DJO SURGICAL 803-00-015Recalls & reported problems
Djo, LLC · HANDLE, FEMORAL STEM INSERTER · Class 1
All models of DJO SURGICAL (42)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888912088541), received between Sep 10, 2018 and May 14, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
Patient problems
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2021Break
PRODUCT FAILLURE - THE TIP OF THE THREADED STEM INSERTER HANDLE BROKE OFF AND IS NO LONGER USABLE BECAUSE IT PHYSICALLY CANNOT THREAD INTO A TAPER FILL OR LINEAR STEM.
Patient problems: Insufficient Information
MAUDE report 11829532 ↗MalfunctionOct 7, 2020Break
INSTRUMENT FAILURE - SURGEON BROUGHT BACK A PATIENT SHOWING INFECTION SYMPTOMS. THE DECISION WAS MADE TO REMOVE ALL PRIMARY HIP COMPONENTS UNTIL INFECTION CLEARED. IN AN ATTEMPT TO REMOVE THE LINEAR FEMORAL STEM IMPLANT, THE THREADED INSERTER HANDLE WAS SCREWED INTO THE IMPLANT AND USED TO EXPLANT THE STEM. DURING THE ATTEMPT TO EXPLANT THE THREADS ON THE HANDLE BROKE OFF. ALL PARTS WERE REMOVED AND THE EXTRACTION CONTINUED TO NO ADVERSE EVENTS.
Patient problems: Unspecified Infection
MAUDE report 10641883 ↗MalfunctionSep 10, 2018Break; Detachment of Device or Device Component
INSTRUMENT FAILURE - DURING IMPACTION OF FEMORAL HIP IMPLANT, INTO THE FEMORAL CANAL, THE TIP OF THE INSERTER BROKE OFF. THE IMPLANT WAS SUCCESSFULLY INSERTED INTO THE CANAL ALONG WITH THE BROKEN TIP INSIDE THE IMPLANT'S THREADED HOLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7862103 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page