT AS THE SURGEON WAS SCREWING IT IN THE ACETABULAR CUP.
Patient problems: No Information
MAUDE report 6899486 ↗InjuryDec 22, 2016Device Slipped
PRIMARY SURGERY - THE ACETABULAR SCREW POST-OPERATIVELY WAS SCREWED THROUGH THE SCREW HOLE. THIS WAS DETERMINED ON THE POST-OP X-RAY.
Patient problems: No Information
MAUDE report 6201434 ↗InjuryJul 27, 2016Break; Detachment Of Device Component
PRIMARY SURGERY - DUE TO THE SCREW HEAD BREAKING OFF WHILE IMPLANTING THE SCREW. IT WAS DISCOVERED ON (B)(6) 2016 THAT PART OF THE SCREW REMAINED IN THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 5826596 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K931665. One submission often covers many models, so check the product and lot details in each record.
- Z-0139-2013 · Sep 28, 2012 · Terminated6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.
- Z-2146-2012 · Jul 16, 2012 · Terminateddjo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page