Wallach 900629Recalls & reported problems
Coopersurgical, Inc. · LL100 Two-Trigger (with Thermocouple) Multi-Tip Cryosurgical System for N2O · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888937009866), received between Dec 9, 2020 and Apr 7, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Defective Device1
- Excessive Cooling1
- Mechanical Problem1
- Physical Resistance/Sticking1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 7, 2021Mechanical Problem; Defective Device; Temperature Problem
THE FOLLOWING MULTI TIP FREEZER OF THE 900629 APPEARS TO BE DEFECTIVE. THERE IS A FAULT IN THE FROST/DEFROST VALVE WHICH PREVENTS PROPER CIRCULATION. INCIDENT OCCURRED BEFORE PROCEDURE. DURING DEFROSTING, THE FREEZING POINT REMAINS BELOW 0 DEGREES DUE TO A MALFUNCTION IN THE FROST-DEFROST CIRCULATION. NO PROBLEMS WITH PATIENT. NO EXTRA ATTENTION WAS NEEDED. 1216677-2021-00063-1 LL100 MULTI TIP W TC 900629 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11628669 ↗MalfunctionJan 11, 2021Excessive Cooling
REPORT SUBMITTED BY CSI SERVICE & REPAIR TEAM- "WOULD NOT DEFROST." REF (B)(4). LL100 MULTI TIP W-TC 900629 E-COMPLAINT: (B)(4).
Patient problems: Insufficient Information
MAUDE report 11148302 ↗MalfunctionDec 9, 2020Physical Resistance/Sticking
"FREEZE VALVE IS STUCK IN THE OPEN POSITION". 900629 LL100 MULTI TIP WTC E-COMPLAINT-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10980613 ↗
FDA clearance and classification
Cleared through 510(k) K813024 on Dec 3, 1981 (decision: Substantially Equivalent). Submitted by Wallach Surgical Devices, Inc. as “Wa1000”. The FDA review took 36 days. FDA record ↗
The FDA regulates product code GEH (unit, cryosurgical, accessories) under 21 CFR 878.4350 as a Class II device. 21 CFR 878.4350 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Wallach 900629
Has the Wallach 900629 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888937009866), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Wallach 900629?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 9, 2020 and Apr 7, 2021. In those reports, the most common specific device problems are “Defective Device” (1), “Excessive Cooling” (1) and “Mechanical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Wallach 900629?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Wallach 900629 FDA approved?
It was cleared through 510(k) K813024 on Dec 3, 1981. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Wallach 900629, and is it still sold?
The FDA's device database lists Coopersurgical, Inc. as the company responsible for it (brand Wallach, model 900629). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, Cryosurgical, Accessories devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths269 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls6 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IceFORCE® 2.1 CX Prostate Cryoablation Kit / Visual-ICE® System FPRPR3605Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
- Three IceFORCE ® 2.1 CX Prostate Cryoablation Kit Visual-ICE® System FPRPR3607Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page